Hostname: page-component-cd9895bd7-8ctnn Total loading time: 0 Render date: 2024-12-23T07:47:53.332Z Has data issue: false hasContentIssue false

Evaluation of intracranial stereotactic treatment plans: a comparison study of CyberKnife and TrueBeam systems

Published online by Cambridge University Press:  22 April 2024

Muhammad Shahban*
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
Muhammad Tariq Rizwan
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
Neelam Shahzadi Nazia
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
Ur Rehman Saeed
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
Mehmood Humera
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
Muhammad Faheem
Affiliation:
AECH NORI, Hanna Road, G8/3, Islamabad, Pakistan
*
Corresponding author: Muhammad Shahban; Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Background:

Stereotactic radiotherapy (SRT) for patients with intracranial tumours are delivered using a dedicated platform or a conventional linear accelerator with a flattening filter-free beam.

Materials and methods:

This study compares treatment plans with intracranial tumours. A total of 29 patients were treated on CyberKnife and planned using the Accuray Precision. The same structure sets ws then exported to Varian Eclipse, and plans were made using a 6 MV FFF beam. Both plans were compared for parameters of target coverage, homogeneity index (HI), new conformity index (nCI), gradient index, selectivity index (SI), volumetric and OAR doses.

Results:

The treatment plans made for CyberKnife exhibit better results in terms of nCI (1·168 ± 0·08 versus 1·173 ± 0·077), SI (0·885 ± 0·05 versus 0·877 ± 0·05) and GI (3·64 ± 0·5 versus 4·45 ± 1·25), while HI values are better for TrueBeam. For OAR doses, in 65·5% and 72% of treatment plans, brainstem and optic pathways received lower doses on CyberKnife, respectively. In terms of dose spillage, Truebeam plans are better for very low doses (V5%), while for V10%, V20% and V50% CyberKnife plans are better.

Conclusion:

CyberKnife is a better modality for the delivery of SRS/SRT to intracranial tumours except for dose homogeneity where TrueBeam offered better results.

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - ND
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives licence (http://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided that no alterations are made and the original article is properly cited. The written permission of Cambridge University Press must be obtained prior to any commercial use and/or adaptation of the article.
Copyright
© Nuclear Medicine Oncology and Radiotherapy Institute, NORI Islamabad, 2024. Published by Cambridge University Press

Introduction

The efficacy and usefulness of stereotactic radiosurgery and stereotactic radiation therapy (SRS/SRT) over conventional radiotherapy have been well-established by many investigators. Reference Antypas and Pantelis1Reference Stafinski, Jhangri, Yan and Menon3 Modern radiotherapy techniques are based on precise target localisation and implementation of multiple coplanar or non-coplanar beams to deliver a highly precise and conformal radiation dose to the tumour while administering the minimal dose to normal tissues. Consequently, the dose per fraction to target the tumour may be increased to a very high level as compared with the conventional fractionation, exempting the need to give time for normal tissue healing. This paves the way to SRS/SRT. Reference Tran, Zhang and Woods4,Reference Zhang, Chiu and Dubas5

The debate on the selection of a suitable platform for the delivery of SRS treatment has been there since the advent of the technique. Currently, dedicated SRS platforms CyberKnife and Gamma Knife are commercially available, while linear accelerators with flattening filter-free (FFF) beams also offer optimal results. Reference Blamek, Grządziel and Miszczyk6 Both platforms have their advantages and disadvantages for treatment delivery. Their comparisons for different sites have been reported in the literature with both phantom and retrospective patient data. Reference Benedict, Perks and Goetsch7Reference Eaton, Lee and Patel13 A few of the research articles presented a comparison of intracranial treatment plans using different modalities. The present study is focused on the comparison of intracranial treatment plans made to be delivered on CyberKnife and TrueBeam.

