Hostname: page-component-cd9895bd7-mkpzs Total loading time: 0 Render date: 2024-12-24T01:54:22.710Z Has data issue: false hasContentIssue false

Body fat and fat-free mass measured by bioelectric impedance spectroscopy and dual-energy X-ray absorptiometry in obese and non-obese adults

Published online by Cambridge University Press:  31 August 2011

Paula Berstad*
Affiliation:
Research Centre, Akershus University Hospital, Box 26, 1478Lørenskog, Norway
Anna Randby
Affiliation:
Division of Medicine, Akershus University Hospital, Lørenskog, Norway Faculty Division, Akershus University Hospital, University of Oslo, Lørenskog, Norway
Gunn Seim Ekeland
Affiliation:
Faculty Division, Akershus University Hospital, University of Oslo, Lørenskog, Norway
Hege Ulveland
Affiliation:
Atlantis Medical College, Oslo, Norway
Torbjørn Omland
Affiliation:
Division of Medicine, Akershus University Hospital, Lørenskog, Norway Faculty Division, Akershus University Hospital, University of Oslo, Lørenskog, Norway
Kari Almendingen
Affiliation:
Research Centre, Akershus University Hospital, Box 26, 1478Lørenskog, Norway Department of Health, Nutrition and Management, Faculty of Health Sciences, Oslo and Akershus University College of Applied Sciences, Norway Institute of Chemistry, Biotechnology and Food Science, Norwegian University of Life Sciences, Ås, Norway
*
*Corresponding author: Dr P. Berstad, fax +47 67968860, email [email protected]
Rights & Permissions [Opens in a new window]

Abstract

The aim of the present study was to compare body fat mass (FM) and fat-free mass (FFM) estimates by bioelectric impedance spectroscopy (BIS), with respective estimates by dual-energy X-ray absorptiometry (DXA), in obese and non-obese subjects. Body composition was measured in ninety-three obese and non-obese men and women by BIS device, BodyScout (Fresenius Kabi, Bad Homburg, Germany) and DXA device, Lunar iDXA (GE Healthcare, Madison, WI, USA). Mean difference between the methods was analysed by t tests, and Bland–Altman plots were generated to further examine the differences between the methods. Mean difference between the estimates by DXA and BIS (ΔDXA − BIS and Bland–Altman 95 % limits of agreement) were as follows: FM 4·1 ( − 2·9, 11·2) kg and 4·5 ( − 2·9, 11·8) %, FFM − 4·1 ( − 11·2, 2·9) kg and − 4·5 ( − 11·9, 2·9) %, indicating large inter-individual variation and statistically significant underestimation of FM and overestimation of FFM by BIS, as compared to DXA. The underestimation of FMkg (FM measured in kg) and overestimation of FFMkg (FFM measured in kg) were more pronounced in men than in women, and the underestimation of FM% (FM measured in percent) and overestimation of FFM% (FFM measured in percent) were more pronounced in normal weight (BMI = 20·0–24·9 kg/m2) than in overweight and obese (BMI ≥ 25·0 kg/m2) subjects. BIS may be suitable for classification of a population into groups according to FM and FFM. However, the large inter-individual variation suggests that this BIS device with the proprietary software is insufficient for estimation of single individual body FM and FFM.

Type
Full Papers
Copyright
Copyright © The Authors 2011

Measurement of body composition is frequently used in the assessment of nutritional status. Low body fat-free mass (FFM) may be a better determinant of malnutrition(Reference Kyle, Morabia and Slosman1) and mortality(Reference Hitzl, Jorres and Heinemann2, Reference Schols, Broekhuizen and Weling-Scheepers3) in patients than low BMI. This is becoming more important as prevalence of obesity increases, and malnutrition in different patient groups may be hidden in obesity(Reference Lamb, Parr and Lamb4, Reference Pressoir, Desne and Berchery5). Consequently, there is need for a simple, valid bedside body composition assessment method.

Bioelectrical impedance is a non-invasive, patient-friendly, portable and relatively inexpensive method for body composition assessment. Bioimpedance technology estimates body water compartments(Reference Kyle, Bosaeus and De Lorenzo6, Reference Matthie7) on the basis of variation in electrical conductivity between different body tissues. In bioelectric impedance spectroscopy (BIS), electric current at a spectrum of frequencies is led through the body via electrodes placed on the skin. Resistivity of extracellular water (ECW) and intracellular water (ICW) compartments against the current varies for different frequencies. High-frequency currents penetrate through cell membranes, whereas low-frequency currents pass through the body, mainly in ECW, without penetration into cells. The measured impedance spectral data are fitted into a physical model, giving estimates for ECW and total body water (TBW) volumes. Using these estimates and measured subject weight and height, ICW, and further, body FFM and fat mass (FM) can be calculated. BIS is not limited to constant body geometry and relative hydration degree of FFM(Reference Kyle, Bosaeus and De Lorenzo6, Reference Deurenberg8, Reference Kyle, Bosaeus and De Lorenzo9), which both vary in obesity(Reference Deurenberg8, Reference Waki, Kral and Mazariegos10). Therefore, as compared to other bioelectrical impedance methods, BIS has higher sensitivity to predict water compartments and FFM in obese subjects or subjects with altered fluid balance(Reference Kyle, Bosaeus and De Lorenzo6, Reference Kyle, Bosaeus and De Lorenzo9, Reference Jaffrin and Morel11).

