Hostname: page-component-cd9895bd7-jkksz Total loading time: 0 Render date: 2024-12-28T13:56:18.597Z Has data issue: false hasContentIssue false

FIRST δ13C RESULTS WITH A NEW CONNECTION BETWEEN THE EA-IRMS SYSTEM AND THE GAS INJECTION SYSTEM AT COLOGNEAMS

Published online by Cambridge University Press:  20 September 2024

Martina Gwozdz*
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
Alexander Stolz
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
Andrea Jaeschke
Affiliation:
University of Cologne, Institute for Geology and Mineralogy, Zuelpicher Str. 49b, Cologne, 50674, Germany
Stefan Heinze
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
Ramona Mörchen
Affiliation:
Bonn University, Institute of Crop Science and Resource Conservation, Soil Science and Soil Ecology, Karlrobert-Kreiten-Strasse 13, Bonn, 53115, Germany
Alfred Dewald
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
Janet Rethemeyer
Affiliation:
University of Cologne, Institute for Geology and Mineralogy, Zuelpicher Str. 49b, Cologne, 50674, Germany
Dennis Mücher
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
Markus Schiffer
Affiliation:
University of Cologne, Institute for Nuclear Physics, Zuelpicher Str. 77, Cologne, 50937, Germany
*
*Corresponding author. Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

This work presents the integration of an elemental analyzer (EA) and an isotope ratio mass spectrometer (IRMS) into the 6 MV AMS system at the Institute for Nuclear Physics, University of Cologne. The AMS measurement of δ13C values for IAEA-C6 reference material resulted in –11.39(226)‰, compared to –10.28(32)‰ obtained by IRMS. The EA-IRMS system was also tested with IAEA-C3, IAEA-C5, and IAEA-C7 reference materials, yielding –24.79(9), –25.18(15), and –14.76(18)‰ respectively. Compared to the IAEA information values given as –24.91(49), –25.49(72) and –14.48(21)‰ respectively. To investigate an observed sample mass dependency, environmental samples from Spitzbergen were examined, showing δ13C values of –25.17(55), –25.80(31), and –26.17‰ in Cologne, while Hamburg recorded –24.8(1), –25.5(1), and –26.2(13)‰. In summary, this new setup could enable online analysis and quasi-simultaneous measurements of 14C, δ13C, and δ15N for ultra-small samples, utilizing precise δ13C values from IRMS for fractionation correction of the 14C/14C isotopic ratio.

Type
Conference Paper
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona

INTRODUCTION

At CologneAMS, it is possible to measure CO2 samples with masses as low as 2 μg (Melchert et al. Reference Melchert, Stolz, Dewald, Gierga, Wischhoefer and Rethemeyer2018). This is achieved by oxidizing the sample within an elemental analyzer (EA) and then directing it, using the Gas Injection System (GIS), to the AMS ion source without the need for graphitization steps.

To meet the increasing demand for 14C/12C isotopic ratio measurements of ultra-small samples (1–20 μg), improvements in isotopic fractionation correction are necessary to enhance measurement accuracy and address dating challenges in archives with low carbon content. Fractionation effects can occur during various stages of the AMS measurement, including sample preparation and oxidation, negative ionization during the sputter process, and stripping process (Hofmann et al. Reference Hofmann, Bonani, Morenzoni, Nessi, Suter and Woelfli1984; Arnold et al. Reference Arnold, Bard, Maurice, Valladas and Duplessy1989; Alderliesten et al. Reference Alderliesten, Van Der Borg and De Jong1998). While AMS can directly determine δ13C values, one study suggest that values derived from IRMS exhibit greater precision, up to two orders of magnitude (Prasad et al. Reference Prasad, Culp and Cherkinsky2019). Furthermore McIntyre et al. (Reference McIntyre, Wacker, Haghipour, Blattmann, Fahrni, Usman, Eglinton and Synal2016) investigated the International Atomic Energy Agency (IAEA) standard IAEA-C6 with an information value δ13C value of –10.80(47)‰ (Rozanski et al. Reference Rozanski, Stichler, Gonfiantini, Scott, Beukens, Kromer and van de Plicht1992) and demonstrated an improvement in the δ13C value from –12.17(224)‰ measured by AMS to –10.50(2)‰ measured by IRMS (McIntyre et al. Reference McIntyre, Wacker, Haghipour, Blattmann, Fahrni, Usman, Eglinton and Synal2016). The notation “12.17(224)‰” indicates the value 12.17‰ with an associated uncertainty of ±2.24‰.

