Hostname: page-component-cd9895bd7-8ctnn Total loading time: 0 Render date: 2024-12-22T10:45:58.310Z Has data issue: false hasContentIssue false

Comparative dosimetry of brachytherapy treatment planning between a volume-based plan by CT and a point-based plan by TAUS in CT datasets for brachytherapy

Published online by Cambridge University Press:  26 November 2021

Chaiyaporn Pintakham
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand Graduate School, Chiang Mai University, Chiang Mai, Thailand
Ekkasit Tharavichitkul*
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand Northern Thailand Radiation Oncology Group, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Somsak Wanwilairat
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Wannapha Nobnop
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand Northern Thailand Radiation Oncology Group, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
*
Author for Correspondence: Ekkasit Tharavichitkul, MD, The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand. Tel: +66-53-935456. E-mail: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Aim:

To evaluate comparative dosimetry of brachytherapy treatment planning between a volume-based plan by computed tomography (CT) and a point-based plan by transabdominal ultrasound (TAUS) in CT datasets for brachytherapy.

Materials and methods:

From 2019 to 2021, 59 different datasets of CT images were collected from 38 patients treated by intracavitary brachytherapy with tandem ovoid or tandem ring applicators. At that time, TAUS was performed to prevent uterine perforation and to evaluate topography of the cervix during application. In volume-based planning by CT, the target dose was used to keep the dose at 90% of high-risk clinical target volume (HR-CTV), to give a dose of at least 7Gy, while in the point-based plan by TAUS, the target dose was used to keep the minimum dose to eight cervix reference points (measured by TAUS), to give a dose of at least 7Gy. The doses to targets and organs at risk were evaluated and compared between volume-based planning by CT and the point-based plan by TAUS.

Results:

Of 59 fractions, a tandem ovoid applicator was used in 48 fractions (81·3%). In the volume-based plan by CT, the mean doses to HR-CTV(D90), intermediate-risk clinical target volume (IR-CTV)(D90), bladder(D2cc), rectum(D2cc) and sigmoid colon(D2cc) were 7·0, 3·9, 4·9, 2·9 and 3·3 Gy, respectively, while in the point-based plan by TAUS, the mean doses to HR-CTV(D90), IR-CTV(D90), bladder(D2cc), rectum(D2cc) and sigmoid colon(D2cc) were 8·2, 4·6, 5·9, 3·4 and 3·9 Gy, respectively. The percentages of mean dose differences between TAUS and CT of HR-CTV(D90), IR-CTV(D90), bladder(D2cc), rectum(D2cc) and sigmoid colon(D2cc) were 17·7, 19·5, 20·5, 19·5, 21·3 and 19·8%, respectively. With the target dose to the point-based plan by TAUS (7 Gy to the cervix reference points), this was close to D98 of HR-CTV with a mean percentage of difference of 0·6%.

Findings:

The point-based plan by TAUS showed higher values to targets and organs at risk than the volume-based plan by CT. With the point-based plan by TAUS, it was close to D98 of HR-CTV.

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press

Introduction

Cervical cancer is one of the most common female cancers worldwide. Reference Arbyn, Weiderpass and Bruni1 In 2017, new cases of cervical cancer in Thailand were more than 8,000· Reference Phoolcharoen, Kantathavorn and Sricharunrat2 The treatments for cervical cancer are surgery, radiotherapy or systemic treatment. For radical radiotherapy with chemotherapy, the combination of external beam radiotherapy (45–50·4 Gy in 23–28 fractions) and brachytherapy (3–5 fractions) is the standard treatment for locally advanced disease to give a dose of at least 80 Gy in EQD210 to the target. Reference Viswanathan, Beriwal and Jennifer3