In order to use BIS for FFM estimation in clinical practice on both overweight and normal weight subjects, it has to be validated in a population with a wide spectrum of BMI. Although BIS is supposed to estimate body fluid compartments accurately in normal, healthy subjects(Reference Kyle, Bosaeus and De Lorenzo9), the validity may be reduced in obesity(Reference Cox-Reijven and Soeters12, Reference Ward, Dyer and Byrne13) and in patients in a malnourished state(Reference Ellegard, Ahlen and Korner14). Dual-energy X-ray absorptiometry (DXA) is a widely used reference method for body FFM and FM measurements(Reference Kyle, Bosaeus and De Lorenzo9). A recent study on cardiac patients concluded that the BIS device, BodyScout (Fresenius Kabi, Bad Homburg, Germany) only slightly overestimated FFM compared to DXA, but that the large inter-individual variation between the methods indicates that the device does not suit for body composition measurement at the individual level(Reference van Venrooij, Verberne and de Vos15). It remains unclear whether the validity of FFM and FM estimates by this BIS device differs between obese and non-obese subjects.

The aim of the present study was to examine the validity of BodyScout estimates of FM and FFM of the device, compared to DXA estimates, in obese and non-obese adults.

Methods

Subjects

The participants in the present study were recruited from The Akershus Sleep Apnea Project, a population-based cross-sectional study in Norway(Reference Hrubos-Strom, Randby and Namtvedt16). The aim of the two-phase main study was to assess the prevalence of sleep apnea in Norway. A total of 30 000 subjects aged 30–65 years and living in three different counties in Norway were randomly drawn from the National Population Registry and invited to respond to a questionnaire. A sub-group of 518 responders was further selected for a clinical examination.

We invited 150 randomly drawn participants for clinical examination to the present sub-study by telephone. Proportion of invited subjects with BMI ≥ 30·0/ < 30·0 kg/m2 was 80/20. A total of 108 subjects gave their consent for participation, whereas forty-two refused to participate. Among those who accepted to participate, eleven subjects were unable to meet up at the measurement appointment. A total of ninety-seven subjects were included in the present study, which included anthropometric measurements, a DXA analysis and a BIS test at the Akershus University Hospital, Lørenskog, Norway. The inclusion criterion for the subjects in the clinical sub-study was not to receive treatment for obstructive sleep apnea. Subjects with significant co-morbid conditions such as cancer, heart failure or severe chronic obstructive pulmonary disease were excluded. Further exclusion criteria in the present study were pregnancy, lack of safe contraception in fertile women and those with intra-cardiac device, cochlea implant or large metallic implants in their body.

The present study was conducted according to the guidelines laid down in the Declaration of Helsinki, and all procedures were approved by the Regional Ethics Committee, The Norwegian Directorate of Health and Norwegian Social Science Data Services. Written informed consent was obtained from all subjects. The radiation dose in DXA (0·05–0·15 Gy) was considered very low, less than a standard chest X-ray.

Measurement procedure

The data were collected between May and September 2008. After fasting for a minimum of 8 h and after avoiding hard physical exercise for 24 h, participants were met at the orthopaedic outpatient clinic in the morning. In all, five subjects reported not being in a fasting state. A trained health specialist (a dietitian (H. U.) or a master of science in physical activity and health (G. S. E.)) performed DXA scanning, the BIS test, and took the blood pressure and anthropometric measurements that included height, weight and waist, hip and arm circumferences. The participants also filled in a lifestyle questionnaire covering questions on physical activity, weight history, recent weight loss and chronic diseases and disabilities limiting physical activity.

Anthropometric measurements

All anthropometric measurements were conducted while the participant was without shoes and wearing only underwear. Height was measured to the nearest cm, and body weight was measured to the nearest 0·1 kg with a digital scale (Seca; Seca Deutchland, Hamburg, Germany). Mid-upper arm circumference was measured and standardised in the non-dominant arm using a scale band at the mid-point of the upper arm. Waist and hip circumferences were measured with a scale band, with the participant standing. Waist circumference was measured between the lowest ribs and crista iliaca after an out-breath. Hip circumference was measured at the level of the thickest point of nates.

Dual-energy X-ray absorptiometry

The DXA instrument used in the present study was Lunar iDXA (GE Healthcare, Madison, WI, USA). Whole-body scans were performed according to the manufacturer's instructions, and body fat, lean tissue mass (LTM) and bone mineral content (BMC) were analysed (software enCore version 11.10; GE Healthcare). The daily quality control procedure of the device was conducted automatically, and included a six-point calibration of three different bone-simulating chambers of known BMC and three different soft tissue-simulating chambers of known fat percentage. The program automatically determined total body measurement mode according to participant thickness (thick >25 cm; standard 13–25 cm and thin < 13 cm)(17). DXA estimate of FFM was calculated as a sum of LTM and BMC estimates. Precision of LTM and BMC estimates of the Lunar iDXA device has been found satisfactory by others (CV 0·5 and 0·6 %, respectively)(Reference Hind, Oldroyd and Truscott18).

A total of nineteen subjects were too large for the DXA instrument. They were scanned one-sided, and the instrument estimated measurement results for the whole body. Agreement in fat and lean mass estimates between whole-body DXA scan and estimates assessed by half-body DXA scan in obese subjects has been reported(Reference Rothney, Brychta and Schaefer19).

Bioelectric impedance spectroscopy

We conducted the measurement after the participant had taken off belts, watches, jewellery and other possibly electromagnetic objects and rested in a supine position for 5 min. During the measurement, the participant was lying in a supine position with four tactile electrodes placed on the skin, two on the dorsal surface of the dominant hand/wrist and two on the parallel foot/ankle according to the manufacturer's instructions. The BIS analyser used was BodyScout (Fresenius Kabi). The BodyScout device covers fifty frequencies in a range of 5 kHz to 1 MHz. The proprietary software extrapolates ECW and ICW using the Hanai equation(Reference Hanai20), as described by Moissl et al. (Reference Moissl, Wabel and Chamney21). The BodyScout estimates of FM and FFM are not sex specific and do not expect a fixed ECW:ICW ratio. The body composition model used in the BodyScout software is based on a principle that whole-body mass (M WB) is a sum of three compartments: normally hydrated adipose tissue mass (M NH_AT), normally hydrated LTM (M NH_LT) and excess fluid (ExF) mass (M ExF), as described by Chamney et al. (Reference Chamney, Wabel and Moissl22, Reference Wabel, Chamney and Moissl23):

(1)
 M _{WB} = M _{NH\lowbar AT} + M _{NH\lowbar LT} + M _{ExF}

The total water (TW), ICW and ECW components of M NH_LT and M NH_AT are defined by equations 27.