Motivated by these findings, a new system has been installed at CologneAMS, which combines the EA and IRMS with the existing GIS. This enables the determination of δ13C values using the IRMS instead of the AMS. During sample oxidation, around 10% of the CO2 is directed to the IRMS for a δ13C measurement, while the remaining gas is directed to the GIS. From the GIS, the sample can be transferred to the HVE SO-110 B ion source, which was installed in 2015 (Stolz et al. Reference Stolz, Dewald, Altenkirch, Herb, Heinze, Schiffer, Feuerstein, Mueller-Gatermann, Wotte, Rethemeyer and Dunai2017). Additionally, this setup will allow for simultaneous IRMS-AMS measurements of gaseous AMS samples from glass ampules, which are cracked within the GIS.

MATERIALS AND METHODS

The setup consists out of four components. The first is the EA, where samples are oxidised and transferred to the second component, the IRMS. Samples are analyzed for their nitrogen and carbon content in the EA while the IRMS measures the δ13C or δ15N values. From there the sample is captured in the syringe of the GIS, where it can be led into the ion source of the AMS.

EA and IRMS

For combustion, all solid samples are placed in cleaned tin boats and then oxidized in the EA. The used device is a vario isotope select which is sold by Elementar, Germany (serial no.15191004). By introducing O2 into the combustion reactor of the EA, the sample undergoes oxidation and becomes gaseous. Helium is used as a carrier gas to transport the sample throughout the system. Within the reduction tube, the NO and NO2 gas are reduced to N2, while other combustion products are filtered out.

After exiting the reduction tube, the gas is separated using thermal programmed desorption (TPD). CO2 is captured at approximately 38°C, while the nitrogen gas reaches the thermal conductivity detector (TCD) of the EA. The TCD signal provides a peak area indicating the concentration of the incoming gas components. The TCD peak areas are calibrated using reference material. The reference material has a known elemental content which is used to calibrate the measured peak areas of the carbon and nitrogen peaks. Once the nitrogen gas passed through, the TPD unit is heated to 210°C, causing the release of CO2, which is then detected by the TCD.

The isoprime precisION IRMS (by Elementar) comprises an ion source that generates positive ions, an electromagnet for momentum separation, and an electrostatic analyzer (ESA). The nitrogen and carbon gases reach the ion source of the IRMS, where they undergo ionization through electron interaction. The positive ions are directed into the electromagnet, and masses 44, 45, and 46 for CO2 and 28, 29, and 30 for N2 are separated. They are individually detected in three Faraday cups.

Gas Injection System

The GIS is a well-established component of the setup at CologneAMS. Its purpose is to capture CO2 gas from the EA using a zeolite trap and subsequently release it into a syringe. Inside the syringe, the CO2 gas is mixed with helium before being transferred to the AMS ion source. The zeolite trap used in this setup has a capacity of up to 1000 μg of carbon. Further details of the GIS can be found in Stolz et al. (Reference Stolz, Dewald, Heinze, Altenkirch, Hackenberg, Herb, Mueller-Gatermann, Schiffer, Zitzer, Wotte, Rethemeyer and Dunai2019, Reference Stolz2020) and Wacker et al. (Reference Wacker, Fahrni, Hajdas, Molnar, Synal, Szidat and Zhang2013).

Connection between EA-IRMS and AMS

This section provides an overview of the newly installed hardware and software connections that link the EA, IRMS, and AMS systems. There are two possibilities to connect these systems. Figure 1 illustrates the general layout of the two setups. In Figure 1 i), the connection is shown, where samples are oxidized in the EA and a percentage of the gas is transferred to the IRMS, while capturing the rest in the GIS. Figure 1 ii) shows the pathway for sample gas which was cracked and then captured in the syringe of the GIS. A portion of the gas can be led to the IRMS, while retaining the remainder for a measurement with the AMS.

Figure 1 The schematic diagram illustrates the connection between the EA, IRMS, GIS, and AMS systems. In i), the connection is established for a sample which is oxidized in the EA. Subsequently, the gaseous sample is transferred through a capillary to a T-piece. Around 10 percent of the sample is directed towards the IRMS through a capillary (cyan) for measurement, while the remaining portion is further transferred to the GIS for subsequent analysis in the AMS ion source. Figure 1 ii) depicts the setup for pre-existing gaseous samples that can be cracked within the ampoule cracker of the GIS. The cracked sample is then transferred to the IRMS under a continuous flow of helium. This is accomplished by splitting an existing capillary, which supplies helium to the GIS, using a T-piece and an additional capillary (red).