Brachytherapy has been used for a long time to escalate the dose to the cervix to achieve the curative goal. 4 Nowadays, brachytherapy has been transformed from a point-based plan (2D) to a volume-based plan (3D), according to many promising publications from the GEC-ESTRO working group. In the volume-based plan, or image-guided adaptive brachytherapy (IGABT), magnetic resonance imaging (MRI) is the standard imaging technique as it provides the best soft tissue discrimination. Reference Haie-Meder, Pötter and Van Limbergen5,Reference Pötter, Haie-Meder and Van Limbergen6 Many international publications have reported promising results from MRI-guided brachytherapy. Reference Pötter, Dimopoulos and Georg7Reference Pötter, Tanderup and Schmid15

Computed tomography (CT)-based IGABT was developed in practice because of the difficulties in using MRI for brachytherapy. Although CT showed poorer tissue discrimination than MRI, CT can support treatment in the volume-based plan as reported in many international publications. Reference Hegazy, Pötter and Kirisits16,Reference Hellebust17 Promising results of CT-based brachytherapy for cervical cancer have been published internationally. Reference Tan, Coles and Hart18Reference Ohno, Noda and Okonogi24 Another method, point-based adaptive planning by transabdominal ultrasound (TAUS), was developed for use in brachytherapy. A few studies have reported the use of a TAUS-based plan and have shown a clinical benefit. Reference Tharavichitkul, Chakrabandhu and Wanwilairat21,Reference van Dyk and Bernshaw25Reference Tharavichitkul, Chakrabandhu and Klunklin31

At our Institute, transformation from 2D (point) to 3D (volume) began in 2008 with CT-based brachytherapy. Unfortunately, the total transformation could not be performed immediately due to our workloads and equipment. In 2011, ultrasound for brachytherapy was installed at our Institute. In 2012, a project to improve the quality of 2D planning using TAUS-guided brachytherapy as 2·5D planning was initiated. From these projects, intermediate-term results were reported in 2013 and 2018, respectively. Reference Tharavichitkul, Chakrabandhu and Wanwilairat21,Reference Tharavichitkul, Chakrabandhu and Klunklin31 The total transformation to a volume-based approach (CT-based plan) succeeded in 2019 when a 4-slice CT instrument was installed in our brachytherapy unit. After that point, our clinical practice changed to volume-based treatment by CT, and TAUS was utilised to prevent uterine perforation and to support the CT-based contouring. However, dose evaluation or correlation between these two approaches was not performed.

With this in mind, we performed the present study to evaluate comparative dosimetry of brachytherapy treatment planning between a volume-based plan by CT, and a point-based plan by TAUS, in CT datasets for brachytherapy.

Materials and Methods

This study was a case–control study approved by the Institutional Review Board of the Faculty of Medicine, Chiang Mai University, with the code RAD-2563–07479.

CT datasets

From January 2019 to January 2021, 59 CT image datasets with 3 mm slice thickness (Alexion, Toshiba Medical Systems, Tochigi, Japan) were collected from 38 patients with locally advanced cervical cancer treated by intracavitary CT-based brachytherapy. High-risk clinical target volume (HR-CTV), intermediate-risk clinical target volume (IR-CTV) and OARs were contoured by experienced radiation oncologists during treatment according to GEC-ESTRO recommendations and the work of Viswanathan et al. Reference Haie-Meder, Pötter and Van Limbergen5,Reference Pötter, Haie-Meder and Van Limbergen6,Reference Viswanathan, Erickson and Gaffney32 All fractions were treated by CT-based brachytherapy (volume-based approach) to give a dose of at least 7 Gy to D90 of HR-CTV. In these fractions, TAUS (Flex Focus 400, BK medical, Harlev, Denmark) was performed by a radiation oncologist to prevent uterine perforation during applicator placement.

Study methods

To perform this study, all CT datasets and structures were imported into Oncentra Master Plan version 4.5.3 (Elekta Brachytherapy, Veenendaal, the Netherlands). Applicator reconstruction was performed by manual reconstruction or library-based registration according to the applicator type. Two planning approaches were performed, and sources were applied with the same dwell positions in both approaches. The dwell times in each dwell position were optimised as the planning aim of each approach.

Approach 1: volume approach

In the volume-based plan by CT, the target of the treatment was HR-CTV, and planning aims were performed to keep the dose at 90% of HR-CTV to give a dose of at least 7 Gy.

