INTRODUCTION
The Chronos 14Carbon-Cycle Facility (henceforth Chronos) was opened in 2019 and houses an Ionplus MIni CArbon DAting System (MICADAS) for accelerator mass spectrometry (AMS) analysis along with a fully automated Ionplus AGE3 graphitization System (Turney et al. Reference Turney, Becerra-Valdivia, Sookdeo, Thomas, Palmer, Haines, Cadd, Wacker, Baker, Andersen, Jacobsen, Meredith, Chinu, Bollhalder and Marjo2021). The facility was designed with the aim of better understanding environmental change, specifically in regions of the world where there are limited long-term (>100 years) instrumental records. The combined AGE3 and MICADAS systems allow for high throughput of samples processed from smaller initial masses of material than other AMS systems (Wacker et al. Reference Wacker, Bonani, Friedrich, Hajdas, Kromer, Němec, Ruff, Suter, Synal and Vockenhuber2010a). This is especially important for analyzing wood samples where annual tree-ring measurements of F14C are needed. However, it is possible for tree rings to be very narrow in larger tropical or arid trees, limiting available sample material. Additionally, due to the heavy logging of Australian forests, many of the larger and older trees remain solely in National Parks, where the analysis must be conducted using small-diameter tree cores. Therefore, accessing sufficient sample material to produce a full-sized graphite target (1000 μg of carbon for a MICADAS system) can be challenging. A Gas Interface System (GIS) for the MICADAS is designed for small sample sizes (Wacker et al. Reference Wacker, Fahrni, Hajdas, Molnar, Synal, Szidat and Zhang2013). However, smaller-sized tree-ring samples typically have a large enough amount of carbon that they can be graphitized to produce a viable target, ranging between 200 and 1000 μgC. Additionally, these narrow rings are often contained within sequences of rings suitable for developing full-sized graphite targets prepared on the AGE3 system. For consistency in analysis, sequences should be measured using the same procedures throughout, which is why it is ideal to only use graphitized targets where possible for wood analysis.
Fortunately, there is precedence for analyzing smaller-sized samples using the AGE3 graphitization system and MICADAS AMS by the Bristol Radiocarbon Accelerator Mass Spectrometry (BRAMS) Facility. Samples containing as little as 200 μg of carbon have been reported to produce reliable measurements when analyzed with size-matched standards and blanks (Knowles et al. Reference Knowles, Monaghan and Evershed2019). BRAMS utilizes both an AGE3 graphitization and MICADAS system (Knowles et al. Reference Knowles, Monaghan and Evershed2019), suggesting that it should be possible to analyze targets down to 200 μg of carbon at Chronos. However, Knowles et al. (Reference Knowles, Monaghan and Evershed2019) did not detail the sample type or specific pretreatment used to achieve reliable dates on small samples, other than the need to size-match standards and blanks. To determine the accuracy of F14C measurement of small samples at Chronos, tests were run using modern known-age Australian wood samples of varying size down to 200 μgC targets across the Southern-Hemisphere “bomb peak” of the 1960s (Turney et al. Reference Turney, Palmer, Maslin, Hogg, Fogwill, Southon, Fenwick, Helle, Wilmhurst, McGlone, Bronk Ramsey, Thomas, Lipson, Beaven, Jones, Andrews and Hua2018; Hua et al. Reference Hua, Turnbull, Santos, Rakowski, Ancapichún, De Pol-Holz, Hammer, Lehman, Levin, Miller, Palmer and Turney2021).
METHODS
A 12-mm-diameter tree core (DMA014E) was subsampled for this study. The core is from a subtropical Araucaria cunninghamii tree located in D’Aguilar National Park, Southeast Queensland, Australia and was part of the DMA master tree-ring chronology developed by Haines et al. (Reference Haines, Gadd, Palmer, Olley, Hua and Heijnis2018a). Annual rings for the CE 1962–1970 period were selected as this includes both the sharp rise and fall of the radiocarbon calibration curve surrounding the 1966 bomb spike. This time period spans an extensive range of F14C values so we can best test the sensitivity of the AMS measurements. The nine rings sampled from this core ranged from 1.8 to 10.8 mm wide, with an average width of 6.4 mm. The annual rings were extracted from the core by cutting each ring apart as close to the ring boundaries as possible, while ensuring no overlap between samples. The nine single-ring samples were then chipped into small matchstick-sized pieces perpendicular to the ring-boundaries so that each piece included the entire annual growth period, with approximately 100 mg worth of matchstick pieces used for analysis of each sample. In addition, three 100-mg samples of Manukau kauri (>140,000 years old; Hogg et al. Reference Hogg, Fifield, Turney, Palmer, Galbraith and Baillie2006; Marra et al. Reference Marra, Alloway and Newnham2006) were prepared into matchstick-sized pieces and pretreated as ancient wood background samples alongside the nine annual ring samples.