Materials and Methods

In this study, CyberKnife, M6 Model (Accuray, Sunnyvale, CA, USA) and TrueBeam (Varian Medical Systems, CA, USA) have been used as SRS treatment modalities. The treatment planning was performed on the Accuray Precision TPS (version 3·3·1) using the Ray Tracing dose calculation algorithm for CyberKnife and on the Varian Eclipse TPS using the Acuros XB algorithm for TrueBeam. The dose rates of CyberKnife and TrueBeam are 950 MU/min and 1000 MU/min, respectively. A 6 MV FFF beam was used for treatment plans of both modalities.

Twenty-nine patients with intracranial tumours of miscellaneous sizes and types were treated on CyberKnife during the 8-month interval of this study. The types of tumour included in this study were meningioma, AVM, glioma, pituitary adenoma, schwannoma and brain mets. The target volume for each plan comprised of a gross tumour volume (GTV) and a planning tumour volume (PTV) where PTV was obtained by adding a 1 mm symmetric margin around GTV. The volume of the PTV ranged between 0·99 and 79·2 cc with an average of 16·53 cc. The organs at risk (OAR) contoured by the oncologists included mostly the normal brain (whole brain minus GTV), optic pathway (sum of all optic structures including the optic chiasm, optic nerves, eyes and lenses) and the brainstem.

For CyberKnife, most of the patients were planned using three or fewer collimators, whereas the number of non-coplanar beams ranged from 88 to 257. All the plans were verified for patient-specific quality assurance (PSQA) by measuring point doses using Sun Nuclear SNC 125c (active volume: 0·125 cc) and Exradin A14SL (active volume: 0·016cc) ion chambers. A commercially available stereo phantom (StereoPHAN) was used for point dose measurements.

For TrueBeam, the treatment plans were prepared with arc therapy using two full arcs of the volumetric arc therapy technique. The PSQA plans were created from the SRS/SRT plans and delivered on an electronic portal imaging device using portal dosimetry. All plans complied with 3%/3 mm criteria with a 95% confidence interval.

In both cases, the outcome for planning was set to be the coverage of PTV and GTV. The optimisation was repeated until minimum doses for OARs were achieved while keeping the target coverage optimised.

The treatment plans prepared for CyberKnife and TrueBeam were evaluated by a combined team comprising physicists and oncologists using the following treatment planning outcomes and indices.

Target coverage

Target coverage is defined as the volume receiving 100% of the prescribed dose. Reference Torrens, Chung and Chung14 Target coverage for both PTV and GTV was compared.

$${{Target}}\;{{coverage}} = {{{{PTV}}\_{{{V}}_{{{PD}}}}} \over {{{PTV}}\_{{V}}}}$$

Selectivity index (SI)

SI is defined as the quotient of the PTV volume and the body receiving the prescribed dose.

$${{SI}} = {{{{PTV}}\_{{{V}}_{{{PD}}}}} \over {{{Body}}\_{{{V}}_{{{PD}}}}}}$$

were PTV_VPD and Body_VPD are volumes of target and body, respectively, within the prescribed dose isodose line. The SI indicates how much of the normal tissue surrounding the target is being irradiated by the prescribed dose. The value 1 of the SI indicates the ideal situation of no irradiation of normal tissue beyond the target volume with the prescribed dose. Its value should be higher than 0·9 while greater than 0·75 for SRS/SRT is also acceptable. Reference Tunstall, Watson and Clarke15

New conformity index (nCI)

The conformity index (CI) is defined as the quotient of volume within the prescription isodose curve with the volume of PTV. These volumes may not overlap each other which leads to a deficit in the calculation of CI value. CI equal to 1 indicates that the volumes of the prescription isodose line and the PTV are equal but cannot identify the non-overlapping region resulting in missed coverage and spillage of prescribed isodose. Reference Torrens, Chung and Chung14,Reference Petkovska, Tolevska and Kraleva16,Reference Paddick17 This issue was addressed by defining the new conformity index (nCI) which indicates the degree of conformality with which the target is being conformed by the prescribed isodose curve. It is a tool to assess the conformity of delivered doses. Reference Al-Rawi, Abouelenein and Nagdy18 nCI equals to 1 indicates 100% PTV coverage with no spillage of isodose to the surrounding normal tissues. The inverse of nCI is called the Paddick conformity index (PCI). Reference Tunstall, Watson and Clarke15,Reference Paddick and Lippitz19

$${{PCI}} = {{PTV\;coverage}} \times {{SI}}$$
$${{nCI}} = {1 \over {{{PCI}}}}$$

Homogeneity index (HI)