(2)
 H _{TW\lowbar NH\lowbar LT} = M _{TW\lowbar NH\lowbar LT}/ M _{NH\lowbar LT}
(3)
 H _{ECW\lowbar LT} = M _{ECW\lowbar NH\lowbar LT}/ M _{NH\lowbar LT}
(4)
 H _{ICW\lowbar NH\lowbar LT} = M _{ICW\lowbar NH\lowbar LT}/ M _{NH\lowbar LT}
(5)
 H _{TW\lowbar NH\lowbar AT} = M _{TW\lowbar NH\lowbar AT}/ M _{NH\lowbar AT}
(6)
 H _{ECW\lowbar NH\lowbar AT} = M _{ECW\lowbar NH\lowbar AT}/ M _{NH\lowbar AT}
(7)
 H _{ICW\lowbar NH\lowbar AT} = M _{ICW\lowbar NH\lowbar AT}/ M _{NH\lowbar AT}

where H TW_NH_LT, H ECW_LT and H ICW_NH_LT are fractional mass of TW, ECW and ICW in normally hydrated lean tissue, and H TW_NH_AT, H ECW_NH_AT and H ICW_NH_AT are fractional mass of TW, ECW and ICW in normally hydrated adipose tissue, respectively. In the BodyScout software, total body FM in kg (FMkg) is related to M NH_AT, as given in equation 8.

(8)
FMkg = M _{NH\lowbar AT}\times (1 - H _{TW\lowbar NH\lowbar AT} - K _{AR})

where K AR is the ratio of residual adipose components (solids, mainly protein and mineral) to M NH_AT, with a value of typically 0·05(Reference Holliday24, Reference Martin, Daniel and Drinkwater25). FFMkg (FFM measured in kg) is calculated by the software by extracting FMkg from body weight.

BIS estimate of FFM (FFMkgTBW/0·73) was also calculated directly from TBW, according to equation 9, based on the assumption of a 73 % hydration of the FFM(Reference Ellis26).

(9)
FFMkg_{TBW/0\cdot 73} = TBW/0\cdot 73

Further, FMkgTBW/0·73 was calculated as FFMkgTBW/0·73 extracted from body weight.

Lifestyle questionnaire

In the lifestyle questionnaire, the participants were asked to report how often during the last year they were involved in non-vigorous and vigorous physical activities, choosing between four levels for both: never, < 1 h/week, 1–2 h/week and 3 h/week or more. They were also asked to report their habitual activity level with four choices: sedate, light activity 4 h/week or more, condition training 4 h/week or more and regular hard physical training. The participants were asked to describe their smoking status by choosing between non-smoker and current smoker, and to give the number of cigarettes smoked daily, years since start and cessation of smoking and how many years totally smoked. The participants were asked to assess their body weight at ages 18, 30, 40, 50 and 60 years, 6 and 12 months before the study date and the lowest and highest weight since the age of 18 years. The questionnaire also included a question on whether the participant had one or more chronic diseases or disabilities limiting physical activity, and if answering ‘Yes’, the participant was asked to write down the disease/disability.

Pilot study

We conducted a pilot study on the anthropometric, BIS and DXA measurements in twenty hospital employees and other voluntary subjects recruited by the project team in order to study variability in measurements between the two project team members who took the measurements (H. U. and G. S. E.). There was no statistically significant difference in the measurements of arm, waist and hip circumference or BIS and DXA measurements between the measurers (one-sample t test, all P ≥ 0·21). There was a statistically significant slight difference between the measurers in height and weight measurements (mean differences 0·1 cm and 0·1 kg), but this difference was without clinical significance.

Statistical analysis

Power calculation

Assuming that the participants have an average weight of 90 kg and 40 % fat gave an estimated total FM of around 36 (sd 3) kg. Given an α of 5 % and statistical power (1 − β) of 80 %, the minimum number of obese subjects to be included in each group, to detect a difference of 2 kg FM, was then n =2 × (3/2)2 × 7·9 = 35 per group. We also invited twenty healthy and normal weight persons to make sure to had a control group with optimal measurement conditions and results.

