Introduction
Patient-specific intensity-modulated radiation therapy (IMRT), volumetric-modulated arc therapy (VMAT) and intensity-modulated proton therapy (IMPT) quality assurance protocols should be implemented in the clinic prior to treatment of the patients Reference Agazaryan, Solberg and DeMarco1 . This typically involves the verification of the dose distributions calculated by the treatment planning system by measurement using quality assurance dosimeters and phantoms. The treatment plans with the dose calculation and optimisation parameters are transferred from the patient CT images to the CT images of a quality assurance phantom. Then, the dose distributions are calculated on the quality assurance phantom and delivered on the machine. The measured and calculated dose distributions on the quality assurance phantom are compared to determine the dose difference, distance-to-agreement and gamma index Reference Low, Harms, Mutic and Purdy2 . Most current quality assurance protocols apply a treatment plan optimised and calculated in 3D on a phantom where the calculated and measured dose distribution are performed under static conditions. These quality assurance procedures are usually used to verify dosimetric accuracy of the dose distributions calculated by the treatment planning system. These methods of using stationary phantoms do not accurately consider patient motion that affects patients treated with radiation therapy at the different stages of the procedure including CT imaging and simulation, treatment planning and dose delivery Reference Ali, Alsbou, Algan, Herman and Ahmad3,Reference Bernatowicz, Keall, Mishra, Knopf, Lomax and Kipritidis4 . Furthermore, these protocols fail to consider the effects of intra- and inter-fractional motion on the dose distribution that are actually received by the mobile patients during the radiation therapy course Reference Yang, Geng, Ding, Kong, Cheung, Chiu, Lam, Cheung and Yu5,Reference Giraud, Morvan, Claude, Mornex, Le Pechoux, Bachaud, Boisselier, Beckendorf, Morelle and Carrere6 .
Patient motion can be caused by the skeletal-muscular, cardiac, gastrointestinal and respiratory systems. Respiratory motion particularly remains one of the forefront challenges towards motion management in radiation therapy because breathing can affect different treatment sites such as lungs, oesophagus, liver, pancreas, breast and prostate Reference Keall, Mageras, Balter, Emery, Forster, Jiang, Kapatoes, Low, Murphy, Murray, Ramsey, Van Herk, Vedam, Wong and Yorke7 . Patient motion can lead to large deviations in the dose coverage of the tumour and sparing of the normal tissue expected from the optimised treatment plans Reference Mageras and Yorke8 . It is important to consider patient motion when performing quality assurance and dose verification on patient treatment plans to accurately simulate patient conditions during therapy. Many studies have been conducted that attempt to introduce an anthropomorphic motion phantoms which have the capabilities of modelling the complex patterns of patient’s respiratory motion Reference Dunn, Kron, Johnston, McDermott, Taylor, Callahan and Franich9–Reference Shiinoki, Fujii, Fujimoto, Yuasa and Sera11 . The aims of this study include investigating the effects of respiratory motion on the delivered dose distributions for different treatment plans that include conformal, IMRT, VMAT and IMPT produced for the treatment of different sites to quantify underdosing of the tumour and overdosing of the normal surrounding tissues. Furthermore, the effects of variations in motion amplitude and frequency on the dose distributions that were optimised in 3D on static CT images using current dose calculation algorithms used for treatment planning were quantified.
