A Multiple Proportional 14C Counter System for Milligram-Sized Samples

Published online by Cambridge University Press:  18 July 2016

Gert Hut
Affiliation:
Isotope Physics Laboratory, University of Groningen The Netherlands
Jan Keyser
Affiliation:
Isotope Physics Laboratory, University of Groningen The Netherlands
Stef Wijma
Affiliation:
Isotope Physics Laboratory, University of Groningen The Netherlands
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

A system of 10 proportional 14C counters has been designed. The counters have a volume of 35ml and are operated with CO2 at a pressure of 1 bar and a voltage of 3300V. The amplified signals of the counters are fed through an analog multiplexer to the inputs of three discriminators. The Geiger signal is fed to a separate discriminator. A logic circuit selects from these discriminators the alpha, beta, muon, and purity counts which are then stored in a microcomputer (ITT 2020 with 48K RAM), where the necessary calculations are also performed. The purity of the gas sample is monitored by counting a part of the muon spectrum. The alpha and beta pulses are stored during 95% of the measuring period; the remaining 5% is used for registration of the muon and purity pulses. The overall accuracy of the system for modern carbon samples is 2% after a counting period of 6 days.

Type
V. General Aspects of 14C Technique
Copyright
Copyright © The American Journal of Science 

References

Currie, LA, 1978, Environmental radiocarbon measurements, in : Univ Rochester, p 372390.Google Scholar
Harbottle, G, Sayre, EV, and Stoenner, RW, 1979, Dating of small samples by proportional counting: Science, v 206, p 683685.Google Scholar
Sayre, EV, Harbottle, G, Stoenner, RW, Otlet, RL, and Evans, GV, 1981, Small gas proportional counter for the 14C measurement of very small samples, in : Berlin, IAEA Symposium, p 391407.Google Scholar
Tans, PP and Mook, WG, 1979, Design, construction and calibration of a high accuracy carbon-14 counting set up: Radiocarbon, v 21, p 2240.Google Scholar