For the software connection the EA and the IRMS can be controlled via an application programming interface (API). The API sends out the requests to start or stop a measurement and also collects the incoming data from the various components. The API allows the gas injection control software (GICS) (Stolz et al. Reference Stolz2020) to communicate with the IRMS, this way the GICS software can fully control the measurement.

For samples combusted in the EA, the EA and IRMS are connected via 1/16-inch diameter tubing. The oxidized sample is transferred through this tubing. A T-piece creates a split, and an additional capillary with a 0.005-inch inner diameter is attached, leading to the IRMS inlet. This setup allows approximately 10% of the sample to enter the IRMS inlet for IRMS measurement, while the remaining portion is directed to the GIS. This is visualized in Figure 1 i). From the EA inlet, the sample gas flows towards a cold zeolite trap, where CO2 is captured. Once all the CO2 is inside the zeolite trap, it is heated to 450°C, allowing the CO2 gas to diffuse into the syringe.

To enable the measurement of pre-existing gaseous samples, an additional connection has been established between the GIS and the IRMS. In future applications, CO2 samples in glass ampules will be cracked using the GIS ampule cracker, and the resulting gas is transferred to the syringe. A small percentage of the sample gas can be directed into the IRMS, while the remaining gas is retained in the syringe for analysis using the AMS setup. Continuous helium flow is necessary for the operation of the IRMS ion source as the carrier gas for CO2, as depicted in Figure 1 ii). The capillary responsible for conveying CO2 from the GIS to the IRMS is equipped with a T-piece, through which helium is introduced. To achieve a low helium flow, a capillary (red) with a 0.005-inch inner diameter is connected via a T-piece to the main helium line, which also supplies the GIS.

The introduction of CO2 into the IRMS is accomplished using an open split. The GIS includes a GIS outlet #2 with a tubing of 1/16-inch diameter, which is inserted into the GC capillary of the IRMS. A piece of plastic is placed around the open split and heated. As the plastic shrinks, it secures both capillaries in place.

RESULTS AND DISCUSSION

EA and IRMS

This section provides an overview of the measurements conducted on different standard materials and environmental samples using the EA and the IRMS. It should be noted that the EA-IRMS setup was not connected to the GIS or AMS for these measurements.

IRMS Standard Samples

Various standard materials were oxidized in the EA, and their δ13C values were determined using the IRMS (see Table 1). The initial tests utilized the following reference material from the IAEA: IAEA-C3, IAEA-C 5, IAEA-C6, as well as IAEA-C7 and IAEA-C8. The δ13C information values of these reference materials are provided relative to the Pee Dee belemnite scale (IAEA 2014).

Table 1 The measured δ13C values, along with their corresponding standard deviations, are compared to the information value values provided by the IAEA. The IAEA information values are denoted as δ13CPDB, representing their expression relative to Pee Dee belemnite (PDB). The data obtained in this work is presented as δ13CCologne.

Over a three-month period, samples of each standard weighing between 0.030 mg and 2.000 mg were analyzed. This extensive range of sample sizes was chosen to thoroughly evaluate the oxidation process in the EA and the results of the IRMS. Each set of reference material underwent at least two measurement sequences, and no reactor changes were made throughout the analysis.

Figure 2 i) presents the results of the measured δ13C values for the reference materials IAEA-C3, IAEA-C5, and IAEA-C7, calibrated using IAEA-C6 and IAEA-C8 as reference material. Notably, it was observed that samples with masses exceeding 1.300 mg led to current saturation in the IRMS Faraday cups. Consequently, the application is limited to samples weighing up to 1.300 mg.

Figure 2 i), the δ13C values of IAEA-C3, IAEA-C5, and IAEA-C7 are presented as initially calculated. These data points exhibit a mass dependency, which is addressed by applying a linear fit (shown as a dotted line) for correction. In Figure 2 ii), the corrected data is displayed. The data is calibrated to the δ13CPDB scale, utilizing measurements from IAEA-C6 and IAEA-C8. Figures iii) and iv) show the same plot but with reference materials IAEA-C2, IAEA-C6 and IAEA-C8, which are calibrated with IAEA-C3 and IAEA-C7.

