Selection of the appropriate cochlear electrode array using a specifically developed research software application
Published online by Cambridge University Press: 18 June 2018
Abstract
To evaluate the usefulness and reliability of a research software application for the estimation of an individual's cochlear duct length as a basis for electrode selection.
In this prospective cohort study, 21 consecutive patients (23 ears) implanted with a cochlear electrode were investigated. The study comprised 19 children (2 bilateral) and 2 adults.
The measured ‘A’ distances (the largest distance from the round window to the contralateral wall) corresponded to cochlear duct lengths of 28.5–36.4 mm. The mean cochlear duct length was 34.05 ± 1.72 mm (33.60 ± 2.27 mm in females and 34.35 ± 1.27 mm in males). Full insertion was achieved in all but two cases. No misplaced electrode array or electrode fold-over was detected. In all but three ears, the electrode was chosen based on the research software application's indication.
The results show a good correlation between the pre-operatively predicted insertion depths using the software application and those post-operatively measured using X-ray. The insertion length predicted by the software was always longer than that measured via X-ray.