Immediate-use steam sterilization and the effect on surgical site infection risk in an acute care facility

Published online by Cambridge University Press:  20 September 2024

Victoria D. Crall
Affiliation:
Department of Infection Control and Hospital Epidemiology, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Casey L. Lewis
Affiliation:
Department of Infection Control and Hospital Epidemiology, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Avisha Risnear
Affiliation:
Department of Surgery, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Jason Miller
Affiliation:
Department of Surgery, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Joanne Sherer
Affiliation:
Department of Surgery, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Ashley M. Ayres
Affiliation:
Department of Infection Control and Hospital Epidemiology, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA
Graham M. Snyder*
Affiliation:
Department of Infection Control and Hospital Epidemiology, UPMC Presbyterian-Shadyside, Pittsburgh, PA, USA Division of Infectious Diseases, Department of Medicine, University of Pittsburgh School of Medicine, Pittsburgh, PA, USA
*
Corresponding author: Graham M. Snyder; Email: [email protected]

Abstract

Immediate-use steam sterilization (IUSS) shortens the time of sterilization but may increase the risk of surgical site infection (SSI). Among 23,919 procedures with 416 (1.7%) procedures resulting in SSI, IUSS was associated with a 1.52 (95% CI, 1.10–2.11) times higher risk of SSI. IUSS should be minimized.

Type
Concise Communication
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - SA
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike licence (http://creativecommons.org/licenses/by-nc-sa/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the same Creative Commons licence is used to distribute the re-used or adapted article and the original article is properly cited. The written permission of Cambridge University Press must be obtained prior to any commercial use.
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America

Introduction

Compared to conventional sterilization techniques, immediate-use steam sterilization (IUSS) shortens the time from sterilization to the aseptic transfer of a medical device onto the surgical sterile field. IUSS should be utilized following the manufacturer’s instructions for use and within evidence-based guidelines, but not for convenience purposes. 1 In the operating room, there are competing pressures such as efficient turnover time, on-time start of surgery, and case cart preparation that factor into the utilization of IUSS and potential surgical site infection (SSI) risk resulting from a non-sterile device. Reference Nania2,Reference Seavey3 Few published data have quantified the risk of SSI following IUSS, compared to the standard sterile reprocessing method. Reference Tantillo, Stapleton and Frane4Reference Leonard, Speroni, Atherton and Corriher6 We aimed to measure the association between IUSS use for surgical instrument reprocessing and SSI risk in a facility where IUSS use increased due to staffing constraints and case volumes.

Methods

This retrospective observational study took place, between January 2022 and December 2023, at a tertiary care hospital with a diverse mix of surgery types. The hospital comprises 2 buildings with 2 separate operating suites and 2 sterile processing departments. Outside of the sterile processing departments, each operation suite houses autoclaves available for prion reprocessing and IUSS within the above evidence-based guidelines 1 . Each autoclave is tested daily to ensure the vacuum system is functioning appropriately, and each IUSS cycle requires documentation of indication and a chemical indicator for validation. Within the sterile processing department, steam and low-temperature gas sterilization methods are used, according to the manufacturer’s instructions for use, for standard reprocessing. For each standardly reprocessed load, a biological and chemical indicator is used to validate the sterilization process. In addition, each tray has exterior sterilization indicator tape and an internal chemical indicator. Regardless of the sterilization method (standard vs IUSS), all instruments go through a manual and/or mechanical decontamination and washing per hospital policy within the sterile reprocessing department.

We utilized internal databases of patient-linked device reprocessing, including sterile reprocessing logs, to document the exposure as “standard reprocessing” or IUSS. SSI outcomes and surgery types were defined using National Health and Safety Network (NHSN) definitions; surgery types were categorized by the surgical service. 7 We compared SSI rates among surgeries using one or more surgical devices preoperatively sterilized using IUSS compared to standard reprocessing methods and calculated a risk ratio (RR) and 95% confidence interval (95% CI) for all NHSN-defined surgery types and stratified by 7 surgical services that had ≥1 reported SSIs in the study time frame. Indication for IUSS was summarized across NHSN and non-NHSN procedures during the study period. Analyses were performed using STATA (version 12.1). This investigation was approved as a quality improvement intervention by the UPMC Quality Review Committee.

Results

Among 23,919 surgical procedures, 416 (1.74%) developed SSIs, including 2.56% (39 of 1,524) and 1.68% (376 of 22,395) of procedures for which IUSS and standard reprocessing was used to sterilize instruments prior to the procedure, respectively (Table 1). IUSS was associated with a 1.52 times higher risk of SSI (95% CI, 1.10–2.11). Two surgical services had statistically significant RRs for SSI development after IUSS: transplant surgery (RR 2.47; 95% CI, 1.32–4.60) and plastic surgery (RR 3.64; 95% CI, 1.13–11.74) (Figure 1, Supplemental Table).

Table 1. Frequency of surgical site infection, by preceding reprocessing method

Figure 1. Risk of national healthcare safety network-defined surgical site infection among procedures with and without preceding immediate-use steam sterilization, by surgery type.