Approach 2: point approach

In the point-based plan by TAUS, eight measurements of cervix dimensions (from intrauterine tandem to uterine wall) were performed at the level of the cervical os, and 2 cm cranially to the cervical os, based on previous work by van Dyk et al. and Tharavichitkul et al. Reference van Dyk and Bernshaw25,Reference Tharavichitkul, Tippanya and Jayavasti28 Eight cervix reference points were generated from cervix dimensions in sagittal and axial views. Planning aims were performed to keep the minimum dose to the cervix reference points to give a dose of at least 7 Gy.

In both approaches, doses at selected parameters were recorded as follows. 4

$${\rm{COIN\; = \;}}{{{{{\rm{V}}_{{\rm{ref,TV}}}}}\over{{\rm{TV}}}}}{\rm{ \times }}{{{{{\rm{V}}_{{\rm{ref,TV}}}}}\over{{{\rm{V}}}_{{\rm{ref}}}}}}$$
$${\rm{DHI\; = \;}}{{{{\rm{(TV}}{{\rm{D}}_{{\rm{ref\;}}}}{\rm{ - \;TV1}}{\rm{.5}}{{\rm{D}}_{{\rm{ref}}}}{\rm{)}}}\over{{\rm{TV}}{{\rm{D}}_{{\rm{ref}}}}}}}$$

Images of the entire study process, measurements of eight cervix dimensions by TAUS, targets (HR-CTV & eight cervix reference points) and isodose distributions of both approaches are shown in Figures 14.

Figure 1. Study process.

Figure 2. Measurements of eight cervix dimensions by TAUS: (a) sagittal view of TAUS, (b) axial view of TAUS at the cervical os and (c) axial view of TAUS at 2 cm cranially from the cervical os.

Figure 3. HR-CTV, IR-CTV and eight cervix reference points (A1-A4 & L1-L4) in the CT datasets.

Figure 4. Isodose distributions of (a) volume-based plan by CT and (b) point-based plan by TAUS.

Statistical analysis

All dose parameters were compared for the volume-based plan by CT and the point-based plan by TAUS. A Wilcoxon sign-ranked test was used to evaluate the comparison.

Percentage dose difference was evaluated from the dose difference to the targets or OARs in the volume-based plan by CT versus the point-based plan by TAUS. Percentage of relative dose difference was evaluated as the difference of the goal dose to the cervix reference points (at least 7 Gy) by TAUS to the dose to HR-CTV (contoured by CT) in terms of D50, D90, D95, D98 and D100. (Figure 5) IBM SPSS version 22 (IBM SPSS, Chicago, IL, USA) was used to evaluate the results.

Figure 5. Percentage dose difference and relative dose difference formulae.

Results

Of 59 fractions, a tandem ovoid applicator was used in 48 fractions. The mean volume of HR-CTV was 15·8 cc. Table 1 shows the characteristics of the studied datasets.

Table 1. Characteristics data

Note: HR-CTV = high-risk clinical target volume, IR-CTV = intermediate-risk clinical target volume.

In the targets, generally, the point-based plan by TAUS yielded a significantly higher dose to D90 of HR-CTV, D90 of IR-CTV, point A, TRAK, DHI and COIN. The percentage of dose difference between the point-based plan by TAUS and the volume-based plan by CT ranged from 17·0 to 21·2% (see Table 2). Interestingly, when we further evaluated the percentage of relative dose difference between the planning aim of the point-based plan by TAUS (at least 7 Gy at cervix reference points) to volume parameters of HR-CTV (from D50 to D100), the TAUS-based planning aim was close to D98 of HR-CTV with a mean of 0·6%. Tables 2 and 3 show the dose parameters, % of dose difference and % of relative dose difference, for the volume-based plan by CT and the point-based plan by TAUS relative to target dose. In normal tissues, the point-based plan by TAUS yielded a significantly higher dose to the bladder, rectum, sigmoid, ICRU points and vaginal points. The percentage of dose difference between the point-based plan by TAUS and the volume-based plan by CT ranged from 17·7 to 24·2% (see Table 4). Table 4 shows the dose parameters and % of dose difference for the volume-based plan by CT and the point-based plan by TAUS in OARs.