Data analysis

There is a considerable difference between the BIS and DXA technologies in estimating FMkg and FFMkg. In DXA, FM is directly estimated in kg, and FFMkg can be calculated as the sum of LTM and BMC estimates in kg. The percentage estimates of FM and FFM are calculated using the sum of all body mass compartments in kg as body weight. In BIS, body weight obtained from an independent measurement (a digital scale in the present study) is inserted into the software before the measurements. FMkg and FFMkg estimates are calculated using the percentage estimates of these and the given body weight in kg. The mean body weight measured by the digital scale in the present study was 0·5 (sd 0·8) kg higher than the mean DXA estimate of body weight. However, this did not significantly differ between obese and other subjects, overweight and normal weight subjects or between half- and whole-DXA-scanned subjects (data not shown). Consequently, the FMkg and FFMkg estimates of DXA were converted to values, the sum of which totalled body weight by the digital scale. We calculated physical activity score as a sum of scores from non-vigorous and vigorous activity, giving scores 0–3 for the four choices of non-vigorous activity and double scores for the respective choices of vigorous activity. This resulted in total physical activity scores ranging from 0 to 9. We further categorised the total physical activity scores into low ( ≤ 3), middle (4–6) and high (7–9) activity groups. We gave score 0–2 for the habitual physical activity choices, combining the two choices with highest activity. Smoking status was reported as never, current and ex-smoker. Weight change during the preceding 12 months was calculated using the self-reported weight for 12 months earlier and the measured weight. Diseases and disabilities limiting physical activity were categorised according to type of disease/disability. Descriptive statistics are reported as distributions and as mean values with standard deviations. Bland–Altman plots(Reference Bland and Altman27), a well-known tool for comparing a new and a referent method, were used to present differences between FM and FFM estimates by DXA and BIS. The limits of agreement between the two methods were defined as mean difference ± 1·96 sd(Reference Bland and Altman27). We tested mean difference between the methods (ΔDXA − BIS) with paired-samples t test, and used independent-samples t test (if two subgroups) or one-way ANOVA (if three subgroups) to analyse differences in ΔDXA − BIS between subgroups. We considered two-tailed P < 0·05 as statistically significant. We conducted the statistical analysis using software SPSS 17.0.1 for Windows (SPSS Inc., Chicago, IL, USA).

Results

Subject characteristics are as described in Table 1. A total of ninety-three participants (61 % males) with successful measurements were included in the study, with a mean age of 51 years. The majority of the participants (fifty-nine subjects, 63 %) were obese (BMI ≥ 30·0 kg/m2), whereas twenty-two subjects (24 %) were overweight (BMI = 25·0–29·9 kg/m2) and twelve subjects (13 %) had a normal BMI (20·0–24·9 kg/m2). Some participants showed variations in the weight measured between the clinical examination of the main study and the present study measurement date, therefore the intended proportion of subjects with BMI ≥ 30·0/ < 30·0 kg/m2 (80/20) varied to some extent. A total of 41 % of the participants reported chronic diseases or disabilities limiting physical activity.

Table 1 Characteristics of study participants

(Mean values and standard deviations, number of participants and percentages, n 93)

DXA, dual-energy X-ray absorptiometry; FM, fat mass; FFM, fat-free mass; BIS, bioelectric impedance spectroscopy; ECW, extracellular water; ICW, intracellular water; R e, resistance of extracellular volume; R i, resistance of intracellular volume.

* Rheumatic, pulmonary or psychiatric problems, fibromyalgi or others.

FFMkg and FFM as a proportion (FFM%) were significantly higher in men than in women, whereas FM as a proportion (FM%) was significantly higher in women, estimated with both DXA and BIS (all P < 0·001). There was no significant difference in FMkg between the sexes (Table 1).

Fat mass comparison

BIS significantly underestimated mean FMkg and FM% as compared to DXA (Table 2 and Fig. 1(a) and (b)). Mean differences of the estimates between the methods (ΔDXA − BIS) and Bland–Altman 95 % limits of agreements were 4·1 ( − 2·9, 11·2) kg and 4·5 ( − 2·9, 11·8) % for FMkg and FM%, respectively (Table 2 and Fig. 1(a) and (b)). The estimates by the two methods were highly correlated (Table 2). There was a statistically significant inverse correlation between the difference and mean values of FM (r Pearson = − 0·27 and − 0·54 for FMkg and FM%, respectively, P < 0·01 for both; Fig. 1(a) and (b)). Underestimation of FMkg was significantly more pronounced in men than in women (ΔDXA − BIS (sd) 4·8 (3·8) v. 3·1 (3·0) kg, P = 0·02; Table 3). Underestimation of FM% was significantly more pronounced in normal weight subjects than in overweight and obese subjects (ΔDXA − BIS (sd) 6·9 (2·7) v. 4·1 (3·8) %, P = 0·005; Table 4). The estimation difference for FM% between the two methods was significantly inversely correlated with the FM% estimate by DXA (r Pearson = − 0·29, P = 0·005; data not shown). The BIS estimates calculated as 73 % hydration of FFM (FMkgTBW/0·73 and FM%TBW/0·73) were significantly higher than the respective BIS estimates by the proprietary software. Consequently, the underestimation of the FM values was significantly smaller than for the estimates given by the BIS software. However, the method differences for FMkgTBW/0·73 and FM%TBW/0·73DXA − BIS(TBW/0·73)) were statistically significant (Table 2) and all results for FMTBW/0·73 estimates were close to identical with FM estimates by BIS software (data not shown).

Table 2 Comparison of fat mass (FM) and fat-free mass (FFM) estimates by dual-energy X-ray absorptiometry (DXA) and bioelectric impedance spectroscopy (BIS) by the proprietary software (BISsoftware) and by direct calculation from total body water (TBW) by BIS

(Mean values and standard deviations, n 93)

*** Mean values between the DXA and BIS methods were significantly different (P < 0·001).

FM and FFM estimated by BIS software, as described by Chamney et al. (Reference Chamney, Wabel and Moissl22).

FM and FFM estimated by direct calculation from TBW by BIS, expecting 73 % hydration of FFM(Reference Ellis26).

Fig. 1 Bland–Altman plots of comparisons of bioelectric impedance spectroscopy (BIS) with dual-energy X-ray absorptiometry (DXA) estimates for (a) body fat mass (FM) in kg (FMkg), (b) body FM as percentage (FM%), (c) body fat-free mass (FFM) in kg (FFMkg) and (d) body FFM as a percentage (FFM%) with limits of agreement (LoA) (dashed lines) and Pearson correlation coefficient (r) (n 93).

Table 3 Difference between the methods (ΔDXA−BIS), where DXA is dual-energy X-ray absorptiometry and BIS is bioelectric impedance spectroscopy, in fat mass (FM) and fat-free mass (FFM) estimates in men and women

(Mean values and standard deviations)

* Mean values were significantly different from men (P < 0·05).

