We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
To evaluate the dosimetric errors associated with the effect of the collimator angle error in volumetric-modulated arc therapy (VMAT) treatments.
Methods and materials
Four patients with different planning target volume (PTV) and localisations treated using VMAT were analysed (high-risk prostate, low-risk prostate, head and neck (H&N) and holocranial with hippocampus protection) in terms of dosimetric variations when errors in the collimator angle were introduced. Original plans underwent modifications of the planned collimator angles of ±0·5°, ±1° and ±1·5°. These modified plans were re-calculated using the same original plan fluencies, and the resulting dose–volume histograms and homogeneity index (HI-ICRU) were compared.
Results
For the high-risk prostate case, there was a noticeable loss of PTV dose coverage for collimator angle errors larger than ±1°, with HI-ICRU relative variations up to 75% in the range analysed. The low-risk prostate case did not present significant changes in organs at risk or PTV dose coverage. For the H&N case, the spinal cord presented changes around 4% for D0·1 cc. In the holocranial case, optic lens showed dose variations up to 5% for collimator angle errors larger than ±1°.
Conclusions
The effect of the collimator error in VMAT increased as the PTV increased.
For selecting the position of the isocentre, one should be cautious, and whenever possible choose a position close to the geometrical centre of the PTVs in order to avoid or minimise errors from the calibration of the collimator angle.