Table 2. Doses to target for volume-based plan by CT and point-based plan by TAUS plans

Note: COIN = conformity index, CT = computed Tomography, D100 = the minimum dose covering 100% of volume, D98 = the minimum dose covering 98% of volume, D95 = the minimum dose covering 95% of volume, D90 = the minimum dose covering 90% of volume, D50 = the minimum dose covering 50% of volume, DHI = dose homogeneity index, HR-CTV = high-risk clinical target volume, IR-CTV = intermediate-risk clinical target volume, TAUS = transabdominal ultrasound, TRAK = total reference air kerma, SD = standard deviation.

* Significant difference.

Table 3. Percentage relative dose difference of HR-CTV in point-based plan by TAUS plan

Note: D100 = the minimum dose covering 100% of volume, D98 = the minimum dose covering 98% of volume, D95 = the minimum dose covering 95% of volume, D90 = the minimum dose covering 90% of volume, D50 = the minimum dose covering 50% of volume, HR-CTV = high-risk clinical target volume, SD = standard deviation.

Table 4. Doses to organs at risk for volume-based plan by CT and point-based plan by TAUS plans.

Note: CT = computed tomography, D0.1cc = dose at 0.1 cc of volume, D1cc = dose at 1 cc of volume, D2cc = dose at 2 cc of volume, ICRU = The International Commission on Radiation Units and Measurements, PIBS = Posterior-Inferior Border of Symphysis, TAUS = transabdominal ultrasound, SD = standard deviation.

* Significant difference.

Discussion

In clinical practice, although the trend in brachytherapy has moved to volume-based planning, some centres still perform point-based planning. Here, TAUS supports reduction of overdose in the bladder and rectal points in comparison to the conventional prescription. Reference Tharavichitkul, Chakrabandhu and Wanwilairat21,Reference Tharavichitkul, Tippanya and Jayavasti28 Our results showed that the point-based prescription to the cervix reference points by TAUS caused a higher dose in comparison to the volume-based prescription to D90 of HR-CTV. Moreover, when we explored the 7 Gy prescription to the cervix reference points, we found the dose to be nearly 7 Gy, close to D98 of HR-CTV. This may support the finding that when we treat by cervix reference points measured by TAUS, the cumulative dose from EBRT plus brachytherapy should focus to at least 75 Gy in EQD210 (the planning aim of D98 of HR-CTV published in the EMBRACE II protocol). Reference Pötter, Tanderup and Kirisits35 For example, if we treated EBRT 46 Gy in 23 fractions, the BT dose by ≥ 29 Gy in EQD210 (such as 21 Gy in 3 fractions or 24 Gy in 4 fractions) will be prescribed to the cervix reference points to get the cumulative dose by at least 75 Gy in EQD210.

This study has some limitations. First, as mentioned above, the cumulative dose of EBRT and BT doses were not evaluated. Second, this study is a case–control study in which we performed two planned approaches on CT datasets of previous treatment. Third, only a single fraction was evaluated. Fourth, these findings were obtained using intracavitary brachytherapy only. Finally, MRI was not performed in these patients. Although there were some limitations, the results of our study revealed that the doses to targets and OARs when we prescribed 7 Gy to the cervix reference points by TAUS were higher than the volume-based prescription to D90 of HR-CTV with an average 20% difference. When we evaluated the correlation of planning aims of 7 Gy to the cervix reference points to the volume of HR-CTV, it showed close correlation to D98 of HR-CTV with a mean percentage of difference of 0·6%.