The results, depicted in Figure 2, indicate a mass dependency in the δ13C values of all three reference materials. To address this issue, a linear fit was applied to the data, resulting in the following fit equations: IAEA-C3: yc3(x) = 0.4298 • x – 0.1232, IAEA-C5: yc5 (x) = 0.4020 • x + 0.0720, and IAEA-C7: yc7(x) = 0.1247 • x - 0.3494. To correct for this mass dependency, the δ13C values are transformed in a way that after transformation the linear fit of the data results in a slope of zero. The corrected values are illustrated in Figure 2 ii). Table 1 shows the standard deviations of the corrected δ13C values.

This mass dependency was also seen for a similar set of measurements of three additional reference materials. The mass dependency in Figure 2 could be interpreted as a dependency to the carbon content, with a lower δ13C value the slope becomes less steep. However, this effect could not be confirmed with the additionally measured reference materials which can be seen in Figure 2 iii) and iv).

Comparing the δ13C values of the IAEA reference materials with those measured at Cologne, it is found that they fall within the expected range, with statistical uncertainties below 0.2‰. This demonstrates a good and reproducible agreement with the IAEA information values. Additionally, it is worth mentioning that the careful selection of calibration reference materials is crucial. In this study, two reference materials were chosen, to provide a large δ13C calibration range. As can be seen in Figure 2 this can lead to deviations from the information values. Therefore, for a more precise calibration technique, Coplen et al. (Reference Coplen, Brand, Gehre, Groening, Meijer, Toman and Verkouteren2006) can be utilized.

Samples from Spitzbergen, Norway

In addition, environmental samples COL5115, COL5111, and COL5119 collected from Spitzbergen, Norway, were analyzed for their δ13C values. These samples had previously been measured at the Institute of Soil Science, University of Hamburg, and are labeled as δ13CHH. Prior to analysis, the environmental samples underwent decalcification using phosphoric acid, following the method described by Rethemeyer et al. (Reference Rethemeyer, Gierga, Heinze, Stolz, Wotte, Wischhoeofer, Berg, Melchert and Dewald2019). The δ13C values were measured using a Delta V instrument (Thermo Scientific) (Knoblauch et al. Reference Knoblauch, Beer, Sosnin, Wagner and Pfeiffer2013). Each environmental sample was measured twice by the Institute of Soil Science and the given uncertainties is the standard deviation between those two points. Detailed δ13C values and their corresponding uncertainties are provided in Table 2.

Table 2 The measured δ13C values and carbon content of the environmental samples from Spitzbergen, along with their corresponding standard deviations, are presented. The δ13CHH values represent the measurements conducted in Hamburg, while the data obtained in this study are provided as δ13CCologne.

A total of 81 samples were analyzed, ranging in sample masses from 0.060 to 1.300 mg. The samples underwent oxidation in the EA using a helium flow rate of 250 ml/min, and their δ13C values were determined in the IRMS using IAEA-C3 and IAEA-C6 for calibration. It is worth noting that the δ13C values of the COL samples are similar to each other, indicating that the calibration procedure using only two reference materials was sufficient for all three samples. However, it is important to exercise caution when calibrating the IRMS system for samples with significantly different δ13C values to ensure accurate results within each range.

Furthermore, the carbon contents of the samples were determined using the EA and are listed in Table 2. Unlike the reference materials, only the environmental sample COL5115 exhibited a noticeable mass dependency, which was subsequently corrected for. This correction is illustrated in Figure 3.

Figure 3 The δ13C values of three environmental samples collected from Spitzbergen and measured at CologneAMS are presented in blue with their average value. These values are compared to measurements from Hamburg, represented by the red line. Furthermore, the δ13C values have been corrected for their mass dependency, and the corrected data is depicted in black.

The δ13C values of the COL samples have a larger standard deviation of up to 0.77‰ compared to those of the standard material with 0.18‰ for IAEA-C7. The reason for the larger scattering can on the one hand be the characteristic of soil samples. On the other hand the standard deviations of the samples in this work result in a statistically significant value, due to the large number of samples, compared to the two measurements per sample done in Hamburg.

Additionally, it is important to acknowledge that the mass dependency effect is not yet fully understood and requires further investigation. To minimize potential uncertainties associated with the mass dependency, future measurements will employ samples and reference materials with similar δ13C values and masses.

δ15N values for Atacama Samples

Environmental samples from the Atacama desert in Chile were analyzed for their δ15N values and compared to measurements conducted at the University of Bonn using an already established EA-IRMS system. The system in Bonn utilizes a Thermo Fisher Scientific Flash EA 1112 series, which is connected to a Conflo III interface and a Delta V Advantage IRMS.

