Article contents
A Source of Error in the Absorptiometric Determination of Inorganic and Total Phosphorus in Sea Water
Published online by Cambridge University Press: 11 May 2009
Extract
The results of phosphate determinations in sea water by the Deniges method, using red light, in an absorptiometer, may be in error if the stannous chloride reagent has low HC1 concentration, and contains stannic tin.
The error is in the reagent blank determination, and seems to be caused by colloidal stannic compounds which increase the opacity of the solution. In sea water the effect may be smaller or immeasurable so that results are over-corrected.
When the stannous chloride solution contains about 5% hydrochloric acid, reagent blank figures are low, and independent of stannic tin concentration. This acidity is recommended.
- Type
- Research Article
- Information
- Journal of the Marine Biological Association of the United Kingdom , Volume 28 , Issue 3 , December 1949 , pp. 701 - 705
- Copyright
- Copyright © Marine Biological Association of the United Kingdom 1949
References
REFERENCES
- 8
- Cited by