Based on the literature, three types of dose comparison studies were performed. First, comparative studies performed to compare between point A and image-guided treatment in the same type of image showed that the image-guided plan improved the dose to targets or reduced overdosage to normal tissues (in comparison to the point A plan). Reference Tharavichitkul, Tippanya and Jayavasti28,Reference De Brabandere, Mousa and Nulens36Reference Tharavichitkul, Mayurasakorn and Lorvidhaya39 Second, comparative studies were performed to compare MRI- versus CT-based contouring. All studies showed that CT-based contouring gave an overestimation and a higher dose to normal tissues in comparison to MRI contouring. Reference Viswanathan, Dimopoulos, Kirisits, Berger and Pötter40,Reference Eskander, Scanderbeg, Saenz, Brown and Yashar41 Third, comparative studies were performed to compare MRI- versus TAUS-based planning. A report by van Dyk et al. has shown the methods to be comparable. Reference van Dyk, Narayan and Fisher26 Table 5 shows selected studies of dose comparisons. Our study is categorised in the first type, but this is modified to compare between cervix reference points versus D90 of HR-CTV in CT images. Ultrasound showed correlation to MRI in dimensional evaluation of the cervix for brachytherapy. The comparison studies mostly compared measurements between TAUS and MRI and showed similar results in measurement. Reference van Dyk and Bernshaw25,Reference Mahantshetty, Khanna and Swamidas42 Moreover, the use of transrectal ultrasound (TRUS)-guided CT contouring showed correlation to MRI in HR-CTV evaluation. Reference Schmid, Nesvacil and Pötter43,Reference Mahantshetty, Khadanga and Gudi44

Table 5. Selected studies of dose comparison.

Note: BT = brachytherapy, CT = computed tomography, D2cc = dose at 2 cc of volume, EQD2 = equivalent dose of 2 Gy, HR-CTV = high-risk clinical target volume, MRI = magnetic resonance imaging, OARs = organs at risk, TAUS = transabdominal ultrasound.

* Standard plan means prescription to point A.

Nowadays, improving treatment quality in brachytherapy for cervical cancer depends on equipment, manpower and workload. At our Institute, we have CT and TAUS for brachytherapy. The use of TAUS-guided planning or CT contouring supported by US (TAUS or TRUS) will be the future approach to improve IGBT for cervical cancer at our institute.

Conclusions

In conclusion, this study has shown that the point-based plan by TAUS yielded a higher dose to targets and OARS in all parameters, with a mean percentage of difference of approximately 20%. In the additional analysis of percentage of relative dose difference, the dose of 7 Gy to the cervix reference points was close to D98 of HR-CTV (with a mean percentage of difference of 0·6%).

Conflicts of Interest

None.

