Hostname: page-component-586b7cd67f-t8hqh Total loading time: 0 Render date: 2024-11-23T17:48:32.218Z Has data issue: false hasContentIssue false

Dosimetry and microdosimetry using LET spectrometer based on the track-etch detector: radiotherapy bremsstrahlung beam, onboard aircraft radiation field

Published online by Cambridge University Press:  14 December 2006

I. Jadrnícková
Affiliation:
Dept. of Radiation Dosimetry, Nuclear Physics Institute AS CR, Na Truhláøce 39/64, 180 86 Prague 8, Czech Republic Dept. of Dosimetry and Application of Ionizing Radiation, Czech Technical University, Bøehová 7, 115 19 Prague 1, Czech Republic
F. Spurný
Affiliation:
Dept. of Radiation Dosimetry, Nuclear Physics Institute AS CR, Na Truhláøce 39/64, 180 86 Prague 8, Czech Republic
Get access

Abstract

The spectrometer of linear energy transfer (LET) based on the chemically etched polyallyldiglycolcarbonate (PADC) track-etch detector was developed several years ago in our institute. LET spectra are estimated through track parameters determination carried out by an automatic optical image analyzer LUCIA G. This LET spectrometer enables determining LET of particles approximately from 10 to 700 keV/µm. From the LET spectra, dose characteristics can be calculated. This contribution presents the use of this spectrometer in some applications – studies in 18 MV radiotherapy bremsstrahlung beam and investigation onboard a commercial aircraft during more than 6 months long exposure. The LET spectra obtained in both experiments represent mainly the contribution of secondary or tertiary particles created through the nuclear reactions of photons (bremsstrahlung beam) or neutrons and neutron-like reacting particles (onboard aircraft) in the detector itself or in the surroundings material. The dose due to the photoparticles represents about 3 × 10-4 of the photon dose; for the dose equivalent this value is about 2 × 10-3 Sv/Gy. Regarding the onboard aircraft exposure, the results obtained using LET spectrometer were compared with those achieved with other methods and a reasonable agreement were stated.

Type
Research Article
Copyright
© EDP Sciences, 2006

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Allen, P.D., Chaudhri, M.A. (1988) Photoneutron production and dosimetry in tissue during high energy bremsstrahlung radiotherapy, Rad. Prot. Dos. 23, 341-344. CrossRef
Bartlett, D.T., McAulay, I.R., Schrewe, U.J., Schnuer, K., Menzel, H.G., Bottollier-Depois, J.-F., Dietze, G., Gmür, K., Grillmaier, R.E., Heinrich, W., Lim, T., Lindborg, L., Reitz, G., Schraube, H., Spurný, F., Tommasino, L. (1997) Dosimetry for occupational exposure to cosmic radiation, Rad. Prot. Dos. 70, 395-404. CrossRef
Charvát J. (1986) LET spectrometry with polymer SSNTD’s, Ph.D. thesis, University of Prague, 1986.
Chibani, O., Ma, C.-M.C. (2003) Photonuclear dose calculations for high-energy photon beams from Siemens and Varian linacs, Med. Phys. 30, 1900-2000. CrossRef
IAEA (1987) Absorbed dose determination in photon and electron beams: an International Code of Practice, IAEA Technical Reports Series 277, IAEA, Vienna.
ICRP Publication 60 (1991) 1990 Recommendations of the Internal Commission on Radiological Protection. Oxford, Pergamon Press.
Schraube H., Leuthold G., Heinrich W., Roessler S., Mareš V., Schraube G. (1992) European program package for the calculation of aviation route doses – User’s manual, GSF-National Research Center on Environmental and Health.
Spurný, F., Bednár, J., Johansson, L., Sätherberg, A. (1996a) LET spectra of secondary particles in CR 39 track etch detectors, Radiat. Meas. 26, 645-649. CrossRef
Spurný, F., Johansson, L., Sätherberg, A., Bednáø, J., Turek, K. (1996b) The contribution of secondary heavy particles to the absorbed dose from high-energy photon beams, Phys. Med. Biol. 41, 2643-2656. CrossRef
Spurný, F., Dachev, Ts. (2003) Long-term monitoring of the onboard aircraft exposure level with a Si-diode based spectrometer, Adv. Space Res. 32, 53-58. CrossRef
Spurný, F., Molokanov, A.G., Bamblevski, V.P. (2004) Spectrometry of linear energy transfer, its development and use, Rad. Prot. Dos. 110, 675-679. CrossRef
Spurný, F., Jadrnícková, I., Molokanov, A.G., Bamblevski, V.P. (2005) Upgrading of LET track-etch spectrometer: Calibration and uncertainty analysis, Radiat. Meas. 40, 343-346. CrossRef
Turek, K., Spurný, F., Alberts, W.G. (1993) On the optimization of the etching of CR39 as fast neutron dosimeter, Nucl. Tracks Radiat. Measur. 21, 299-302. CrossRef