Published online by Cambridge University Press: 22 March 2021
This study aimed to evaluate hearing outcomes and device safety in a large, single-surgeon experience with the totally implantable active middle-ear implants.
This was a retrospective case series review of 116 patients with moderate-to-severe sensorineural hearing loss undergoing implantation of active middle-ear implants.
Mean baseline unaided pure tone average improved from 57.6 dB before surgery to 34.1 dB post-operatively, signifying a mean gain in pure tone average of 23.5 dB (p = 0.0002). Phonetically balanced maximum word recognition score improved slightly from 70.5 per cent to 75.8 per cent (p = 0.416), and word recognition score at a hearing level of 50 dB values increased substantially from 14.4 per cent to 70.4 per cent (p < 0.0001). Both revision and explant rates were low and dropped with increasing surgeon experience over time.
This study showed excellent post-operative hearing results with active middle-ear implants with regard to pure tone average and word recognition score at a hearing level of 50 db. Complication rates in this case series were significantly lower with increasing experience of the surgeon. Active middle-ear implants should be considered in appropriate patients with moderate-to-severe sensorineural hearing loss who have struggled with conventional amplification and are good surgical candidates.
Dr D Morrison takes responsibility for the integrity of the content of the paper