References

Arbyn, M, Weiderpass, E, Bruni, L et al. Estimates of incidence and mortality of cervical cancer in 2018: a worldwide analysis. Lancet Glob Health 2020; 8 (2): e191e203.CrossRefGoogle ScholarPubMed
Phoolcharoen, N, Kantathavorn, N, Sricharunrat, T et al. A population-based study of cervical cytology findings and human papillomavirus infection in a suburban area of Thailand. Gynecol Oncol Rep 2017; 21: 7377.CrossRefGoogle Scholar
Viswanathan, A N, Beriwal, S, Jennifer, F et al. American Brachytherapy Society consensus guidelines for locally advanced carcinoma of the cervix. Part II: high-dose-rate brachytherapy. Brachytherapy 2012; 11 (1): 4752.CrossRefGoogle ScholarPubMed
International Commission on Radiation Units and Measurement. Dose and Volume Parameters for Prescribing, Recording, and Reporting Brachytherapy, Alone and Combined with External-Beam Radiotherapy. J ICRU 2016; 13: 105122.Google Scholar
Haie-Meder, C, Pötter, R, Van Limbergen, E et al. Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): concepts and terms in 3D image-based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol 2005; 74 (3): 235245.CrossRefGoogle Scholar
Pötter, R, Haie-Meder, C, Van Limbergen, E et al. Recommendations from gynaecological (GYN) GEC ESTRO working group (II): concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy—3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol 2006; 78 (1): 6777.CrossRefGoogle Scholar
Pötter, R, Dimopoulos, J, Georg, P et al. Clinical impact of MRI assisted dose volume adaptation and dose escalation in brachytherapy of locally advanced cervix cancer. Radiother Oncol 2007; 83 (2): 148155.CrossRefGoogle ScholarPubMed
Pötter, R, Georg, P, Dimopoulos, JC et al. Clinical outcome of protocol-based image (MRI) guided adaptive brachytherapy combined with 3D conformal radiotherapy with or without chemotherapy in patients with locally advanced cervical cancer. Radiother Oncol 2011; 100(1): 116123.CrossRefGoogle ScholarPubMed
Lindegaard, J C, Fokdal, L U, Nielsen, S K et al. MRI-guided adaptive radiotherapy in locally advanced cervical cancer from a Nordic perspective. Acta Oncol 2013; 52 (7): 15101519.CrossRefGoogle ScholarPubMed
Sturdza, A, Pötter, R, Fokdal, L U et al. Image guided brachytherapy in locally advanced cervical cancer: improved pelvic control and survival in Retro-EMBRACE, a multicenter cohort study. Radiother Oncol 2016; 120 (3): 428433.CrossRefGoogle Scholar
Tinkle, C L, Weinberg, V, Chen, L M et al. Inverse Planned High-Dose-Rate Brachytherapy for Locoregionally Advanced Cervical Cancer: 4-Year Outcomes. Int J Radiat Oncol Biol Phys 2015; 92 (5): 10931100.CrossRefGoogle ScholarPubMed
Gill, B S, Kim, H, Houser, C J et al. MRI-guided high-dose-rate intracavitary brachytherapy for treatment of cervical cancer: The University of Pittsburgh experience. Int J Radiat Oncol Biol Phys 2015; 91 (3): 540547.CrossRefGoogle ScholarPubMed
Castelnau-Marchand, P, Chargari, C, Maroun, P et al. Clinical outcomes of definitive chemoradiation followed by intracavitary pulsed-dose rate image-guided adaptive brachytherapy in locally advanced cervical cancer. Gynecol Oncol 2015; 139 (2): 288294.CrossRefGoogle ScholarPubMed
Mahanshetty, U, Krishnatry, R, Hande, V et al. Magnetic resonance image guided adaptive brachytherapy in locally advanced cervical cancer: an experience from a tertiary cancer center in a low-and middle-income countries setting. Int J Radiat Oncol Biol Phys 2017; 99 (3): 608617.CrossRefGoogle Scholar
Pötter, R, Tanderup, K, Schmid, M P et al. MRI-guided adaptive brachytherapy in locally advanced cervical cancer (EMBRACE-I): a multi-centre prospective cohort study. Lancet Oncol 2021; 22 (4): 538547.CrossRefGoogle Scholar