Table 4 Difference between the methods (ΔDXA−BIS), where DXA is dual-energy X-ray absorptiometry and BIS is bioelectric impedance spectroscopy, in fat mass (FM) and fat-free mass (FFM) estimates in normal weight (BMI <25 kg/m2) and overweight (BMI ≥25 kg/m2) subjects

(Mean values and standard deviations)

** Mean values were significantly different from normal weight subjects (P < 0·01).

There was neither a statistically significant difference in the underestimation of FMkg or FM% by BIS between obese (BMI ≥ 30·0 kg/m2) and other subjects nor between very obese (BMI ≥ 34·0 kg/m2, eighteen subjects) and other subjects, or between subjects with half- and whole-body DXA scan (data not shown). There was no difference in the underestimation through age groups, physical activity scores or recent weight change groups (data not shown).

When grouping subjects into FMkg and FM% quartiles by BIS and DXA estimates, 74 and 73 % of the subjects, respectively, were classified into the same quartile by the two methods, whereas 26 and 25 %, respectively, were classified into same ± 1 quartile (Table 5).

Table 5 Joint classification of subjects by quartiles of fat mass (FM) and fat-free mass (FFM) estimated by dual-energy X-ray absorptiometry (DXA) and bioelectric impedance spectroscopy (BIS) (n 93)

Fat-free mass comparison

BIS significantly overestimated mean FFMkg and FFM% as compared to DXA (Table 2 and Fig. 1(c) and 1(d)). ΔDXA − BIS and Bland–Altman limits of agreements were − 4·1 ( − 11·2, 2·9) kg and − 4·5 ( − 11·9, 2·9) % for FFMkg and FFM%, respectively (Table 2 and Fig. 1(c) and (d)). The estimates by the two methods were highly correlated (Table 2). There was a statistically significant inverse correlation between the difference and mean values of FFM (r Pearson = − 0·36 and − 0·53, for FFMkg and FFM%, respectively, P < 0·01 for both; Fig. 1(c) and (d)). Overestimation of the absolute FFMkg was significantly more pronounced in men than in women (ΔDXA − BIS (sd) − 4·8 (3·8) v. − 3·1 (3·0) kg, P = 0·03; Table 3). Additionally, overestimation of FFM% was significantly more pronounced in normal weight subjects than in overweight and obese subjects (ΔDXA − BIS (sd) − 6·9 (2·8) v. − 4·1 (3·8) %, P = 0·006; Table 4). The estimation difference for FFM% between the two methods was significantly inversely correlated with the FFM% estimate by DXA (r Pearson = − 0·33, P = 0·001; data not shown). The BIS estimates calculated as 73 % hydration of FFM (FFMkgTBW/0·73 and FFM%TBW/0·73) were significantly lower than the respective BIS estimates by the proprietary software. Consequently, the overestimation of the FFM values was significantly smaller than for the estimates given by the BIS software. However, the method differences for FFMkgTBW/0·73 and FFM%TBW/0·73DXA − BIS(TBW/0·73)) were statistically significant (Table 2) and all results for FFMTBW/0·73 estimates were close to identical with FFM estimates by BIS software (data not shown).

There was neither statistically significant difference in the overestimation of FFMkg or FFM% by BIS between obese (BMI ≥ 30·0 kg/m2) and other subjects nor between very obese (BMI ≥ 34·0 kg/m2, eighteen subjects) and other subjects, or between subjects with half- and whole-body DXA scans (data not shown). There was no difference in the overestimation through age groups, physical activity scores or recent weight change groups (data not shown).

When grouping subjects into FFMkg and FFM% quartiles by BIS and DXA estimates, 74 and 71 % of the subjects, respectively, were classified into the same quartile by the two methods, whereas 26 and 26 %, respectively, were classified into same ± 1 quartile (Table 5).

Discussion

In the present validation study, we observed large inter-individual variation in FM and FFM estimates between BIS and DXA. We observed a significant underestimation of FM and overestimation of FFM by BIS, as compared to DXA. The underestimation of FMkg and the overestimation of FFMkg were higher in men than in women. The underestimation of FM% and the overestimation of FFM% were both higher at normal compared to high BMI. Furthermore, underestimation of FM% was more pronounced at low FM% and overestimation of FFM% was more pronounced at high FFM%. The FFM and FM estimates calculated from TBW by BIS were closer to these estimates from DXA. The large inter-individual variation and heterogeneity in validity between subgroups indicate that BIS is not a suitable method for single body FM or FFM assessment at an individual level. The present BIS device, BodyScout, with the proprietary software was initially designed to estimate body liquid compartments. The present study demonstrates that it may not be equally suitable for estimation of body FM and FFM.

In the present study population with obese and non-obese subjects, using the present BIS device, the BIS estimates of FM and FFM will be significantly better if these estimates are calculated directly from TBW by BIS, expecting 73 % hydration of FFM. Such observation has been done earlier by Ellegard et al. (Reference Ellegard, Ahlen and Korner14). Earlier comparisons in different population groups have, in line with the present results, found large inter-individual variation in mean difference between BIS and DXA, accompanied by overestimation of FFM(Reference van Venrooij, Verberne and de Vos15, Reference Carlsson, Bosaeus and Nordgren28, Reference Svantesson, Zander and Klingberg29) and underestimation of FM(Reference Svantesson, Zander and Klingberg29) by BIS. Underestimation of FFM by BIS has also been reported, but in incurable cancer patients(Reference Ellegard, Ahlen and Korner14). In contrast to the present study, the overestimation of FFM by BIS was, in most previous studies, in adults not related to BMI, other anthropometric variables, sex(Reference Carlsson, Bosaeus and Nordgren28, Reference Svantesson, Zander and Klingberg29) or magnitude of FFM and FM estimates(Reference Svantesson, Zander and Klingberg29). In one study, overestimation of FFM by BIS compared to DXA was related to overweight(Reference Ward, Dyer and Byrne13). In children, BIS has been found to overestimate FM in lean and underestimate FM in overweight subjects(Reference Fors, Gelander and Bjarnason30). Higher validity of FFM% and FM% estimates in overweight than in normal weight subjects is, to our knowledge, reported only by the present study. The number of normal weight subjects in the present study (n 12) probably is too low to draw conclusions on this. However, the significant correlations between the method difference (ΔFM%DXA − BIS and ΔFFM%DXA − BIS) and the FM% and FFM% values, respectively, indicate an inverse dose–response association between validity of these estimates and adiposity of the subject.