For the analysis, the samples were weighed at approximately 200 mg in both institutes and measured alongside IAEA-N1 (δ15N = +0.43‰) and IAEA-N2 (δ15N = +20.41‰) reference materials to calibrate the δ15N values. The environmental samples exhibit very low nitrogen content, averaging at 0.03 percent. Due to the low nitrogen content, larger sample sizes were required compared to standard carbon samples.

The first test series in Cologne was conducted on three different days of two weeks, during which the EA reactors remained unchanged. The samples were measured in ascending order based on their names. Each data point represents the average of approximately three samples of the same sample material. A more detailed listing can be found in Table 3. Figure 4 i) presents a comparison of the δ15N values obtained in Cologne and Bonn. The data indicates a discrepancy towards lower δ15N values for the samples measured last (Loa10, Loa11, Loa12). Further investigation revealed that these samples were only partially oxidized, likely accounting for the observed differences in δ15N values.

Table 3 δ15N along with their corresponding standard deviations of environmental samples from the Atacama desert measured with the EA-IRMS system at CologneAMS.

Figure 4 The two test series of the Atacama environmental samples are shown. The first test series, i) were measured in ascending order of the samples names. The second test series, ii), the samples were measured in descending order and with an extended O2 window during the combustion in the EA. This is done to ensure the full combustion of the samples. The errorbars show the standard deviation of the individual set of samples. If the data point does not have an errobar, only one sample was measured. The identity line is depicted as a dotted line in the figures, serving as a reference.

To test this hypothesis, a second series was conducted in reverse order, starting with the Loa12 sample and proceeding in descending order of the sample names. This second test series is depicted in Figure 4 ii). The samples were measured on two different days without changing the reactors, and the O2 window for combustion was extended from 80 s to 100 s. Due to limited sample material, most of the samples in the second test series could only be measured once or twice.

The δ15N values in the second series exhibit greater scattering compared to the first series, likely due to the low number of measured samples and the heterogeneous nature of the soils. Additionally, it can be observed that the scattering no longer depends on the order in which the samples were measured, indicating that the deviation is now independent of sample order.

Excluding the sample Loa10, Loa11 and Loa12, the results of the first test series demonstrate that this setup is capable of reliably measuring δ15N values from environmental samples, and the results are comparable to those from other institutes. However, results for Loa10–Loa12 emphasize on a careful execution of the combustion, to ensure that unknown or large sample are fully oxidized.

The standard deviations of the sample material, in comparison to the reference materials, show an increase. While the standard deviation of the IAEA-N2 reference materials is 0.08‰, the samples exhibit standard deviations of up to 2.5‰. This difference may be attributed to the non-homogeneous nature of the samples and warrants further investigation with a larger sample size. The effect could also stem from an incomplete reduction of NOx to N2, but the chromatograms of m/z 28 did not show any abnormalities. Overall, Figure. 4 i) illustrates that the δ15N values obtained from the new setup largely exhibit good consistency with the results from Bonn University.

Connection

In this section the two setups shown in Figure 1 are tested with standard material.

GIS to IRMS

The connection between the GIS and the IRMS was assessed by introducing varying masses of Ox-II gas into the system. The Ox-II gas is a mixture consisting of 5 percent Ox-II and the remaining portion being helium. The carbon mass was determined by measuring the pressure within a known volume of syringe, tubes, and fittings (Stolz et al. Reference Stolz2020). The bottled Ox-II gas was directly transferred to the syringe of the GIS. Target carbon sample masses of 0.495(3) μg, 0.745(1) μg, 0.981(17) μg, 1.896(7) μg, 2.888(9) μg, 3.922(9) μg, 4.628(7) μg, 5.942(5) μg, 6.610(202) μg, and 7.714(2) μg were transferred from the GIS syringe to the IRMS. Four measurements were conducted for each mass, and the average δ13C value was plotted against the mass in Figure 5. The δ13C measurements were calibrated using the 7.71 μg Ox-II sample.

Figure 5 The graphic shows the test series for the connected GIS-IRMS system. Different sample sizes of Ox-II are transferred into the IRMS and its δ13C value is measured. The data is calibrated using the 8 μg sample. The errorbars show the standard deviation of the set of one sample mass.

The test of the connected GIS and IRMS system, shown in Figure 5, yielded reliable δ13C values, except for carbon masses below 2 μg, where the δ13C values exhibited scattering. While the δ13C value for 6.610 μg could stem from the larger mass scattering of ±0.202 μg the δ13C value for 4.628 μg cannot be explained by this. Nonetheless the δ13C value is still within the range of the information value.