Homogeneity index is defined as the quotient of the maximum and minimum dose received by the PTV. Reference Torrens, Chung and Chung14,Reference Ito, Kawamura and Mori20 HI equals to 1 signifies no hot spots or cold spots within or outside the target volume.

$${{HI}} = {{{{PTV}}\_{{{D}}_{{{max}}}}} \over {{{PTV}}\_{{{D}}_{{{min}}}}}}$$

Gradient index (GI)

GI is defined as the quotient of volumes of the body receiving 100–50% of the prescribed dose. The GI signifies the rate of decrease of dose outside the target volume. For SRS/SRT, its optimal value is less than or equal to 3, but it should not exceed 5. Reference Tunstall, Watson and Clarke15

$${{GI}} = {{{{Body}}\_{{{V}}_{{{PD}}}}} \over {{{Body}}\_{{{V}}_{{\rm 50\%} \;{{of}}\;{{PD}}}}}}$$

where ${\rm{Body}}\_{{\rm{V}}_{{\rm{PD}}}}$ is the volume of the body receiving the prescribed dose and ${\rm{Body}}\_{{\rm{V}}_{50{\rm{\% \;of\;PD}}}}$ is the volume of the body receiving 50% of the prescribed dose. Reference Paddick and Lippitz19

The comparison of CyberKnife and TrueBeam plans was made as the quotient of CyberKnife value and TrueBeam value. A value greater than unity signifies the higher value of the parameter for the TrueBeam plan and vice versa.

For comparison, the ratio of indices values for CyberKnife to TrueBeam (CK-TB) was calculated.

$${{{{CyberKnife}}\;\left( {{{CK}}} \right)} \over {{{TrueBeam}}\;\left( {{{TB}}} \right)}} = {{{{Index}}\;{{value}}\;{{for}}\;{{CyberKnife}}} \over {{{Index}}\;{{value}}\;{{for}}\;{{TrueBeam}}}}$$

Volumetric doses

Lower dose spillage was compared by evaluation of doses received by the body with 5, 10, 20, and 50% of the prescribed doses, denoted as V5%, V10%, V20% and V50%, respectively.

OAR dose constraints

Dose constraints for intracranial treatment plans including Dmax, D0·03cc and D10% of the brainstem, optic pathway (Boolean of all optic structures), spinal cord planning risk volume (PRV) (including the medulla) and V12Gy, V18Gy and V5Gy of the normal brain were evaluated and compared. PRVs for OARs were drawn by adding a 1 mm margin to the respective OARs. Reference Hanna, Murray and Patel21,Reference Diez, Hanna and Aitken22 Treatment planning goals for targets and dose constraints for OARs that were considered in this study are given in Table 1.

Table 1. Acceptability criteria for target volumes and OARs in case of intracranial SRS/SRT Reference Hanna, Murray and Patel21

OARs, organs at risk; SRS/SRT, stereotactic radiosurgery and stereotactic radiation therapy; GTV, gross tumour volume; PTV, planning tumour volume.

* GTV coverage was achieved 100 in 99% of the cases, while it was greater than 99.5 in 100% of cases.

Results and Discussion

A total of 29 SRS/SRT plans with 1, 3 and 5 fractions were evaluated. Out of these 28 patients, 8 were treated with single fraction, 9 patients were treated with 3 fractions, and 12 patients were treated with 5 fractions. All the plans were evaluated, and a comprehensive comparison was made, based on treatment planning parameters and evaluation indices. The results of these comparisons are given in Table 2 and discussed below.

Table 2. Comparison of evaluation indices in treatment plans of CyberKnife versus TrueBeam

PTV, planning tumour volume; GTV, gross tumour volume; SI, selectivity index; nCI, new conformity index; HI, homogeneity index; GI, gradient index.