BIS validation studies against those methods other than DXA have compared BIS estimates of ECW with dilution methods and ICW with total body potassium. These have found reduced BIS validity in disease states or obesity(Reference Cox-Reijven and Soeters12, Reference Moissl, Wabel and Chamney21). A suggested explanation for reduced validity of BIS measurements in obesity is the sensitivity of BIS to detect imbalance in body water compartments. BIS estimates of ECW and ICW are altered by changes in these, as found by Moissl in patients with renal failure and healthy subjects(Reference Moissl, Wabel and Chamney21). The ECW:ICW ratio is greater in obese subjects due to an increased amount of ECW and TBW(Reference Waki, Kral and Mazariegos10), as compared to non-obese subjects, leading to overestimation of FFM and underestimation of FM in obesity. Although BIS is sensitive to body water changes in normal, healthy subjects(Reference Kyle, Bosaeus and De Lorenzo9), accuracy of BIS estimates of body water decreases in obesity(Reference Cox-Reijven and Soeters12). The opposite, a larger measurement error by BIS in normal weight subjects in the present study, cannot be explained by this reduced BIS sensitivity in an abnormal body fluid balance. However, we observed larger measurement error by BIS in normal weight subjects only in the percentage proportions of FFM and FM. The reduced validity of BIS in altered fluid balance has been associated rather with the absolute estimates of ICW and ECW in litre(Reference Cox-Reijven and Soeters12, Reference Moissl, Wabel and Chamney21).

Another possible explanation for the large inter-individual variability between BIS and DXA may be a reduced validity of DXA measurements in obese subjects. Although DXA has been found to be an accurate method for body fat and lean tissue measurements(Reference Svendsen, Haarbo and Hassager31), underestimation of FFM in women and reduced validity in FFM estimates by truncal adiposity(Reference Hansen, Raja and Aslani32) and increasing body size(Reference Williams, Wells and Wilson33) have been reported. DXA has been suspected to underestimate FM in severely obese subjects as a result of increased photon absorption(Reference Levitt, Beckman and Mager34). If the DXA measurements in the present study were similarly biased, the observed underestimation of FM by BIS in obese subjects may have been smaller than the real underestimation in these subjects. In that case, the difference in BIS measurement error between normal weight and overweight subjects would, in reality, be smaller than observed. We did not observe a differential method difference between the subjects scanned on one side by DXA (n 19) and others. We can therefore not suspect a reduced precision of the DXA device in estimation of total body FM and FFM from the one-sided DXA scan in the present study.

It has been suggested that BIS is suitable for assessment of body composition in groups or individual body composition changes rather than determination of single individual body composition(Reference Buchholz, Bartok and Schoeller35). BIS estimates of individual TBW measurements have not been considered to be valid in a population with a wide range of BMI(Reference Moon, Tobkin and Roberts36), but BIS has been reported to be a reproducible method within an individual(Reference Andersen, Jodal and Arveschoug37, Reference Moon, Stout and Smith38) and track intra-individual changes in TBW accurately, regardless of BMI(Reference Moon, Tobkin and Roberts36, Reference Moon, Stout and Smith38). Estimating changes in body composition may be the most applicable use of BIS in clinical practice, although a recent study suggested that BIS is inappropriate for this purpose also because of the large inter-individual variability(Reference van Venrooij, Verberne and de Vos15). We found that BIS classifies the population relatively accurately into quartiles according to FM and FFM. This may be satisfactory for epidemiological studies. Body composition assessment is, however, mainly used in clinical studies and accurate individual measurements are necessary. Hence, correct ranging into groups is not sufficient for validity of the device.

Careful participant recruitment and standardised examination procedures are strengths of the present study. The pilot study conducted before the data collection showed agreement in measurements between the two measurers. Corrections were made in height and weight measurement procedures in order to minimise the slight measurer difference. The small number of normal weight subjects is a limitation of the present study. Use of validated regression equations has been warranted as a condition of clinical bioimpedance technology use in subjects with abnormal body shape(Reference Kyle, Bosaeus and De Lorenzo6). We did not, for the purpose of the present study, develop new predicted equations based on BodyScout estimates of body fluid compartments, resistance and BMI, as the aim of the present study was to validate the BodyScout device ‘as is’. The number of normal weight subjects in the present study sample would probably have been too low to develop equations useful for clinical practice.

We conclude that we found a large inter-individual agreement in FM and FFM estimates between the present BIS device and DXA. We also observed that this BIS device underestimates FM and overestimates FFM. In agreement with a recent validation study(Reference van Venrooij, Verberne and de Vos15), we conclude that this BIS device using the proprietary software has too low an accuracy to be used in estimation of body FM and FFM at the individual level. The absolute BIS estimates in kg may be less valid in men than in women, and may be less comparable with DXA estimates than the proportional estimates. In epidemiological studies, a population can be satisfactorily classified into groups by the FFM and FM estimates. Finally, we suggest that future validation studies on BIS among obese subjects include ‘gold standard’ methods for estimation of body fluids.

