No CrossRef data available.
Published online by Cambridge University Press: 25 February 2011
An experimental protocol using secondary ion mass spectrometry to measure Al in vapor phase decomposition (VPD) solutions has been developed to achieve a detection limit of 108 atoms/cm2. The analytical technique is called VPD/SIMS. The procedure utilizes sample preparation techniques developed for microvolume SIMS. Preliminary SIMS analysis of simulated VPD solutions consisting of dilute HF indicate that a theoretical detection limit of 5.7×105 Al atoms/cm2 is possible from a 6″ wafer.