In focused ion beam (FIB) fabrication of cross-sectional transmission electron microscopy (X-TEM) specimens, highly accelerated ion beams sometimes cause serious damage. The damage can be induced in both the specimen surface and in the side walls. We used X-TEM observations to investigate the side-wall damage induced by FIB fabrication in crystalline silicon. The damaged layer was found to be about 20 nm thick in the case of 30-keV FIB etching. We tried to reduce the damage by several methods, such as gas-assisted etching (GAE) with iodine, broad argon ion milling and wet etching. The damaged layer was 19 nm for GAE and 12 nm for argon ion milling with a beam current of 70 mA and the tilt angle between the beam and the specimen of 15 degrees. Wet etching using a mixture of nitric and hydrofluoric acid removes most of the damaged layer.