Hegazy, N, Pötter, R, Kirisits, C et al. High-risk clinical target volume delineation in CT-guided cervical cancer brachytherapy: impact of information from FIGO stage with or without systematic inclusion of 3D documentation of clinical gynecological examination. Acta Oncol 2013; 52 (7): 13451352.CrossRefGoogle ScholarPubMed
Hellebust, T. Place of modern imaging in brachytherapy planning. Cancer/Radiothérapie 2018; 22 (4): 326333.CrossRefGoogle ScholarPubMed
Tan, L T, Coles, C E, Hart, C et al. Clinical impact of computed tomography-based imaged-guided brachytherapy for cervix cancer using the tandem-ring applicator-the Addenbrooke’s experience. Clin Oncol 2009; 21 (3): 175182.CrossRefGoogle Scholar
Kang, H C, Shin, K H, Park, S Y et al. CT-based high-dose-rate brachytherapy for cervical cancer: clinical impact on late rectal bleeding and local control. Radiother Oncol 2010; 97 (3): 507513.CrossRefGoogle ScholarPubMed
Murakami, N, Kanamatsu, T, Wakita, A et al. CT based three dimensional dose-volume evaluations for high-dose rate intracavitary brachytherapy for cervical cancer. BMC Cancer 2014; 14: 447.CrossRefGoogle ScholarPubMed
Tharavichitkul, E, Chakrabandhu, S, Wanwilairat, S et al. Intermediate-term results of image-guided brachytherapy and high-technology external beam radiotherapy in cervical cancer: Chiang Mai University experience. Gynecol Oncol 2013; 130 (1): 8185.CrossRefGoogle Scholar
Kusada, T, Toita, T, Ariga, T et al. Computed tomography–based image-guided brachytherapy for cervical cancer: correlations between dose–volume parameters and clinical outcomes. J Radiat Res 2018; 59 (1): 6776.CrossRefGoogle ScholarPubMed
Zolciak-Siwinska, A, Gruszczynska, E, Bijok, M et al. Computed tomography–planned high-dose-rate brachytherapy for treating uterine cervical cancer. Int J Radiat Oncol Biol Phys 2016; 96 (1): 8792.CrossRefGoogle ScholarPubMed
Ohno, T, Noda, S E, Okonogi, N et al. In-room computed tomography-based brachytherapy for uterine cervical cancer: results of a 5-year retrospective study. J Radiat Res 2017; 58 (4): 543551.CrossRefGoogle ScholarPubMed
van Dyk, S, Bernshaw, D. Ultrasound-based conformal planning for gynaecological brachytherapy. J Med Imaging Radiat Oncol 2008; 52 (1): 7784.CrossRefGoogle ScholarPubMed
van Dyk, S, Narayan, K, Fisher, R et al. Conformal brachytherapy planning for cervical cancer using transabdominal ultrasound. Int J Radiat Oncol Biol Phys 2009; 75 (1): 6470.CrossRefGoogle ScholarPubMed
Narayan, K, van Dyk, S, Bernshaw, D et al. Ultrasound guided conformal brachytherapy of cervix cancer: survival, patterns of failure, and late complications. J Gynecol Oncol 2014; 25 (3): 206213.CrossRefGoogle ScholarPubMed
Tharavichitkul, E, Tippanya, D, Jayavasti, R et al. Two-year results of transabdominal ultrasound-guided brachytherapy for cervical cancer. Brachytherapy 2015; 14 (2): 238244.CrossRefGoogle ScholarPubMed
Spartacus, R, Dana, R, Gaur, P et al. Ultrasound Image Based High Dose Rate (HDR) Brachytherapy in Locally Advanced Cancer Cervix. International Journal of Cancer Management 2017; 10 (8): e7392 CrossRefGoogle Scholar
Nayak, A, Engineer, R, Mahantshetty, U et al. Transabdominal ultrasonography based high dose rate conformal intracavitary brachytherapy in cervical carcinoma. Int J Radiat Oncol Biol Phys. 2017; 99: e304.CrossRefGoogle Scholar
Tharavichitkul, E, Chakrabandhu, S, Klunklin, P et al. Intermediate-term results of trans-abdominal ultrasound (TAUS)-guided brachytherapy in cervical cancer. Gynecol Oncol 2018; 148 (3): 468473.CrossRefGoogle ScholarPubMed
Viswanathan, A N, Erickson, B, Gaffney, D K et al. Comparison and consensus guidelines for delineation of clinical target volume for CT-and MR-based brachytherapy in locally advanced cervical cancer. Int J Radiat Oncol Biol Phys 2014; 90 (2): 320328.CrossRefGoogle ScholarPubMed
Westerveld, H, Pötter, R, Berger, D et al. Vaginal dose point reporting in cervical cancer patients treated with combined 2D/3D external beam radiotherapy and 2D/3D brachytherapy. Radiother Oncol 2013; 107 (1): 99105.CrossRefGoogle ScholarPubMed
