Hostname: page-component-cd9895bd7-gvvz8 Total loading time: 0 Render date: 2024-12-23T13:18:39.236Z Has data issue: false hasContentIssue false

An unusual occurrence in cochlear implantation surgery: misplaced electrode

Published online by Cambridge University Press:  04 November 2008

R Ramalingam
Affiliation:
Head and Neck department of the KKR, ENT Hospital and Research Institute, Chennai, India
K K Ramalingam*
Affiliation:
Head and Neck department of the KKR, ENT Hospital and Research Institute, Chennai, India
H S Padmaja
Affiliation:
Head and Neck department of the KKR, ENT Hospital and Research Institute, Chennai, India
*
Address for correspondence: Prof K K Ramalingam, KKR ENT Hospital and Research Institute (P) Ltd, 274 Poonamallee High Road, Kilpauk, Chennai 600010, India. Fax: 91 44 26412727 E-mail: [email protected]

Abstract

Objective:

To highlight the possibility of misplacement of electrodes during cochlear implantation surgery, to stress the importance of routine neural response telemetry, and also to emphasise the value of conventional radiography in confirming electrode positioning in the immediate post-operative period.

Case report:

A two-year-old boy presented with bilateral, profound, sensorineural hearing loss and underwent conventional cochlear implantation surgery. During insertion of the implant, there was doubt regarding the direction of passage of the electrode array. Instead of the usual smooth passage of the electrode, some resistance was felt. Neural response telemetry, performed at the end of the procedure, showed absent responses. A confirmatory X-ray in Stenver's view confirmed the extra-cochlear positioning of the electrodes in the superior semicircular canal. The wound was reopened, the electrode array removed and the cochleostomy was positioned a little more antero-inferiorly on the promontory with respect to the round window. Correct insertion of the electrode was then performed without difficulty.

Conclusion:

The misplacement of electrodes during cochlear implantation surgery is rare but can occur. Neural response telemetry and conventional radiography are invaluable in assessing the placement of the electrode intra-operatively. Conventional radiographs are very cost-effective and aid in confirming the position of the electrode array.

Type
Online Only Clinical Records
Copyright
Copyright © JLO (1984) Limited 2008

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1 Mecca, AM, Wagle, W, Lupinetti, A, Parnes, S. Complication of cochlear implantation surgery. AJNR Am J Neuroradiol 2003;24:2089–91Google ScholarPubMed
2 Ball, JB Jr, Miller, GW, Hepfner, ST. Computed tomography of single-channel cochlear implants. AJNR Am J Neuroradiol 1986;7:41–7Google ScholarPubMed
3 Balkany, T, Gantz, BJ, Steenerson, RL, Cohen, NL. Systematic approach to electrode insertion in the ossified cochlea. Otolaryngol Head Neck Surg 1991;114:411CrossRefGoogle Scholar
4 NIH Consensus Conference. Cochlear implants in adults and children presented at NIH consensus conference. JAMA 1995;274:1955–61CrossRefGoogle Scholar
5 Tykocinski, M, Saunders, E, Cohen, LT. The contour electrode array: safety study and initial patient trials of a new perimodiolar design. Otol Neurotol 2001;22:3341CrossRefGoogle ScholarPubMed
6 Gstoettner, WK, Adunka, O, Franz, P. Perimodiolar electrodes in cochlear implant surgery. Acta Otolaryngol 2001;121:216–19Google ScholarPubMed
7 Lo, WWM. Imaging of cochlear and auditory brain stem implantation. AJNR Am J Neuroradiol 1998;19:1147–54Google ScholarPubMed
8 Harnsberger, HR, Dart, DJ, Parkin, JL, Smoker, WR, Osborn, AG. Cochlear implant candidates: assessment with CT and MR imaging. Radiology 1987;164:53–7CrossRefGoogle ScholarPubMed
9 Marsh, MA, Xu, J, Blamey, PJ. Radiologic evaluation of multichannel intracochlear implant insertion depth. Am J Otol 1993;14:386–91Google ScholarPubMed
10 Shpizner, BA, Holliday, RA, Roland, JT, Cohen, NL, Waltzman, SB, Shapiro, WH. Postoperative imaging of the multichannel cochlear implant. AJNR Am J Neuroradiol 1996;16:1517–24Google Scholar