We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Shape-memory abnormalities are seen in some nitinol atrial septal occluders. Variably described as cobra-head, tulip, and others, their incidence, mechanisms, clinical impact, and outcome have not been systematically analysed.
Methods:
We retrospectively reviewed all consecutive device closures in the last 6 years for deformations. Type and size of the occluder, deployment technique, size, and angulation/kinking of the delivery sheath were analysed. Procedural success, duration, and other complications were studied.
Results:
A total of 112 devices (11.8%) among 950 occluders used in 936 patients showed deformities. Fourteen of 936 received 2 devices. Deformities were transient and self-correcting in 40%. Multivariate analysis showed significant associations with oversized sheaths (p = 0.004), kinked/angulated sheaths (p < 0.001), special deployment techniques (p < 0.001), and twist in the device waist (p = 0.011). Despite more frequent deformities with Figulla (15.6%) and Amplatzer (13.9%) occluders than Cera occluders (6.6%) and larger devices (>24 mm – 14.6%) than smaller devices (less than or equal to 24 mm – 9.7%), they were not significant on multivariate analysis. In vivo manipulations corrected most deformities; nineteen needed in vitro reformations and four needed a change of device. Despite prolongation of the procedure, repeated attempts (mean 2.76 ± 1.7 attempts, with a range from 1 to 9 attempts), and supraventricular tachycardia in two patients, there were no serious adverse effects.
Conclusions:
Deformations were frequent in 11.8% of atrial septal occluders on a targeted search. Oversized and angulated/kinked sheaths, special techniques like pulmonary vein deployment and twist in device waist during procedure predisposed to deformities. While most deformities were corrected with manipulations, removal of the device was infrequently needed and change of device was rarely required. Long procedural time and multiple attempts for deployment did not affect procedural success.