Tourniquets (TQs) save lives. Although military-approved TQs appear more effective than improvised TQs in controlling exsanguinating extremity hemorrhage, their bulk may preclude every day carry (EDC) by civilian lay-providers, limiting availability during emergencies.
The purpose of the current study was to compare the efficacy of three novel commercial TQ designs to a military-approved TQ.
Nine Emergency Medicine residents evaluated four different TQ designs: Gen 7 Combat Application Tourniquet (CAT7; control), Stretch Wrap and Tuck Tourniquet (SWAT-T), Gen 2 Rapid Application Tourniquet System (RATS), and Tourni-Key (TK). Popliteal artery flow cessation was determined using a ZONARE ZS3 ultrasound. Steady state maximal generated force was measured for 30 seconds with a thin-film force sensor.
Success rates for distal arterial flow cessation were 89% CAT7; 67% SWAT-T; 89% RATS; and 78% TK (H 0.89; P = .83). Mean (SD) application times were 10.4 (SD = 1.7) seconds CAT7; 23.1 (SD = 9.0) seconds SWAT-T; 11.1 (SD = 3.8) seconds RATS; and 20.0 (SD = 7.1) seconds TK (F 9.71; P <.001). Steady state maximal forces were 29.9 (SD = 1.2) N CAT7; 23.4 (SD = 0.8) N SWAT-T; 33.0 (SD = 1.3) N RATS; and 41.9 (SD = 1.3) N TK.
All novel TQ systems were non-inferior to the military-approved CAT7. Mean application times were less than 30 seconds for all four designs. The size of these novel TQs may make them more conducive to lay-provider EDC, thereby increasing community resiliency and improving the response to high-threat events.