Materials and Methods
Treatment planning
In this study, various 3D-optimised treatment plans using conformal beam, IMRT, VMAT and IMPT were delivered to a 2D detectors placed on a motion platform (Standard Imaging Inc, Middleton, WI). The mobile platform moved in one dimension sinusoidally producing different motion patterns with adjustable amplitude and frequency. The MapCheck2 phantom (Sun Nuclear, Melbourne, FL) made from multiple array diodes was used to measure the dose distributions for the conformal, IMRT and VMAT treatment plans delivered by a Varian Trilogy and Truebeam-STX Linear Accelerators (Varian Medical Systems, Inc., Palo Alto, CA) Reference Jursinic, Sharma and Reuter12 . The OCTAVIUS phantom (PTW, Freiburg, GmbH) made from 2D multiple array ionisation chambers was used to measure the dose distributions for the IMPT treatment plans delivered by the Mevion-S250i HYPERSCAN proton system (MEVION Medical Systems, Littleton, MA, USA) Reference Decabooter, Roijen, Martens, Unipan, Bosmans and Vilches-Freixas13,Reference Urso, Lorusso, Marzoli, Corletto, Imperiale, Pepe and Bianchi14 . The treatment plans 10 in total represented different treatment sites of head and neck, thorax, abdomen, and pelvis. The plans were first delivered for a static condition which was used as a reference representing typical conditions for patient-specific quality assurance testing. Then, it was followed by dose delivery with the mobile platform which moved at amplitudes in the range 0–35 mm using frequencies of 0·2–0·33 Hz. The mobile phantom was placed on the treatment couch where different sinusoidal motion patterns were applied in the superior–inferior direction during beam delivery as shown in Figure 1.
The 2D dose distributions from the photon plans were measured with a multiple-diode array MapCheck2 phantom which consisted of 1,527 diodes arranged in arrays that are offset by 5 mm as shown in Figure 1a Reference Rinaldin, Perna, Agnello, Pallazzi, Cattaneo, Fiorino and Calandrino15 . The overall effective detection area of the detector is 32×26 cm2 with an array geometry of 10 mm detector spacing parallel to the X- and Y-axis and row spacing of 5 mm. The minimal distance between the centre of the detectors was uniform throughout the array at 7·07 mm. The active single detector area and volume were 0·64 mm2 and 0·019 mm3, respectively. The diodes were located at 12 mm depth within the phantom, which was equivalent to 20 mm of solid water build-up. The diodes were encapsulated with an additional 3 cm of solid water on the top to provide sufficient depth for dose measurement mimicking tumour depths and eliminating dose artefacts in the build-up region and electron contaminations. Another 3 cm phantom was layered beneath the detector to eliminate backscatter issues. The dose distributions for the IMPT plans were measured using the OCTAVIUS Detector 729XDR Reference Jeevanandam, Agnew, Irvine and McGarry16,Reference Stelljes, Harmeyer, Reuter, Looe, Chofor, Harder and Poppe17 , which wss made from a multiple-array ion chamber matrix commonly used for quality control and patient plan verification using protons or heavy ions. This detector, shown in Figure 1b, consisted of 729 vented plane-parallel ion chambers arranged in an effective detection area of 27×27 cm2. The size of each vented plane-parallel ion chamber is 5×5×3 mm3, and they were aligned in rows and columns with a separation of 10 mm. In contrast to the diodes, ion chambers were more resilient to the damage from the neutrons that were produced from the interaction of the proton beam. The treatment plans were mapped on the OCTAVIUS detector encapsulated in phantom with different thicknesses to simulate the tumour depth in the patient-specific plans for proper dose verification of the spread-out Bragg peak, the proximal and distal ranges of the proton beams calculated by the treatment planning system. The 2D dose distributions from the photon and proton plans measured with mobile phantom were compared with the corresponding dose distributions measured with the stationary phantom. Different dosimetric parameters that included dose difference, dose deficit in the tumour, dose spread-out and the gamma index were calculated to quantify the effect of phantom motion on the dose distributions.
Results
Dose distributions dependence on motion amplitude
Conformal plans with open fields
The variations of the 2D dose distributions from open fields with motion amplitudes ranging 0–35 mm at a constant cycle interval of 3 s is shown in Figure 2. As the motion amplitude increased, the spread-out of the dose distributions at the beam edge increased. The extent of the dose spread-out correlated with the motion amplitude, with the edge blurring equating to half the motion amplitude from each side of the dose distributions along the direction of motion.