The data points above 2 μg exhibit a standard deviation of the δ13C values ranging from 0.03 to 0.06‰, except for the 4.63 μg and 6.61 μg samples which have standard deviations of 0.19‰ and 0.16‰, respectively.

For sample masses below 2 μg, deviations from the information value become more significant, potentially due to constant contamination. In this setup, the deviation from the information value can reach up to ±0.6‰. Although the standard deviation for the 0.49 μg and 0.75 μg samples is ±0.2‰, this would still result in reliable δ13C measurements, if the calibration gas volume is kept the same as the sample gas, but it could result in larger uncertainties.

Further testing of this setup is necessary to determine if the deviations from the information value are caused by constant contamination. The data collected thus far is not sufficient to draw definitive conclusions.

Additionally, as depicted in the Chapter IRMS Standard Samples, the δ13C values of solid samples analyzed using the EA-IRMS display a dependency on sample mass. In contrast, the gaseous samples discussed in this chapter do not exhibit such a dependency. It is important to highlight that the sample masses measured in both setups are comparable, as indicated by the similar peak areas measured with the IRMS. This observation suggests that this effect could stem from either the sample or the EA system itself. Further investigation needs to be conducted to find the cause of the mass dependency.

EA-IRMS to GIS

This section presents the first results of the combustion of samples in the EA and the quasi-simultaneous measurement of δ13C with the IRMS and AMS.

In the first test measurement, 32 IAEA-C6 samples with a δ13C value of –10.8‰ were measured using the setup shown in Figure 1 i). Alongside, eight Ox-II samples and two blank samples were also measured for calibration. All samples had a carbon mass of around 100 μg.

The IAEA-C6 δ13C obtained from the AMS measurements was corrected using the blank and Ox-II samples, while the IAEA-C6 data from the IRMS measurements was calibrated and corrected using the data from the Ox-II samples and two empty tin boats that were measured as well. The evaluated data is presented in Figure 6.

Figure 6 Shown are the δ13C values of IAEA-C6 which has an information value of δ13C PDB= –10.80(47)‰. The blue data represents the values measured by the IRMS, while the green is the data measured by the AMS. The dotted lines show the average values respectively and the filled in area is the uncertainties of the average. The error bars show the standard deviation of the complete set of either AMS or IRMS measured samples.

Figure 6 shows that the AMS measurements of the δ13C have a standard deviation seven times larger than the IRMS data. Additionally, the δ13C values obtained from the AMS measurements fall within the standard deviation of the IAEA-C6 measured using IRMS. This is similar to McIntyre et al. (Reference McIntyre, Wacker, Haghipour, Blattmann, Fahrni, Usman, Eglinton and Synal2016), where the IRMS δ13C value lies within the given standard deviation.

Reasons for the more precise δ13C could lie within the different ion sources that are used in the AMS and IRMS. But the results presented in this work lie close to another and could give rise to the possibility to use the δ13C from the IRMS for the fractionation correction. In the future further test will be conducted to find if the δ13C from the AMS could have a systematical offset compared to the IRMS δ13C data.

CONCLUSION

The 6 MV AMS system at the University of Cologne, which is used for routine 14CO2 measurements, has been expanded with the integration of an EA-IRMS system, to enable the quasi-simultaneous measurement of δ13C. The entire system is now computer-controlled using a single software.

While the mass dependency found in Chapter IRMS Standard Samples needs to be further investigated to find its origin, standard materials and environmental samples consistently yield reproducible δ13C values that are comparable to the results reported by Knoblauch et al. (Reference Knoblauch, Beer, Sosnin, Wagner and Pfeiffer2013). Additionally, a good agreement was observed between the δ15N values of Atacama environmental samples measured using the new EA-IRMS system and those obtained from the established EA-IRMS system at Bonn University.

Furthermore, our tests demonstrate that the new EA-IRMS-GIS system allows for the transfer of gaseous samples from the GIS to the IRMS. While this setup is still in progress and requires further testing, the initial measurements of δ13C values have been reliably obtained with sample masses as low as 2 μg. However, the measurement of δ13C of samples below 2 μg need to be treated with caution.

As an initial step, the δ13C of IAEA-C6 standard material was successfully measured using both the IRMS and AMS. This system enables precise measurements of a small amount of sample gas for δ13C values while the rest is transferred to the AMS system. Eventually, this development could enable the use of IRMS δ13C values to correct for fractionation in AMS measurements.