Chemical pretreatment was undertaken at Chronos to transform all raw wood samples and backgrounds into cellulose following a base-acid-base-acid-bleaching (BABAB) method modified slightly from Turney et al. (Reference Turney, Becerra-Valdivia, Sookdeo, Thomas, Palmer, Haines, Cadd, Wacker, Baker, Andersen, Jacobsen, Meredith, Chinu, Bollhalder and Marjo2021). The samples and ancient wood were first washed three times with 10 mL of acetone, with each wash left for 30 min in a heat block at 55°C. After the three washes, the samples underwent two Milli-Q water rinses. The first base step was then conducted using three 10 mL 1M NaOH base washes for 30 min each in a heat block at 75°C. After the three washes, the samples were again rinsed twice with Milli-Q water. The remaining ABAB steps outlined in Turney et al. (Reference Turney, Becerra-Valdivia, Sookdeo, Thomas, Palmer, Haines, Cadd, Wacker, Baker, Andersen, Jacobsen, Meredith, Chinu, Bollhalder and Marjo2021) were adapted for processing the larger weight of wood samples (i.e., ca. 100 mg) rather than the reported 10–40 mg sample size noted in Turney et al. (Reference Turney, Becerra-Valdivia, Sookdeo, Thomas, Palmer, Haines, Cadd, Wacker, Baker, Andersen, Jacobsen, Meredith, Chinu, Bollhalder and Marjo2021). To ensure all the wood was properly pretreated into cellulose when using these larger weights, the chemical volume was doubled for each chemical processing step, and the bleaching step was extended to 3 hr.
Four different graphite sizes were collected from each of the nine single rings that were pretreated as well as from the three ancient wood background samples. This process was undertaken using a combustion-graphitization system composed of an Elementar vario ISOTOPE select elemental analyzer (EA) and an Ionplus automated graphitization system (AGE3; Wacker et al. Reference Wacker, Němec and Bourquin2010b). Iron is used as a catalyst for graphitization and is dispensed using an Ionplus FED dispenser, providing on average 3.9 mg of iron for the catalyzation of each sample. In the EA-AGE3 system, the pure carbon dioxide captured in the graphitization reactor tube is reduced to graphite at 580°C with hydrogen on iron powder. The graphite sizes analyzed included the typical Ionplus factory specified “full” graphitized sample size of 1000 μgC, and the smaller graphitized sample sizes of 500 μgC, 350 μgC, and 200 μgC desired for this study. The EA-AGE3 instrument method is set in the software program for these instruments but it can be modified to capture specific sizes of carbon for graphitization by incrementally reducing the pressure remaining in the graphitization reactor to a determined setpoint. The Chronos EA-AGE3 instrument method was therefore set at the following levels for this study: 570 mbar for 1000 μgC samples, 290 mbar for 500 μgC, and 200 mbar for 350 μgC. To capture 200 μgC samples accurate weights of cellulose needed to be calculated as the EA-AGE3 instrument method was limited to 200 mbar as the minimum reactor pressure setting at the time this experiment was run (an updated Ionplus EA-AGE3 software version now permits lower reactor pressure settings). Cellulose samples were weighed into tin boats with 2.8–3.2 mg of cellulose used for the 1000 μgC samples, 1.54–1.85 mg of cellulose used for the 500 μgC samples, 0.77–1.05 mg of cellulose used for the 350 μgC samples, and 0.513–0.582 mg of cellulose used for the 200 μgC samples. The best practice for selecting cellulose prescribes using an entire matchstick or a portion of matchstick cut smaller lengthwise and representative of the whole ring. However, when collecting cellulose for the smaller 350 μgC and 200 μgC weights, some of the wider rings had matchsticks that needed to be cut smaller crosswise to obtain the desired weight. Where this was undertaken, sampling was performed to capture as much of the annual representation of the ring as possible.
Standards of oxalic acid (OXII, NIST SRM 4990c), IAEA-C8 (oxalic acid, 15.03 ± 0.07 pMC; Le Clercq et al. Reference Le Clercq, van der Plicht and Gröning1997), and chemical blanks (phthalic anhydride, PhA; SigmaAldrich, PN-320064-10 g) were size matched to produce 1000 μgC, 500 μgC, and 350 μgC of graphite using the mbar levels on the EA-AGE3. All samples, backgrounds, standards, and blanks were pressed pneumatically at 300 MPa into clean aluminium cathodes using an Ionplus press (PSP).