Figures 3a shows the dose profiles normalised to the central axis dose (100%) for the 10×10 cm2 open field along the direction of motion (Y-axis). While Figure 3b shows the percent dose difference of the mobile dose distribution relative to the static dose distribution for the motion amplitudes 5–35 mm. For the static condition, the penumbra showed a sharp dose fall-off at the beam edge indicating the dose distributions from conformal fields. As the motion amplitude increased, the penumbra experienced a spread-out about the 50% isodose line that caused the beam edge to broaden. With increasing motion amplitude, the region inside the treatment volume experienced underdosing, while the region outside of the treatment volume experienced overdosing. The spread-out of this region on one edge of the dose profile, from the overdosing outside the volume and underdosing inside the volume, was approximately half the size of the motion amplitude. The magnitudes of the overdosing and underdosing regions were further investigated quantitatively in Figure 3, with the smallest motion amplitude of 5 mm yielded an overdosing of 40% outside the treatment volume and underdosing of 20% inside the treatment volume. While the largest motion amplitude of 35 mm produced an overdose of approximately 75% in the surrounding normal tissue and an underdosing of nearly 60% within the treatment volume. For this conformal plan that delivered a flat uniform dose, the central region of the beam remained unchanged for all mobile conditions for the range of motion amplitudes (0–35 mm) used in this study. This may change for small fields with their length equal to the motion amplitudes.
IMRT-optimised plans
Figure 4(a, b, c, d) shows the 2D dose distributions for IMRT plans for different treatment sites that include: (a) head and neck, (b) lung, (c) abdomen and (d) pelvis. The dose distributions were displayed with increasing motion amplitude, beginning with the static condition and then increasing in motion amplitude from 5 mm to 35 mm. Similar to the open field, the dose distributions were spread out at the beam edge along the direction of motion which increased as the motion amplitude increased.
Similar behaviour was noted in the normalised dose profiles of the different IMRT plans for the different treatment sites as shown in Figure 5a. The dose distributions spread out with an overdosing outside the treatment volume and underdosing inside the treatment volume. The spread-out in the penumbra regions increased as the motion amplitude increased; however, it depended on the local dose peaks of each IMRT dose profile for the different treatment plans. Furthermore, the motion artefacts also induced variations in the local doses and the dose distributions inside the PTV. The right column in Figure 5(b, d, f, h) showed percentage dose differences of underdosing in the tumour and overdosing in the surrounding tissues relative to the static dose distributions which were optimised without consideration of patient motion. The spread-out of the dose distributions from the IMRT plans produced similar effects with increasing severity as the amplitude increased. In most cases, the smallest motion amplitude of 5 mm used in this study resulted underdosing inside the treatment volume by 5–20% and overdosing of 20–30% in the surrounding tissues. While the largest motion amplitude of 35 mm resulted in underdosing from 40 to 60% inside the treatment volume and overdosing from 60–80% in the nearby tissues.
VMAT-optimised plans
Figure 6 shows the 2D dose distributions obtained from a (a) Head and Neck, (b) Lung and (c) Abdomen VMAT plans delivered by a Varian TrueBeam linear accelerator. The measured dose distributions for the VMAT plans showed similar behaviour to those obtained for the IMRT plans where the spread-out at the beam edges increased along the direction of the amotion with increasing motion amplitude from 0 to 35 mm as shown in Figure 7. However, the dose deviations from the motion artefacts were more pronounced inside the tumour as shown in figure. This resulted from several factors that include increased intensity modulation, rotation of the gantry around the patient and cross-talking between the phantom motion, MLC-motion and gantry motion in the VMAT plans.