Dosimetric parameters and evaluation indices

The mean value of the indices along with their range and standard deviation from all the treatment plans are stated. The CyberKnife plans recorded better values of GTV coverage, SI, GI and nCI in comparison to TrueBeam plans. On the contrary, the HI is relatively better achieved in a treatment plan for TrueBeam. The evaluated results for CyberKnife and TrueBeam are given in Table 2.

The ratio of the dosimetric parameters and evaluation indices values from the treatment plans for CyberKnife and TrueBeam are plotted against the patient number as shown in Figure 1. The graphs show that CyberKnife is superior to TrueBeam as nCI and GI are better for CyberKnife in 15 and 23 out of 29 treatment plans, respectively, while HI for TrueBeam is better in 27 treatment plans out of 29.

Figure 1. The graphs showing planning endpoints for both CyberKnife and TrueBeam plans. For each parameter, a point having a value above 1 signifies higher values for that parameter for the CyberKnife plan and vice versa.

Dose spillage

Although GI provides the measure of dose spillage beyond the PTV boundaries, it signifies the dose drop of 50% isodose line. Reference Paddick and Lippitz19 GI does not consider low doses which are absorbed in the surrounding normal tissues. For the assessment of low doses, values of V5%, V10%, V20% and V50% as a percentage of total body volume are evaluated. The results of the comparison are given in Table 3.

Table 3. Dose spillage in treatment plans of CyberKnife versus TrueBeam

where X, Y%, (Z1–Z2%), X = number of patients for which the index has a negative value, Y% = mean of difference between CyberKnife and TrueBeam results, and Z1–Z2% = range of difference.

Figure 2 shows a comparison of V5%, V10%, V20% and V50% doses as CK–TB values. For V5%, more treatment plans with greater than 0 value signify the higher value for CyberKnife treatment plans. For V10%, V20% and V50%, more points are below the zero line which shows that the values of these parameters are less in the CyberKnife treatment plans than in the TrueBeam.

Figure 2. Low doses received by the patient’s body from CyberKnife and TrueBeam plans as CK–TB values for the parameter. A negative value signifies the lower volume irradiated in CyberKnife plans.

Doses of OARs

The doses received by OAR are linked to their proximity to the PTV. In some cases, these OARs about the GTV and PTV are making it difficult to achieve the desired tumour coverage. The D0·035cc doses for brainstem and optic pathways were evaluated and compared as shown in Table 4.

Table 4. Doses of OARs in treatment plans of CyberKnife versus TrueBeam

where X, Y%, (Z1–Z2%), X = number of patients for which the dose to the OAR is lower, Y = mean of difference between CyberKnife and TrueBeam results with uncertainty, and Z1–Z2% = minimum to maximum doses.

OAR, organs at risk.

The doses received by the brainstem and optic pathway are linked to their proximity towards the target volume. In some of the treatment plans, these OARs are being abutted by the tumour, making it difficult to achieve the desired tumour coverage while sparing the normal tissues. The tolerance doses D0·035cc of the brainstem and optic pathways were evaluated in treatment plans prepared for CyberKnife and TrueBeam.

Figure 3 illustrates the status of individual patient doses for both OARs as the difference between CyberKnife and TrueBeam treatment plans. A negative value exhibits less dose received by the OAR from the CyberKnife treatment plan while a positive value shows a higher dose received from the CyberKnife treatment plan. To summarise the results, in 65·5 and 72·5% of patients, the brainstem and optic pathway received lesser doses in the CyberKnife treatment plans, respectively.

Figure 3. Doses received by the brainstem and optic pathway in treatment plans for CyberKnife and TrueBeam plans. A positive value signifies a higher dose from TrueBeam plans and vice versa.

For normal brain percent doses, V5Gy, V12Gy and V18Gy were evaluated and compared for CyberKnife and TrueBeam.