The various-sized graphite samples are produced on a fixed amount of Fe catalyst, resulting in graphitized targets with varying ratios of Fe and C. Three magazines based on sample graphitization size were prepared for routine 14C measurements on the MICADAS; magazine 1 with cathodes containing 1000 μg of carbon, magazine 2 with cathodes containing 500 μg of carbon, and magazine 3 with both the cathodes containing 350 μg of carbon and those containing 200 μg of carbon. The sample graphitization size and the corresponding composition of the graphitized sample produce different C- currents on the MICADAS AMS, thus requiring size-matched samples, backgrounds, standards, and blanks to be measured concurrently in a single magazine of cathodes. The solid graphite targets were measured at moderate C- currents (LE), 20–70 μA. Tuning procedures and operating parameters of the MICADAS are described in Turney et al. (Reference Turney, Becerra-Valdivia, Sookdeo, Thomas, Palmer, Haines, Cadd, Wacker, Baker, Andersen, Jacobsen, Meredith, Chinu, Bollhalder and Marjo2021).
RESULTS AND SUMMARY
The 36 radiocarbon F14C values obtained from subtropical A. cunninghamii tree rings are presented in Table 1 and Figure 1. These values were calibrated using OxCal version 4.4 (Bronk Ramsey Reference Bronk Ramsey2009) on the SH1-2 post-bomb atmospheric curve (Hua et al. Reference Hua, Barbetti and Rakowski2013; Hogg et al. Reference Hogg, Heaton, Hua, Palmer, Turney, Southon, Bayliss, Blackwell, Boswijk, Bronk Ramsey, Petchey, Reimer, Reimer and Wacker2020). As the age of each ring was already known and confirmed through dendrochronological analysis (Haines et al. Reference Haines, Gadd, Palmer, Olley, Hua and Heijnis2018a), only the relevant position on either the rising or falling portion of the SH1-2 post-bomb curve was assessed in this study. These results demonstrate that for all four graphite sizes analyzed here, similar F14C values were obtained, which closely follows the Southern Hemisphere post-bomb radiocarbon calibration curve (Figure 1). Samples that rest on flatter portions of the calibration curve demonstrate that the four target sizes either fall within error of each other or come very close to being within error, as can be noted in the inset of Figure 1. However, samples that fall on the steepest part of the calibration curve (CE 1963 and 1964) demonstrate a larger range of values (Figure 1). This reflects the fact that the carbon fixed across these years represents a 7–8-month accumulation period from mid-August to late March (Haines et al. Reference Haines, Olley, English and Hua2018b), during which there were rapid and extreme changes in atmospheric 14C content (Turney et al. Reference Turney, Palmer, Maslin, Hogg, Fogwill, Southon, Fenwick, Helle, Wilmhurst, McGlone, Bronk Ramsey, Thomas, Lipson, Beaven, Jones, Andrews and Hua2018). This highlights the need for careful “matchstick” sampling across an entire growth period.
* Year assigned to each ring is based on dendrochronological analysis.
† F14C error reported here represents one sigma error.
The precision of the F14C results of the different target sizes (Figure 1, Table 1) was assessed by calculating the smaller targets’ paired differences to that of the “standard” 1000 μgC target. The assumption being the paired differences should ideally be zero. For example, the F14C value for the 500 μgC target of 1962 was subtracted from the F14C value of the 1000 μgC sample for the same year. The results are presented in Table 2 and Figure 2 using a Δ (delta) notation to indicate paired differencing. This approach effectively sets up paired sample t-test comparisons (McDonald Reference McDonald2014). To test if the paired differences were normally distributed, we used a Shapiro-Wilk test of normality where a threshold of 0.05 indicates a normal distribution. This determined that Δ200 was skewed and so the non-parametric Wilcoxon test was therefore applied in this instance instead of the paired sample t-test. Overall, the mean of the paired differences for each of the three smaller-sized samples were not significantly different from the 1000 μgC standard, based on statistical test results (Table 2). This means the smaller weight results were of similar precision to that of the conventional sample weight results. However, the skewed (i.e., biased) values for the 200 μgC targets indicate their use must be treated with caution. Some of this error could be from the incomplete sampling of the entire growth ring with larger rings; a risk not normally an issue with a narrow tree ring that only provides enough material for a ∼200 μgC target. Figure 1 and Table 1 clearly demonstrate that the 200 μgC targets produced F14C values beyond the range of the other samples for wide rings and narrower rings where the entire growth season was sampled, so it would seem other error sources were the cause. It is also known that smaller-sized samples are proportionally affected by any form of chemical or surface contamination during graphitization resulting in a greater possibility of a skewed result (Fahrni et al. Reference Fahrni, Wacker, Synal and Szidat2013). While the use of the Gas Interface System would eliminate some of this contamination due to less handling of a smaller-sized sample, the precision of analysis on the MICADAS is much lower, creating a larger error range for the value provided. Therefore, we believe it is possible to graphitize smaller-sized wood samples as long as greater vigilance is used when analyzing the F14C results.