ACKNOWLEDGMENTS

The research is supported by the CRC1211. We thank Christian Knoblauch and Ramona Mörchen for their supply of the EA-IRMS data.

Footnotes

Selected Papers from the 24th Radiocarbon and 10th Radiocarbon & Archaeology International Conferences, Zurich, Switzerland, 11–16 Sept. 2022

References

REFERENCES

Alderliesten, C, Van Der Borg, K, De Jong, AFM. 1998. Contamination and fractionation effects in AMS-measured 14C/12C and 13C/12C ratios of small samples. Radiocarbon 40(1), 215221. doi: 10.1017/S0033822200018075 CrossRefGoogle Scholar
Arnold, M, Bard, E, Maurice, P, Valladas, H, Duplessy, JC. 1989. 14C dating with the Gif-sur-Yvette Tandetron accelerator: status report and study of isotopic fractionation in the sputter ion source. Radiocarbon 31(3), 284291. doi: 10.1017/s0033822200011814 CrossRefGoogle Scholar
Coplen, TB, Brand, WA, Gehre, M, Groening, M, Meijer, HAJ, Toman, B, Verkouteren, RM. 2006. New Guidelines for δ13C Measurements. Analytical Chemistry 78(7), 24392441. doi: 10.1021/ac052027c CrossRefGoogle ScholarPubMed
Hofmann, H, Bonani, G, Morenzoni, E, Nessi, M, Suter, M, Woelfli, W. 1984. Charge state distributions and resulting isotopic fractionation effects of carbon and chlorine in the 1–7 MeV energy range. Nuclear instruments and Methods in Physics Research B: Beam Interactions with Materials and Atoms 5(2), 254258. doi: 10.1016/0168-583X(84)90522-6 CrossRefGoogle Scholar
IAEA (International Atomic Energy Agency Department of Nuclear Sciences and Applications IAEA Environment Laboratories). 2014. Reference sheet for quality control material. https://nucleus.iaea.org/sites/ReferenceMaterials/Shared Google Scholar
Knoblauch, C, Beer, C, Sosnin, A, Wagner, D, Pfeiffer, EM. 2013. Predicting long-term carbon mineralization and trace gas production from thawing permafrost of northeast Siberia. Glob. Change Biology 19, 11601172. doi: 10.1111/gcb.12116.CrossRefGoogle ScholarPubMed
McIntyre, CP, Wacker, L, Haghipour, N, Blattmann, TM, Fahrni, S, Usman, M, Eglinton, TI, Synal, HA. 2016. Online 13C and 14C gas measurements by EA-IRMS-AMS at ETH Zuerich. Radiocarbon 59(3), 893903. doi: 10.1017/rdc.2016.68 CrossRefGoogle Scholar
Melchert, JO, Stolz, A, Dewald, A, Gierga, M, Wischhoefer, P, Rethemeyer, J. 2018. Exploring sample size limits of AMS gas ion source 14C analysis at CologneAMS. Radiocarbon 61(6), 17851793. doi: 10.1017/RDC.2019.143 CrossRefGoogle Scholar
Prasad, GVR, Culp, R, Cherkinsky, A. 2019. 13C correction to AMS data: Values derived from AMS vs IRMS values. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms 455, 244249. doi: 10.1016/j.nimb.2019.01.034 CrossRefGoogle Scholar
Rethemeyer, J, Gierga, M, Heinze, S, Stolz, A, Wotte, A, Wischhoeofer, P, Berg, S, Melchert, JO, Dewald, A. 2019. Current sample preparation and analytical capabilities of the radiocarbon laboratory at CologneAMS. Radiocarbon 61(5), 14491460. doi: 10.1017/RDC.2019.16 CrossRefGoogle Scholar
Rozanski, K, Stichler, W, Gonfiantini, R, Scott, EM, Beukens, RP, Kromer, B, van de Plicht, J. 1992. The IAEA 14C Intercomparison Exercise 1990. Radiocarbon 34(3), 506519. doi: 10.1017/S0033822200063761 CrossRefGoogle Scholar
Stolz, A. 2020. Einrichtung und Weiterentwicklung eines 14CO2 Systems am 6 MV TANDETRON Beschleuniger des CologneAMS [dissertation].Google Scholar
Stolz, A, Dewald, A, Altenkirch, R, Herb, S, Heinze, S, Schiffer, M, Feuerstein, C, Mueller-Gatermann, C, Wotte, A, Rethemeyer, J, Dunai, TJ. 2017. Radiocarbon measurements of small gaseous samples at CologneAMS. Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms 406(A), 283286. doi: 10.1016/j.nimb.2017.03.031 CrossRefGoogle Scholar
Stolz, A, Dewald, A, Heinze, S, Altenkirch, R, Hackenberg, G, Herb, S, Mueller-Gatermann, C, Schiffer, M, Zitzer, G, Wotte, A, Rethemeyer, J, Dunai, TJ. 2019. Improvements in the measurement of small 14CO2 samples at CologneAMS. Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms 439, 7075. doi: 10.1016/j.nimb.2018.12.008 CrossRefGoogle Scholar
Wacker, L, Fahrni, SM, Hajdas, I, Molnar, M, Synal, HA, Szidat, S, Zhang, YL. 2013. A versatile gas interface for routine radiocarbon analysis with a gas ion source. Nuclear Instruments and Methods in Physics Research B: Beam Interactions with Materials and Atoms 294, 315319. doi: 10.1016/j.nimb.2012.02.009 CrossRefGoogle Scholar
Figure 0