The reasons for IUSS rather than standard reprocessing among the 2,069 IUSS instances were insufficient reprocessing time before surgery started (1,284, 62.1%), missing instruments in set identified during surgery (646, 31.2%), crucial device contaminated in the operating room (112, 5.4%), incoming vendor trays (24, 1.2%), and implants (3, 0.1%).

Discussion

In this observational study of SSI risk following NHSN-defined procedures for which surgical devices were reprocessed prior to the procedure with either IUSS or standard reprocessing, we observed a 1.52 times higher overall risk of SSI following IUSS and statistically significantly higher risk for transplant and plastic surgery procedures. These findings support the avoidance of IUSS to reduce the risk of SSI.

The most robust study to date defining the SSI risk associated with IUSS identified no significant difference in SSI risk among 111 propensity-matched patients (61 IUSS, 50 non-IUSS reprocessing) undergoing orthopedic or spine surgeries. Reference Tantillo, Stapleton and Frane4 Reasons for findings discordant with this investigation may relate to the procedure types studied: the familiarity, acceptance, indication, and workflow disruption associated with IUSS may vary between procedures, making the strength of causal or incidental association with SSI differ between surgery types. For example, a surgical procedure for which IUSS is commonly used and part of the perioperative workflow may not incur significant SSI risk compared to a procedure where IUSS represents an interruption of usual process and urgency that may lead to other lapses in SSI prevention practices (eg, operating room entries).

Our study is limited by a lack of modeling of patient and procedure-associated factors associated with SSI risk. Reference Zuckerman, Parikh, Moore and Talbot8 Due to the nature of documentation of autoclave device indication and IUSS use prior to a surgical case, we could not report the SSI rate stratified by IUSS indication. During the course of this investigation, there were no new interventions to modify these factors, reducing the likelihood that these differed in the IUSS versus standard reprocessing groups except by random chance.

Ethical reasons prohibit a randomized trial of IUSS versus standard reprocessing. Larger studies in diverse patients, procedures, and facility types with quasi-experimental study designs are needed to provide a clearer estimation of SSI risk associated with IUSS.

Given our findings and regulatory guidance, 1 IUSS utilization should be minimized. The reasons for IUSS use may be complex. An in-depth root cause analysis—as our team performed in this investigation—can reveal issues broader than and indirectly related to surgical device reprocessing, such as staffing levels, supply chain operations, surgical staff education, patient throughput, and organizational culture. A regular review of IUSS use rates with root cause analyses may help identify underlying factors related to quality of care.

In this observational study of SSI risk among patients undergoing surgical procedures with surgical equipment reprocessed prior to the procedure using IUSS or standard methods, we identified a 1.52 times higher risk of SSI following IUSS device reprocessing and significantly increased procedure-specific risk for transplant and plastic surgeries. Although the analysis does not precisely identify why IUSS confers the risk of SSI, our findings add the most robust information to a very limited amount of published data supporting recommendations to minimize IUSS use.

Supplementary material

To view supplementary material for this article, please visit https://doi.org/10.1017/ash.2024.408.

Financial support

None reported.

Competing interests

All authors report no conflicts of interest relevant to this article.

Footnotes

Findings from this study were presented at the SHEA Spring 2024 conference.

References

The Joint Commission. Instrument Reprocessing - Immediate Use Steam Sterilization (IUSS): What are important considerations associated with Immediate-Use Steam Sterilization?, 5/12/2017. https://www.jointcommission.org/standards/standard-faqs/ambulatory/infection-prevention-and-control-ic/000002122/. Accessed May 16, 2024.Google Scholar
Nania, P. Immediate use steam sterilization: it’s all about the process. AORN J 2013;98:3238.Google Scholar
Seavey, R. Immediate use steam sterilization: moving beyond current policy. Am J Infect Control 2013;41:S46S48.Google Scholar
Tantillo, TJ, Stapleton, EJ, Frane, N, et al. The association of immediate-use steam sterilization with the incidence of orthopaedic surgical site infections: a propensity score-matched cohort study. J Bone Joint Surg Am 2022;104:988994.Google Scholar
Hutzler, L, Kraemer, K, Iaboni, L, Berger, N, Bosco, JA. A hospital-wide initiative to eliminate preventable causes of immediate use steam sterilization. AORN J 2013;98:597607.Google Scholar
Leonard, Y, Speroni, KG, Atherton, M, Corriher, J. Evaluating use of flash sterilization in the OR with regard to postoperative infections. AORN J 2006;83:672680.Google Scholar
Centers for Disease Control and Prevention. Surgical Site Infection Event (SSI), 2024. https://www.cdc.gov/nhsn/pdfs/pscmanual/9pscssicurrent.pdf. Accessed May 16, 2024.Google Scholar
Zuckerman, SL, Parikh, R, Moore, DC, Talbot, TR. An evaluation of immediate-use steam sterilization practices in adult knee and hip arthroplasty procedures. Am J Infect Control 2012;40:866871.Google Scholar
Figure 0

Table 1. Frequency of surgical site infection, by preceding reprocessing method

Figure 1

Figure 1. Risk of national healthcare safety network-defined surgical site infection among procedures with and without preceding immediate-use steam sterilization, by surgery type.