Poddar, J, Sharma, A D, Suryanarayan, U et al. Calculation of dose volume parameters and indices in plan evaluation of HDR interstitial brachytherapy by MUPIT in carcinoma cervix. Indian J Cancer 2018; 55 (3): 238.CrossRefGoogle ScholarPubMed
Pötter, R, Tanderup, K, Kirisits, C et al. The EMBRACE II study: the outcome and prospect of two decades of evolution within the GEC-ESTRO GYN working group and the EMBRACE studies. Clin Transl Radiat Oncol 2018; 9: 4860.CrossRefGoogle Scholar
De Brabandere, M, Mousa, A G, Nulens, A et al. Potential of dose optimization in MRI-based PDR brachytherapy of cervix carcinoma. Radiother Oncol 2008; 88 (2): 217226.CrossRefGoogle Scholar
Lindegaard, JC, Tanderup, K, Nielsen, S K et al. MRI-guided 3D optimization significantly improves DVH parameters of pulsed-dose-rate brachytherapy in locally advanced cervical cancer. Int J Radiat Oncol Biol Phys 2008; 71 (3): 756764.CrossRefGoogle ScholarPubMed
Zwahlen, D, Jezioranski, J, Chan, P et al. Magnetic resonance imaging-guided intracavitary brachytherapy for cancer of the cervix. Int J Radiat Oncol Biol Phys 2009; 74 (4): 11571164.CrossRefGoogle ScholarPubMed
Tharavichitkul, E, Mayurasakorn, S, Lorvidhaya, V et al. Preliminary results of conformal computed tomography (CT)-based intracavitary brachytherapy (ICBT) for locally advanced cervical cancer: a single institution’s experience. J Radiat Res 2011; 52 (5): 634640.CrossRefGoogle ScholarPubMed
Viswanathan, AN, Dimopoulos, J, Kirisits, C, Berger, D, Pötter, R. Computed tomography versus magnetic resonance imaging-based contouring in cervical cancer brachytherapy: results of a prospective trial and preliminary guidelines for standardized contours. Int J Radiat Oncol Biol Phys 2007; 68 (2): 491498.CrossRefGoogle ScholarPubMed
Eskander, RN, Scanderbeg, D, Saenz, CC, Brown, M, Yashar, C. Comparison of computed tomography and magnetic resonance imaging in cervical cancer brachytherapy target and normal tissue contouring. Int J Gynecol Cancer 2010; 20 (1): 4753.CrossRefGoogle ScholarPubMed
Mahantshetty, U, Khanna, N, Swamidas, J et al. Trans-abdominal ultrasound (US) and magnetic resonance imaging (MRI) correlation for conformal intracavitary brachytherapy in carcinoma of the uterine cervix. Radiother Oncol 2012; 102 (1): 130134.CrossRefGoogle ScholarPubMed
Schmid, M P, Nesvacil, N, Pötter, R et al. Transrectal ultrasound for image-guided adaptive brachytherapy in cervix cancer–an alternative to MRI for target definition? Radiother Oncol 2016; 120 (3): 467472 CrossRefGoogle ScholarPubMed
Mahantshetty, U, Khadanga, C, Gudi, S et al. A Prospective Comparison of Computed Tomography with Transrectal Ultrasonography Assistance and Magnetic Resonance Imaging-Based Target-Volume Definition During Image Guided Adaptive Brachytherapy for Cervical Cancers. Int J Radiat Oncol Biol Phys 2018; 102 (5): 14481456.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. Study process.

Figure 1

Figure 2. Measurements of eight cervix dimensions by TAUS: (a) sagittal view of TAUS, (b) axial view of TAUS at the cervical os and (c) axial view of TAUS at 2 cm cranially from the cervical os.

Figure 2

Figure 3. HR-CTV, IR-CTV and eight cervix reference points (A1-A4 & L1-L4) in the CT datasets.

Figure 3

Figure 4. Isodose distributions of (a) volume-based plan by CT and (b) point-based plan by TAUS.

Figure 4

Figure 5. Percentage dose difference and relative dose difference formulae.

Figure 5

Table 1. Characteristics data

Figure 6

Table 2. Doses to target for volume-based plan by CT and point-based plan by TAUS plans

Figure 7

Table 3. Percentage relative dose difference of HR-CTV in point-based plan by TAUS plan

Figure 8

Table 4. Doses to organs at risk for volume-based plan by CT and point-based plan by TAUS plans.

Figure 9

Table 5. Selected studies of dose comparison.