Abstract
The BioCD is a spinning biochip that uses quadrature laser interferometry to detect captured protein on the disc surface. We describe the detection limits of protein binding on this optical biosensor. The fundamental metrology limit is 1 picometer for a single 100-micron diameter spot. Under assay conditions for prostate specific antigen, we can detect 25 pg/ml at 10 assays per disc.