This study aimed to assess the prevalence of “silent” valvar regurgitation in healthy Brazilian children; to determine whether variations in the gain settings of the machine interfered with its documentation; and to compare Doppler color flow mapping with the pulsed wave Doppler. Twenty-five healthy children were selected randomly. Echocardiographic equipment used was ALOKA 870. Fixed parameters were 3.5 MHz transducer, 400 Hz filter, reject of 7/14 and aperture of 30/80. Variable parameters were Doppler color flow mapping and pulsed wave Doppler gains. A protocol for evaluation of each cardiac valve was developed. Regurgitation was noted in 93% for the pulmonary valve, 74% for the tricuspid valve and 27% for the mitral valve. In the aortic valve, jets suggestive of mild valvar regurgitation were observed with the Doppler color flow mapping technique in two children but these findings were not corroborated by pulsed wave Doppler. Only in the aortic valve were findings from Doppler color flow mapping not in agreement with those from pulsed wave Doppler. Variations in the settings for gain did not interfere significantly in the documentation of regurgitation. Inter-observer variations were not significant (two independent operators). The clinical implications of these findings are yet to be established. Labeling these regurgitant jets as “normal” may be oversimplifying their meaning and perhaps overlooking the possibility of predisposition to endocarditis or progression of the valvar abnormality. Conversely, labeling them as “abnormal” can generate a more serious problem, which is that of iatrogenic heart disease. Prospective follow-up studies should be carried out in these subjects.