IMPT-optimised plans
Figure 8 shows the 2D dose distributions of a left breast treatment plan using the MEVION-S250i proton system for motion amplitudes ranging 0–35 mm. The dose spread-out at the beam edge increased with increasing motion amplitude similar to the IMRT and VMAT photon plans. However, the dose distributions from the spot scanning proton beam within the tumour were distorted substantially by motion. The flat dose regions and local dose distributions in the proton plans changed within the tumour depending on the motion amplitude due to substantial cross-talking effect of the scanning proton spot beam with the phantom motion. Figure 8a shows the normalised percentage dose profiles normalised to the central axis dose acquired with the static phantom. The motion effects close to the penumbra for the proton plan did not follow similar behaviour as seen in the previous IMRT and VMAT treatment plans. While the photon dose distributions showed spread-out for all motion amplitudes around the 50% dose point along the direction of motion, there was no systematic blurring of the dose distributions around the 50% point for the dose distributions from the spot scanning proton beams. In addition, the central region of the dose distributions of the proton plans deviated significantly with the different motion amplitudes, which resulted displacement of the hot spots within the treatment volume. The central region of the fields showed non-uniform percent differences between each motion amplitude as shown in Figure 8c. While the dose close to the central axis in the IMRT and VMAT plans deviated from 0–5%, the IMPT plans central doses deviated by up to 50% for large motion amplitudes as shown in Figure 9a,b. The large deviations in the dose distribution close to the penumbra and near the central axis regions resulted from the complex interplay effects between the proton spot scanning beam and the phantom motion. As the proton delivery system used a scanning pencil beam and not a flat field of variable intensity produced by the linear accelerators, the dose delivered by the spot scanning pencil beam would vary depending on its overlap with the phantom motion cycle.
Dose distributions dependence on cycle interval
Figure 10 shows the variations of the dose distributions delivered motion cycle intervals of 2 sec/cycle and 6 sec/cycle. Small changes in the dose profiles were measured with two motion cycle intervals for both motion amplitudes of 5 and 35 mm. The percentage dose deviations between the two cycle intervals were within 3% as shown in Figure 10c. The dose differences were nearly 5% at the field edge due to spatial mismatching of the dose profiles because of the uncertainty in the detector positioning during measurement.
Quality assurance testing: Gamma Index
Figure 11 shows the gamma index passing rates with a criterion of 3%/3 mm calculated for different IMRT and VMAT plans with motion amplitudes 3–35 mm that were used on the mobile phantom. The gamma test was calculated using the dose distributions recorded with the different motion amplitudes relative to that recorded under static conditions. The gamma passing rate decreased as the motion amplitude increased because of the deviations introduced by motion artefacts. The decline in the gamma passing rates were more significant for the IMRT plans compared to that of the VMAT plans which was due to the shorter delivery time of the VMAT plans that resulted less motion artefacts.
Discussion
The combination of the 2D detectors including the MapCheck2 or OCTAVIUS with the mobile phantom set-up provided this 4D dosimeteric system that was used to quantify the discrepancies induced by respiratory motion on conformal, IMRT, VMAT and IMPT plans. These dose discrepancies mimic the deviations between the dose distributions obtained from 3D-optimised plans using static CT imaging and the actual delivered dose distributions with patient breathing. The dose distributions of the mobile phantom spread outside the tumour volume with reproducible characteristics of underdosing of the tumour and overdosing of the surrounding area as indicated by other studies Reference Sarudis, Karlsson, Nyman and Bäck18,Reference Yoshioka, Akino, Shiomi, Hirata, Kai, Ogawa and Koizumi19 . These artefacts are not accounted for in treatment plans that do not consider the respiratory motion in the treatment planning and dose delivery processes. The 4D dosimeter developed in this work provides a novel tool to quantify the dose deviation induced by respiratory motion compared with previous studies Reference Choi, Sung, Dong, Shin, Kim and Lim20–Reference Marants, Vandervoort and Cygler23 . Innovative dosimetric tools have to be developed to measure motion artefact, and new approaches are needed to account for these dose artefacts in the treatment planning, quality assurance and dose delivery techniques to ensure accurate dose delivery.