Figure 4 shows the comparison of doses received by the volume of normal brain in CyberKnife treatment plans versus TrueBeam. A negative value exhibits a better treatment plan with CyberKnife. Figure 4 shows that for the majority of patients (22 out of 29), the volume of the brain receiving 5 Gy dose is lower for CyberKnife plans than TrueBeam. A similar trend has been observed for V12Gy where 18 out of 10 treatment plans showed negative values. For V18Gy, the comparison of the values is insignificant. The absolute difference in values of volumetric doses for the normal brain is higher for V5Gy (mean 2·87 ± 1·77 Gy) than the difference for V12Gy (mean 1·18 with an exception of 16·7 Gy) and V18Gy (mean 0·34 ± 0·38 Gy).

Figure 4. Doses received by normal brain with CyberKnife and TrueBeam treatment plans. A negative number indicates that a smaller volume of the normal brain has received the respective dose in the case of the CyberKnife treatment plan.

Several research articles, previously published, support these results. These studies either include retrospective patient data or are done on a phantom.

Makoto Ito et al. Reference Ito, Kawamura and Mori20 compared treatment plans for TrueBeam and CyberKnife made on dummy targets on a phantom. The targets include cube and spherical shapes of different sizes. A total of 16 plans for each modality were made. They reported higher means target doses, less homogeneity and similar conformity in CyberKnife plans as compared with the TrueBeam plans.

Gevaert et al. Reference Gevaert, Levivier and Lacornerie23 compared the treatment plans of patients treated on Gamma Knife with the plans made on Novalis (Brain lab) and CyberKnife. They compared conformity, homogeneity and gradient indices of the plans made on these three machines. Their sample size was a total of 15 patients (5 acoustic neuromas and 10 AVM). Although they reported that Gamma Knife plans are superior to the other two, CyberKnife showed better values in all three indices as compared with the conventional linear accelerator (LINAC)-based plans.

Dutta et al. Reference Dutta, Subramanian and Murli24 compared seven patient plans retrospectively treated on CyberKnife diagnosed with acoustic neuromas. Their results also support our study with similar conclusions. They reported comparable values of CI in both plans but better OAR sparing in CyberKnife plans as compared with the plans made on Brain lab. Their results are different from ours in the aspect of low-dose spread. Their results showed that lesser volumes are being irradiated with low doses in Brain lab plans as compared with CyberKnife.

David Kaul et al. Reference Kaul, Badakhshi and Gevaert25 also did a retrospective study of patients diagnosed with meningioma. They compared plans of ten patients made on CyberKnife and LINAC-based systems (Novalis). They concluded that CyberKnife plans are superior in terms of conformity and homogeneity.

Most of the studies are confined to either single or two tumour types, with smaller volumes and only to a single fraction. The analysis of dose spillage into the normal tissue is also lacking in previous studies. Our study encompassed almost all types of brain tumours being treated for radiosurgery with different sizes including large tumours. It also includes plans comparison both with SRS and SRT techniques. The results are all supported by previous studies except few parameters.

Conclusion

The treatment plans made on CyberKnife showed better values of conformity and gradient indices, while the plans made on TrueBeam are more homogenous. The CyberKnife presented better results in sparing brainstem and optic structures as compared with the TrueBeam plans with similar target coverage. The normal brain was also more spared in the CyberKnife plans than TrueBeam. Low-dose spread is also a very important factor to consider while planning radiation therapy. For very low doses (5% of the prescribed dose), the TrueBeam showed better results. The volume of the body irradiated with a low dose is smaller for TrueBeam than for CyberKnife. But for higher doses, that is, 10, 20 and 50% of the prescribed doses, CyberKnife has shown an evident advantage over TrueBeam. From the results, it is concluded that CyberKnife is superior to TrueBeam for the delivery of SRS/SRT treatment to patients with intracranial tumours.

Acknowledgements

None.

