Hostname: page-component-586b7cd67f-rcrh6 Total loading time: 0 Render date: 2024-11-26T01:55:39.648Z Has data issue: false hasContentIssue false

Limiting Factors for Diffraction in the Gandolfi X-Ray Camera

Published online by Cambridge University Press:  19 May 2016

Dennis V. Canfield
Affiliation:
Department of Polymer Science, Box 10076
Daniel A. Sundeen
Affiliation:
Department of Geology, Box 8196, University of Southern Mississippi, Hattiesburg, Mississippi 39406, U.S.A.

Abstract

A discrepancy was discovered between an experimental X-ray diffraction pattern of sodium acetate trihydrate collected with the Gandolfi camera and a calculated pattern based on a single crystal study of the same compound. The experimental d spacings matched those of the calculated pattern with the exception of the strongest reflection d = 3.002 Å (h k l = 40 − 2) which was missing in the experimental pattern. Another experimental Gandolfi pattern obtained from a powdered sample of sodium acetate trihydrate gave excellent agreement with the calculated d spacings and intensities.

The results of our investigation into this discrepancy show that the angle θ, crystal alignment, camera diameter, and the crystal system can affect diffraction in the Gandolfi camera. It is shown that all reciprocal lattice points with θ angles less then 10.08° would diffract in the film plane of the 114.6 mm Gandolfi camera and points with θ angles less then 20.49° would diffract in the film plane for the 57.3 mm camera regardless of the crystal system or alignment. However, reciprocal lattice points aligned along the spindle axis of the 114.6 mm Gandolfi camera having reciprocal lattice points with Z ≥ 0.35 reciprocal lattice units (r.l.u.) and X and Y ≤ Z-0.35 r.l.u. will not diffract into the recording plane of the Gandolfi camera. The reciprocal lattice points aligned along the spindle axis of the 57.3 mm Gandolfi camera with Z ≥ 0.70 r.l.u. and X and Y ≤ Z-0.70 r.l.u. will not diffract into the recording plane of the camera.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1988

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Canfield, D. and De Forest, P. R. (1977). The Use of the Gandolfi Camera as a Screening and Confirmation Tool in the Analysis of Explosives Residues, J. Forensic Sci. 22(2), 337347.CrossRefGoogle Scholar
Santoro, A., and Zocchi, M. (1964). Geometrical Properties of a Four-Circle Neutron Diffractometer for Measuring Intensities at an ‘Optimum’ Azimuth of the Reflecting Planes, Acta Crystallogr. 17, 597602.Google Scholar
Wei, K-T and Ward, D. L. (1977). Sodium Acetate Trihydrate: A Redetermination. Acta Crystallogr. B33, 522526.Google Scholar
Yvon, K., Jeitschko, W., and Parthe, E. (1977). LAZY PULVERIX. Laboratoire De Cristallographie Aux Rayons-X, Université De Genève, 24 Quai Ernest Ansermet, CH 1211 Geneva 4, Switzerland.Google Scholar