The conformal plans showed systematic spread-out of the dose distributions around the 50% isodose line along the direction of motion with a flat dose near the central axis as indicated by previous studies Reference Gauer, Sothmann, Blanck, Petersen and Werner24,Reference Sothmann, Gauer and Werner25 . However, the IMRT and VMAT plans showed large variations in the dose distributions overall including the tumour volume because of shifts in the local doses due to the phantom motion. The features of the dose distributions were consistent in the treatment plans for IMRT and VMAT with photon beams. However, different patterns were produced noticeably in the proton IMPT plans by motion artefacts as reported in previous studies Reference Mastella, Molinelli, Pella, Vai, Maestri, Vitolo, Baroni, Valvo and Ciocca26,Reference Pfeiler, Bäumer, Engwall, Geismar, Spaan and Timmermann27 . While the conformal, IMRT and VMAT spread-out systematically around a single point where the 50% isodose line with increasing motion amplitude, the dose patterns were not reliably reproducible or predictable in the IMPT plans. In addition, the location of the peak doses or hot spots within the central region of the treatment volume varied significantly in the measured dose distribution of the proton plans compared to photon plans. This resulted because the dose from the photon beams were delivered uniformly over the tumour volume and the overall dose in the central region averaged out, while in the IMPT plan the dose was delivered by a scanning pencil beam locally in one region and gradually until the dose coverage of the entire volume was achieved. The proton dose delivery method was subject to larger interplay effects with the mobile phantom, where the location of the phantom voxels changed constantly during dose delivery with spot scanning proton beam. This resulted in large dose variations ranging from 10 to 20% in the central region of the dose distribution in the proton plan compared to just 0–5% the photon dose distributions due to motion artefacts. One limitation of this study is the use of treatment plans with intermediate to large PTV, where the treatment beams were larger than the motion amplitudes (5–35 mm) used. We expect that the use of treatment plans for small targets comparable or smaller than the motion amplitudes will induce more substantial artefacts in the dose distributions measured with this 4D dosimeter.
While the variations in motion amplitude resulted in large deviations in the dose distributions with the mobile phantom relative to stationary, the variations in the motion frequency did not result in substantial dose effects. This can be explained by dose averaging effects due to elongated dose delivery on the machine (nearly 60 sec per beam) in comparison with the motion frequency (3–4 sec cycles), where the dose delivery occurred over multiple motion cycles and the dose rates were too low for the motion frequency to have considerable effects. Another limitation of this study is the use of large cumulative doses with relatively low dose rates. The motion frequency might strongly affect the dose distributions delivered with high dose rates over short time periods comparable to the periods of the respiratory cycle. Further investigation is required using treatment plans with higher dose rates to quantify the effects that motion frequency may have on dose distributions in future studies. The gamma index analysis of IMRT and VMAT treatment plans showed a steady decline in passing rates as the range of motion applied on the phantom increased, indicating the importance of motion management techniques for accurate dose delivery and quality assurance of patient-specific treatment plans to consider motion artefacts. The dose spread-out for the photon plans was reproducible and could be modelled for the cyclic motion of the mobile phantom system. These features can be employed to develop motion management techniques for adaptive radiation therapy and dose calculation algorithms with capabilities to perform 4D optimisation of the treatment plans that can account for motion in real time with new innovative techniques that can be developed in the future Reference Wu, Li, Wu and Yin28,Reference Yan, Vicini, Wong and Martinez29 .
Conclusion
Phantom motion caused significant variations in the dose distributions measured for conformal, IMRT, VMAT and IMPT plans. The motion artefact increased with increasing motion amplitude, and it was more prominent in proton plans compared to photon plans. The dose distributions in all cases showed spread-out effects at the beam edge and changes in the local dose distribution in the tumour which were simulated for varying motion amplitudes and frequencies. The measured dose distributions with this mobile phantom provided a quantitative characterisation of the changes in the dose distributions induced by sinusoidal motion that simulated respiratory motion of cancer patients treated with radiation therapy. The 4D dosimeter developed and tested in this work provided a prototype of a noble quality assurance system that can be used to quantify dose deviations induced by respiratory motion that were present during dose delivery for actual patients. It is also a superior tool that can be used for quality assurance and dose verification of futuristic 4D dose optimisation and calculation algorithms that aim to compensate for respiratory motion effects in order to provide conformal dose distributions and achieve the goals of adaptive radiation therapy.