Acknowledgements

The authors are grateful to the subjects who participated in the present study and to the orthopaedic outpatient clinic at Akershus University Hospital for providing the facilities, and, in particular, to Hanna Småstu for practical help and advice. Eastern Norway Regional Health Authority RHF (no. 2006094) is acknowledged for their financial support. P. B. conducted the statistical analysis and wrote the manuscript. A. R., K. A. and T. O. designed the study. A. R. included the study subjects and coordinated data collection. G. S. E. and H. U. conducted the data collection. T. O. and K. A. obtained funding. K. A. is project leader for the ASAP study (main study) and K. A. and T. O. supervised the study. All authors critically revised the manuscript. The authors declare no conflict of interest.

References

1 Kyle, UG, Morabia, A, Slosman, DO, et al. (2001) Contribution of body composition to nutritional assessment at hospital admission in 995 patients: a controlled population study. Br J Nutr 86, 725731.CrossRefGoogle ScholarPubMed
2 Hitzl, AP, Jorres, RA, Heinemann, F, et al. (2010) Nutritional status in patients with chronic respiratory failure receiving home mechanical ventilation: impact on survival. Clin Nutr 29, 6571.CrossRefGoogle ScholarPubMed
3 Schols, AM, Broekhuizen, R, Weling-Scheepers, CA, et al. (2005) Body composition and mortality in chronic obstructive pulmonary disease. Am J Clin Nutr 82, 5359.CrossRefGoogle ScholarPubMed
4 Lamb, CA, Parr, J, Lamb, EI, et al. (2009) Adult malnutrition screening, prevalence and management in a United Kingdom hospital: cross-sectional study. Br J Nutr 102, 571575.CrossRefGoogle Scholar
5 Pressoir, M, Desne, S, Berchery, D, et al. (2010) Prevalence, risk factors and clinical implications of malnutrition in French Comprehensive Cancer Centres. Br J Cancer 102, 966971.CrossRefGoogle ScholarPubMed
6 Kyle, UG, Bosaeus, I, De Lorenzo, AD, et al. (2004) Bioelectrical impedance analysis-part II: utilization in clinical practice. Clin Nutr 23, 14301453.CrossRefGoogle ScholarPubMed
7 Matthie, JR (2008) Bioimpedance measurements of human body composition: critical analysis and outlook. Expert Rev Med Devices 5, 239261.CrossRefGoogle ScholarPubMed
8 Deurenberg, P (1996) Limitations of the bioelectrical impedance method for the assessment of body fat in severe obesity. Am J Clin Nutr 64, 449S452S.CrossRefGoogle ScholarPubMed
9 Kyle, UG, Bosaeus, I, De Lorenzo, AD, et al. (2004) Bioelectrical impedance analysis–part I: review of principles and methods. Clin Nutr 23, 12261243.CrossRefGoogle ScholarPubMed
10 Waki, M, Kral, JG, Mazariegos, M, et al. (1991) Relative expansion of extracellular fluid in obese vs. nonobese women. Am J Physiol 261, E199E203.Google ScholarPubMed
11 Jaffrin, MY & Morel, H (2008) Body fluid volumes measurements by impedance: a review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods. Med Eng Phys 30, 12571269.CrossRefGoogle ScholarPubMed
12 Cox-Reijven, PL & Soeters, PB (2000) Validation of bio-impedance spectroscopy: effects of degree of obesity and ways of calculating volumes from measured resistance values. Int J Obes Relat Metab Disord 24, 271280.CrossRefGoogle ScholarPubMed
13 Ward, LC, Dyer, JM, Byrne, NM, et al. (2007) Validation of a three-frequency bioimpedance spectroscopic method for body composition analysis. Nutrition 23, 657664.CrossRefGoogle ScholarPubMed
14 Ellegard, LH, Ahlen, M, Korner, U, et al. (2009) Bioelectric impedance spectroscopy underestimates fat-free mass compared to dual energy X-ray absorptiometry in incurable cancer patients. Eur J Clin Nutr 63, 794801.CrossRefGoogle ScholarPubMed
15 van Venrooij, LM, Verberne, HJ, de Vos, R, et al. (2010) Preoperative and postoperative agreement in fat free mass (FFM) between bioelectrical impedance spectroscopy (BIS) and dual-energy X-ray absorptiometry (DXA) in patients undergoing cardiac surgery. Clin Nutr 29, 789794.CrossRefGoogle ScholarPubMed
16 Hrubos-Strom, H, Randby, A, Namtvedt, SK, et al. (2010) A Norwegian population-based study on the risk and prevalence of obstructive sleep apnea. J Sleep Res 20, 162170.CrossRefGoogle Scholar
17 GE Medical Systems (2005) Lunar iDXA Safety and Specification Manual no. L11347. Madison, WI: GE Medical Systems Lunar.Google Scholar
18 Hind, K, Oldroyd, B & Truscott, JG (2011) In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adults. Eur J Clin Nutr 65, 140142.CrossRefGoogle ScholarPubMed
19 Rothney, MP, Brychta, RJ, Schaefer, EV, et al. (2009) Body composition measured by dual-energy X-ray absorptiometry half-body scans in obese adults. Obesity (Silver Spring) 17, 12811286.CrossRefGoogle ScholarPubMed
20 Hanai, T (1968) Electrical properties of emulsions. In Emulsion Science, pp. 354477. London: Academic Press.Google Scholar
21 Moissl, UM, Wabel, P, Chamney, PW, et al. (2006) Body fluid volume determination via body composition spectroscopy in health and disease. Physiol Meas 27, 921933.CrossRefGoogle ScholarPubMed
22 Chamney, PW, Wabel, P, Moissl, UM, et al. (2007) A whole-body model to distinguish excess fluid from the hydration of major body tissues. Am J Clin Nutr 85, 8089.CrossRefGoogle ScholarPubMed
23 Wabel, P, Chamney, P, Moissl, U, et al. (2009) Importance of whole-body bioimpedance spectroscopy for the management of fluid balance. Blood Purif 27, 7580.CrossRefGoogle ScholarPubMed
24 Holliday, MA (1986) Body composition and energy need during growth. In Human growth, pp. 101117. New York, NY: Plenum Press.Google Scholar
25 Martin, AD, Daniel, MZ, Drinkwater, DT, et al. (1994) Adipose tissue density, estimated adipose lipid fraction and whole body adiposity in male cadavers. Int J Obes Relat Metab Disord 18, 7983.Google ScholarPubMed
26 Ellis, KJ (2000) Human body composition: in vivo methods. Physiol Rev 80, 649680.CrossRefGoogle ScholarPubMed
27 Bland, JM & Altman, DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8, 135160.CrossRefGoogle ScholarPubMed
28 Carlsson, E, Bosaeus, I & Nordgren, S (2004) Body composition in patients with short bowel syndrome: an assessment by bioelectric impedance spectroscopy (BIS) and dual-energy absorptiometry (DXA). Eur J Clin Nutr 58, 853859.CrossRefGoogle ScholarPubMed
29 Svantesson, U, Zander, M, Klingberg, S, et al. (2008) Body composition in male elite athletes, comparison of bioelectrical impedance spectroscopy with dual energy X-ray absorptiometry. J Negat Results Biomed 7, 1.CrossRefGoogle ScholarPubMed
30 Fors, H, Gelander, L, Bjarnason, R, et al. (2002) Body composition, as assessed by bioelectrical impedance spectroscopy and dual-energy X-ray absorptiometry, in a healthy paediatric population. Acta Paediatr 91, 755760.CrossRefGoogle Scholar
31 Svendsen, OL, Haarbo, J, Hassager, C, et al. (1993) Accuracy of measurements of body composition by dual-energy X-ray absorptiometry in vivo. Am J Clin Nutr 57, 605608.CrossRefGoogle ScholarPubMed
32 Hansen, RD, Raja, C, Aslani, A, et al. (1999) Determination of skeletal muscle and fat-free mass by nuclear and dual-energy X-ray absorptiometry methods in men and women aged 51–84 y (1–3). Am J Clin Nutr 70, 228233.CrossRefGoogle ScholarPubMed
33 Williams, JE, Wells, JC, Wilson, CM, et al. (2006) Evaluation of Lunar Prodigy dual-energy X-ray absorptiometry for assessing body composition in healthy persons and patients by comparison with the criterion 4-component model. Am J Clin Nutr 83, 10471054.CrossRefGoogle ScholarPubMed
34 Levitt, DG, Beckman, LM, Mager, JR, et al. (2010) Comparison of DXA and water measurements of body fat following gastric bypass surgery and a physiological model of body water, fat, and muscle composition. J Appl Physiol 109, 786795.CrossRefGoogle Scholar
35 Buchholz, AC, Bartok, C & Schoeller, DA (2004) The validity of bioelectrical impedance models in clinical populations. Nutr Clin Pract 19, 433446.CrossRefGoogle ScholarPubMed
36 Moon, JR, Tobkin, SE, Roberts, MD, et al. (2008) Total body water estimations in healthy men and women using bioimpedance spectroscopy: a deuterium oxide comparison. Nutr Metab (Lond) 5, 7.CrossRefGoogle ScholarPubMed
37 Andersen, TB, Jodal, L, Arveschoug, A, et al. (2011) Precision and within- and between-day variation of bioimpedance parameters in children aged 2–14 years. Clin Nutr 30, 326331.CrossRefGoogle ScholarPubMed
38 Moon, JR, Stout, JR, Smith, AE, et al. (2010) Reproducibility and validity of bioimpedance spectroscopy for tracking changes in total body water: implications for repeated measurements. Br J Nutr 104, 13841394.CrossRefGoogle ScholarPubMed
Figure 0