References

Antypas, C and Pantelis, E Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system. Phys Med Biol 2008; 53 (17): 46974718. doi: 10.1088/0031-9155/53/17/016.CrossRefGoogle ScholarPubMed
Boyd, ME Advantages of stereotactic radiosurgery (SRS) over other radiotherapy techniques. In: Photonic Therapeutics and Diagnostics, Bartels, KE, Bass, LS, de Riese, WTW et al., editors. Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series. Vol. 5685, 2005: 291300. doi: 10.1117/12.580873.CrossRefGoogle Scholar
Stafinski, T, Jhangri, GS, Yan, E and Menon, D Effectiveness of stereotactic radiosurgery alone or in combination with whole brain radiotherapy compared to conventional surgery and/or whole brain radiotherapy for the treatment of one or more brain metastases: a systematic review and meta-analysis. Cancer Treatment Rev 2006; 32 (3): 203213. doi: 10.1016/j.ctrv.2005.12.009.CrossRefGoogle ScholarPubMed
Tran, A, Zhang, J, Woods, K et al. Treatment planning comparison of IMPT, VMAT and 4Π radiotherapy for prostate cases. Radiat Oncol 2017; 12 (1): 19. doi: 10.1186/s13014-016-0761-0.CrossRefGoogle ScholarPubMed
Zhang, Y, Chiu, T, Dubas, J et al. Benchmarking techniques for stereotactic body radiotherapy for early-stage glottic laryngeal cancer: LINAC-based non-coplanar VMAT versus CyberKnife planning. Radiat Oncol 2019; 14 (1): 110. doi: 10.1186/s13014-019-1404-z.CrossRefGoogle Scholar
Blamek, S, Grządziel, A and Miszczyk, L Robotic radiosurgery versus micro-multileaf collimator: a dosimetric comparison for large or critically located arteriovenous malformations. Radiat Oncol 2013; 8:205. [Online]. http://www.ro-journal.com/content/8/1/205 CrossRefGoogle Scholar
Benedict, SH, Perks, JR, Goetsch, S et al. Stereotactic radiation therapy. Comprehensive Biomed Phys 2014; 9:505527. doi: 10.1016/B978-0-444-53632-7.00925-4. Elsevier.CrossRefGoogle Scholar
Kaul, D, Badakhshi, H, Gevaert, T et al. Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of meningioma. Acta Neurochir (Wien) 2015; 157 (4): 559564. doi: 10.1007/s00701-014-2272-9.CrossRefGoogle ScholarPubMed
Paik, K, Kim, MS, Choi, CW et al. Dosimetric comparison of volumetric modulated arc therapy with robotic stereotactic radiation therapy in hepatocellular carcinoma. Radiat Oncol J 2015; 33 (3): 233241. doi: 10.3857/roj.2015.33.3.233.CrossRefGoogle ScholarPubMed
Borowicz, DM, Skrobała, A, Kruszyna-Mochalska, M et al. Does CyberKnife improve dose distribution versus IMRT and VMAT on a linear accelerator in low-risk prostate cancer? Radiol Oncol 2022; 56 (2): 259266. doi: 10.2478/raon-2022-0010.CrossRefGoogle ScholarPubMed
Weidlich, GA, Hacker, F, Bellezza, D et al. Ventricular tachycardia: a treatment comparison study of the CyberKnife with conventional linear accelerators. Cureus 2018; 10(10): e3445. doi: 10.7759/cureus.3445.Google ScholarPubMed
Maria Tomatis, S, Navarria, P, Franceschini, D et al. Comparing Varian EDGE and Gamma Knife Perfexion for brain metastases radiosurgery. Preliminary results. Poster presented at ESTRO 35 held in Turin, Italy, 2016. doi: 10.3252/pso.eu.ESTRO2016.2016.CrossRefGoogle Scholar
Eaton, DJ, Lee, J, Patel, R et al. Stereotactic radiosurgery for benign brain tumors: results of multicenter benchmark planning studies. Pract Radiat Oncol 2018; 8 (5): e295e304. doi: 10.1016/j.prro.2018.02.006.CrossRefGoogle ScholarPubMed
Torrens, M, Chung, C, Chung, HT et al. Standardization of terminology in stereotactic radiosurgery: report from the Standardization Committee of the International Leksell Gamma Knife Society: special topic. J Neurosur 2014; 121 (2): 215. doi: 10.3171/2014.7.gks141199.CrossRefGoogle ScholarPubMed
