Divergence slit corrections for Bragg–Brentano diffractometers with rectangular sample surface

Published online by Cambridge University Press:  10 January 2013

Reinhard X. Fischer
Affiliation:
Institut für Geowissenschaften der Universität, D-55099 Mainz, Germany

Abstract

Algorithms are presented to correct intensities affected by divergence slit attenuation in the low 2θ region of Bragg–Brentano powder diffractometers with rectangular sample holders. The intensity loss below a limiting angle 2θ1 occurs when the X-ray beam cross section exceeds the sample surface at decreasing angles. For identification purposes, the observed intensities can be scaled up to values comparable with intensities unaffected by divergence slit effects.