The capability for analyzing smaller-sized samples was also assessed by looking at the target scatter noted for the backgrounds and standards analyzed in this study at the 1000 μgC, 500 μgC, and 350 μgC sizes producible via carbon capture in the EA-AGE3 system (Table 3). Given the small sample size of this study, the target scatter was calculated for all backgrounds and standards run on the Chronos MICADAS since the facility opened. These values demonstrate that when standard deviation is accounted for, the average F14C values for all standards, blanks, and backgrounds fall within range of each other regardless of target size. The only exception is the 350 μgC sized PhA chemical blanks however, given the small amount of F14C and the fact that any surface contamination on the tin boats used in graphitization, the aluminium cathodes used in the AMS, as well as on the tools used to transfer the graphite will result in proportionally higher backgrounds in both older and smaller samples this value slightly outside of the standard deviation range can be expected. It should also be noted that for the three magazines run for this project, all standards, blanks, and background F14C values fell within the standard deviation of the average values reported in Table 3, with the exception of some of the 350 μgC PhA cathodes which can be explained by the possibility of contamination noted above. The consistent and comparable results between the different sized standards, blanks, and backgrounds demonstrate the ability of the Chronos MICADAS to normalize and correct the values for samples of all sizes discussed in this study. It is also important to ensure that all samples are run against standards and backgrounds which have been size matched to provide the most comparable analysis (Knowles et al. Reference Knowles, Monaghan and Evershed2019).
We also tested the accuracy of our F14C measurements against the calibration curve. Using the program OxCal version 4.4 (Bronk Ramsey Reference Bronk Ramsey2009), each of the four different target weight sequences were calibrated to the Southern Hemisphere region 1-2 (SH1-2) post-bomb atmospheric calibration curve. Our 9-year study period includes, at the start, the steep rise to the bomb peak and this was considered to unduly bias any “wiggle-matching” statistical testing. We therefore selected the shortened window of CE 1966–1970, that is post bomb-spike, to determine how accurate the wiggle-matched sequences (Bronk Ramsey et al. Reference Bronk Ramsey, van der Plicht and Weninger2001; Hogg et al. Reference Hogg, Lowe, Palmer, Boswijk and Bronk Ramsey2011) of values were for the four different target weights. The chi-squared analysis results are 65.2, 111.6, 79.4, and 0.6 for 1000, 500, 350, and 200 μgC targets respectively, with all but the 200 μgC target sequence passing the statistical significance threshold. This suggests all samples from 350 μgC upwards represented accurate F14C measurements in relation to the atmospheric calibration curve.
An additional test of the accuracy of the F14C measurements, specifically on the steep upward climb of the bomb-spike, saw us review the four samples from CE 1964, where the greatest variance in F14C can be found in a single year. These values were plotted in OxCal on the SH1-2 post-bomb atmospheric curve and compared to each other (Figure 3). For the 1000 μgC, 500 μgC, and 350 μgC samples, the probable calendar date on the upwards pre-bomb slope as assigned by OxCal was CE 1964 (Figure 3a–c, respectively). For the 200 μgC sample, the most probable calendar date was given as CE 1963; however, as seen in Figure 3d, the suggestion of CE 1963 is based on a spike in F14C at the end of this year, but the probability distribution extends into CE 1964. Given that most analyses of modern wood samples would involve a mixture of radiocarbon dating and dendrochronological analysis, the ability exists for the correct date to be assigned to this ring with all four graphite sample sizes tested here. However, the result for the 200 μgC sample once again illustrates the need for cautious interpretation.
Overall, these results show that Chronos can produce accurate and precise F14C measurements in small-sized wood samples down to 350 μgC through the use of the standard pretreatment, operation, and analysis procedures. However, great care is needed for samples <350 μgC as there is potential for lower accuracy and precision in replicating atmospheric F14C. This is particularly important for Australian dendrochronological analysis as tree rings in many species from tropical, arid, and some temperate locations may only provide very small amounts of wood per annual ring. When performing radiocarbon analysis on minimal pieces of wood, or any material for that matter, extra care needs to be taken, as any contamination will have a greater effect on a small sample than a larger one. It should also be noted that this study was undertaken on a modern tree core in good condition. While it is likely that the capability of the facility to measure small graphite targets extends back beyond modern wood samples, further study is needed to confirm such analysis is also applicable to much older samples, samples that have been degraded prior to sampling, or are from another dateable material type (e.g., macrofossils, charcoal, peat).