Tunstall, C, Watson, R, Clarke, E et al. Thames Valley and Wessex Radiotherapy Operational Delivery Network (ODN) Radiotherapy Protocols. Intracranial Stereotactic Radiosurgery/Radiotherapy, 2023. https://www.tvw-radiotherapy-odn.org.uk/protocols Google Scholar
Petkovska, S, Tolevska, C, Kraleva, S et al. Conformity Index for Brain Cancer Patients, 2010. https://inis.iaea.org/search/search.aspx?orig_q=RN:43026160 Google Scholar
Paddick, I A simple scoring ratio to index the conformity of radiosurgical treatment plans. J Neurosurg 2000; 93 (3): 219222.CrossRefGoogle ScholarPubMed
Al-Rawi, SAI, Abouelenein, H, Nagdy, ME et al. Assessment of dose gradient index variation during simultaneously integrated boost intensity-modulated radiation therapy for head and neck cancer patients. Precis Radiat Oncol 2022; 6 (3): 216224. doi: 10.1002/pro6.1166.CrossRefGoogle Scholar
Paddick, I and Lippitz, B A simple dose gradient measurement tool to complement the conformity index. J Neurosur 2006; 105:194201.CrossRefGoogle Scholar
Ito, M, Kawamura, T, Mori, Y et al. Dose distributions of high-precision radiotherapy treatment: a comparison between the cyber knife and true beam systems. Int J Radiat Res 2018; 16 (4): 395402. doi: 10.18869/acadpub.ijrr.16.4.395.Google Scholar
Hanna, GG, Murray, L, Patel, R et al. UK consensus on normal tissue dose constraints for stereotactic radiotherapy. Clin Oncol 2018; 30 (1): 514. doi: 10.1016/j.clon.2017.09.007.CrossRefGoogle ScholarPubMed
Diez, P, Hanna, GG, Aitken, KL et al. UK 2022 consensus on normal tissue dose-volume constraints for oligometastatic, primary lung and hepatocellular carcinoma stereotactic ablative radiotherapy. Clin Oncol 2022; 34 (5): 288300. doi: 10.1016/j.clon.2022.02.010.CrossRefGoogle ScholarPubMed
Gevaert, T, Levivier, M, Lacornerie, T et al. Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of arteriovenous malformations and acoustic neuromas. Radiother Oncol 2013; 106 (2): 192197. doi: 10.1016/j.radonc.2012.07.002.CrossRefGoogle ScholarPubMed
Dutta, D, Subramanian, SB, Murli, V et al. Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma. J Neuro-Oncol 2012; 106:637642. doi: 10.1007/s11060-011-0703-5.CrossRefGoogle ScholarPubMed
Kaul, D, Badakhshi, H, Gevaert, T et al. Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of meningioma. Acta Neurochir (Wien) 2015; 157 (4): 559564. doi: 10.1007/s00701-014-2272-9.CrossRefGoogle ScholarPubMed
Figure 0

Table 1. Acceptability criteria for target volumes and OARs in case of intracranial SRS/SRT21

Figure 1

Table 2. Comparison of evaluation indices in treatment plans of CyberKnife versus TrueBeam

Figure 2

Figure 1. The graphs showing planning endpoints for both CyberKnife and TrueBeam plans. For each parameter, a point having a value above 1 signifies higher values for that parameter for the CyberKnife plan and vice versa.

Figure 3

Table 3. Dose spillage in treatment plans of CyberKnife versus TrueBeam

Figure 4

Figure 2. Low doses received by the patient’s body from CyberKnife and TrueBeam plans as CK–TB values for the parameter. A negative value signifies the lower volume irradiated in CyberKnife plans.

Figure 5

Table 4. Doses of OARs in treatment plans of CyberKnife versus TrueBeam

Figure 6

Figure 3. Doses received by the brainstem and optic pathway in treatment plans for CyberKnife and TrueBeam plans. A positive value signifies a higher dose from TrueBeam plans and vice versa.

Figure 7

Figure 4. Doses received by normal brain with CyberKnife and TrueBeam treatment plans. A negative number indicates that a smaller volume of the normal brain has received the respective dose in the case of the CyberKnife treatment plan.