Table 1 Characteristics of study participants(Mean values and standard deviations, number of participants and percentages, n 93)

Figure 1

Table 2 Comparison of fat mass (FM) and fat-free mass (FFM) estimates by dual-energy X-ray absorptiometry (DXA) and bioelectric impedance spectroscopy (BIS) by the proprietary software (BISsoftware) and by direct calculation from total body water (TBW) by BIS(Mean values and standard deviations, n 93)

Figure 2

Fig. 1 Bland–Altman plots of comparisons of bioelectric impedance spectroscopy (BIS) with dual-energy X-ray absorptiometry (DXA) estimates for (a) body fat mass (FM) in kg (FMkg), (b) body FM as percentage (FM%), (c) body fat-free mass (FFM) in kg (FFMkg) and (d) body FFM as a percentage (FFM%) with limits of agreement (LoA) (dashed lines) and Pearson correlation coefficient (r) (n 93).

Figure 3

Table 3 Difference between the methods (ΔDXA−BIS), where DXA is dual-energy X-ray absorptiometry and BIS is bioelectric impedance spectroscopy, in fat mass (FM) and fat-free mass (FFM) estimates in men and women(Mean values and standard deviations)

Figure 4

Table 4 Difference between the methods (ΔDXA−BIS), where DXA is dual-energy X-ray absorptiometry and BIS is bioelectric impedance spectroscopy, in fat mass (FM) and fat-free mass (FFM) estimates in normal weight (BMI <25 kg/m2) and overweight (BMI ≥25 kg/m2) subjects(Mean values and standard deviations)

Figure 5

Table 5 Joint classification of subjects by quartiles of fat mass (FM) and fat-free mass (FFM) estimated by dual-energy X-ray absorptiometry (DXA) and bioelectric impedance spectroscopy (BIS) (n 93)