Hostname: page-component-cd9895bd7-q99xh Total loading time: 0 Render date: 2024-12-23T12:56:05.844Z Has data issue: false hasContentIssue false

Adding another dimension to plan evaluation: visualising the dose–volume histogram band in head and neck radiotherapy and exploring its utility

Published online by Cambridge University Press:  18 April 2017

S. Samanta
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
S. Balukrishna*
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
K.M. Rafic
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
B.S. Timothy Peace
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
I.R.R. Singh
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
S.P. Pavamani
Affiliation:
Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore, Tamil Nadu, India
*
Correspondence to: Balu Krishna Sasidharan, Department of Radiation Oncology, Dr Ida B Scudder Cancer Center, Christian Medical College, Vellore 632004, Tamil Nadu, India. Tel: +91 96 2626 2296. Fax: +91 41 6228 2035. E-mail: [email protected]

Abstract

Background

To introduce a method to generate a ‘dose–volume histogram (DVH) band’ for plan evaluation of photon therapy and explore its various potentials.

Materials and methods

Intensity-modulated radiotherapy (IMRT) plans for head and neck cancer patients were analysed, retrospectively, for setup errors noted during treatment. From the maximum observed random errors, absolute displacement was calculated using Euclidian formula. The original plan with same beam parameters and leaf sequence were used to generate six plans with shifts applied in three axes in six directions. The DVH curves from these six plans were superimposed to form the DVH band. Plans were reviewed with set tolerance criteria.

Results

Method to generate and visualise DVH band was developed. DVH bands were created for 20 patients with head and neck cancer who underwent treatment with IMRT. It was found that seven out these 20 plans were rejected as they crossed the set tolerance criteria using DVH band as an evaluation tool.

Conclusions

DVH band in photon therapy can help the clinician visualise the impact of setup errors at planning and may help select the plan with lesser influence of setup errors over another.

Type
Original Articles
Copyright
© Cambridge University Press 2017 

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1.Dose–volume histograms. http://www.sciencedirect.com/science/article/pii/0360301691901684. Accessed on 14th February 2016.Google Scholar
2. Murthy, K K, Al-Rahbi, Z, Sivakumar, S S, Davis, C A, Ravichandran, R, El Ghamrawy, K. Verification of setup errors in external beam radiation therapy using electronic portal imaging. J Med Phys 2008; 33 (2): 4953.Google ScholarPubMed
3. Booth, J T, Zavgorodni, S F. Set-up error & organ motion uncertainty: a review. Australas Phys Eng Sci Med 1999; 22 (2): 2947.Google ScholarPubMed
4. Hurkmans, C W, Remeijer, P, Lebesque, J V, Mijnheer, B J. Set-up verification using portal imaging; review of current clinical practice. Radiother Oncol 2001; 58 (2): 105120.Google Scholar
5. Cheng, C W, Das, I J. Treatment plan evaluation using dose-volume histogram (DVH) and spatial dose-volume histogram (zDVH). Int J Radiat Oncol Biol Phys 1999; 43 (5): 11431150.CrossRefGoogle ScholarPubMed
6. Trofimov, A, Unkelbach, J, Delaney, T F, Bortfeld, T. Visualization of a variety of possible dosimetric outcomes in radiation therapy using dose-volume histogram bands. Pract Radiat Oncol 2012; 2 (3): 164171.Google Scholar
7. Deza, E, Deza, M M. Encyclopedia of Distances. Springer, p. 94; (Internet) http://www.springer.com/us/book/9783642002342.Google Scholar
8. Journal of the ICRU, vol 10, no 1 (2010); Report 83, Oxford University Press, doi:10.1093/jicru/ndq001 Google Scholar