Figure 1 The schematic diagram illustrates the connection between the EA, IRMS, GIS, and AMS systems. In i), the connection is established for a sample which is oxidized in the EA. Subsequently, the gaseous sample is transferred through a capillary to a T-piece. Around 10 percent of the sample is directed towards the IRMS through a capillary (cyan) for measurement, while the remaining portion is further transferred to the GIS for subsequent analysis in the AMS ion source. Figure 1 ii) depicts the setup for pre-existing gaseous samples that can be cracked within the ampoule cracker of the GIS. The cracked sample is then transferred to the IRMS under a continuous flow of helium. This is accomplished by splitting an existing capillary, which supplies helium to the GIS, using a T-piece and an additional capillary (red).

Figure 1

Table 1 The measured δ13C values, along with their corresponding standard deviations, are compared to the information value values provided by the IAEA. The IAEA information values are denoted as δ13CPDB, representing their expression relative to Pee Dee belemnite (PDB). The data obtained in this work is presented as δ13CCologne.

Figure 2

Figure 2 i), the δ13C values of IAEA-C3, IAEA-C5, and IAEA-C7 are presented as initially calculated. These data points exhibit a mass dependency, which is addressed by applying a linear fit (shown as a dotted line) for correction. In Figure 2 ii), the corrected data is displayed. The data is calibrated to the δ13CPDB scale, utilizing measurements from IAEA-C6 and IAEA-C8. Figures iii) and iv) show the same plot but with reference materials IAEA-C2, IAEA-C6 and IAEA-C8, which are calibrated with IAEA-C3 and IAEA-C7.

Figure 3

Table 2 The measured δ13C values and carbon content of the environmental samples from Spitzbergen, along with their corresponding standard deviations, are presented. The δ13CHH values represent the measurements conducted in Hamburg, while the data obtained in this study are provided as δ13CCologne.

Figure 4

Figure 3 The δ13C values of three environmental samples collected from Spitzbergen and measured at CologneAMS are presented in blue with their average value. These values are compared to measurements from Hamburg, represented by the red line. Furthermore, the δ13C values have been corrected for their mass dependency, and the corrected data is depicted in black.

Figure 5

Table 3 δ15N along with their corresponding standard deviations of environmental samples from the Atacama desert measured with the EA-IRMS system at CologneAMS.

Figure 6

Figure 4 The two test series of the Atacama environmental samples are shown. The first test series, i) were measured in ascending order of the samples names. The second test series, ii), the samples were measured in descending order and with an extended O2 window during the combustion in the EA. This is done to ensure the full combustion of the samples. The errorbars show the standard deviation of the individual set of samples. If the data point does not have an errobar, only one sample was measured. The identity line is depicted as a dotted line in the figures, serving as a reference.

Figure 7

Figure 5 The graphic shows the test series for the connected GIS-IRMS system. Different sample sizes of Ox-II are transferred into the IRMS and its δ13C value is measured. The data is calibrated using the 8 μg sample. The errorbars show the standard deviation of the set of one sample mass.

Figure 8

Figure 6 Shown are the δ13C values of IAEA-C6 which has an information value of δ13C PDB= –10.80(47)‰. The blue data represents the values measured by the IRMS, while the green is the data measured by the AMS. The dotted lines show the average values respectively and the filled in area is the uncertainties of the average. The error bars show the standard deviation of the complete set of either AMS or IRMS measured samples.