Hostname: page-component-cd9895bd7-gvvz8 Total loading time: 0 Render date: 2024-12-25T05:30:59.162Z Has data issue: false hasContentIssue false

An in vitro study of tracheostomy tube cuff herniation and inflation characteristics

Published online by Cambridge University Press:  29 June 2007

L. J. Hollis*
Affiliation:
Department of Otolaryngology – Head and Neck Surgery, St Mary's Hospital, Praed Street, London, UK.
J. S. Almeyda
Affiliation:
Department of Otolaryngology – Head and Neck Surgery, St Mary's Hospital, Praed Street, London, UK.
G. Mochloulis
Affiliation:
Department of Otolaryngology – Head and Neck Surgery, St Mary's Hospital, Praed Street, London, UK.
K. S. Patel
Affiliation:
Department of Otolaryngology – Head and Neck Surgery, St Mary's Hospital, Praed Street, London, UK.
*
Address for correspondence: Mr L. J. Hollis, Flat 7, The Helme, 10 Lake Road, Wimbledon, London SW19 7EN

Abstract

Seven of the most commonly used tracheostomy tubes used in the UK were tested for cuff herniation and creasing in synthetic tracheas corresponding to the shapes and sizes found in vivo. Results demonstrated that only two tubes of one particular brand herniated and that creasing occured in tubes disproportionaltely large for the trachea used. A discussion of modern tracheostomy tube manufacture is included.

Keywords

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 1996

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Bar-Lavie, Y., Gatot, A., Tovi, F. (1995) Intraoperative herniation of a tracheostomy tube cuff. Journal of Laryngology and Otology 109: 159160.CrossRefGoogle ScholarPubMed
Bernhard, W. N., Yost, L., Joynes, D., Cothalis, S., Turndorf, H. (1985) Intracuff pressures in endotracheal and tracheostomy ubes. Chest 87: 720725.CrossRefGoogle Scholar
Davidson, I., Zimmer, S. (1989) Cuff herniation. Anaesthesia 44: 938939.CrossRefGoogle ScholarPubMed
Mackenzie, C. F., Shin, B., Whitley, N., Schellinger, D. (1979) The relationship of human tracheal size to body habitus. Anesthesiology 51(3): S378.CrossRefGoogle Scholar
Robbie, D. S., Pearce, D. J. (1959) Some dangers of armoured tubes. Anaesthesia 14(4): 379385.CrossRefGoogle ScholarPubMed
Stanley, T. H. (1975) Nitrous oxide and pressures and volumes of high and low pressure endotracheal tube cuff in intubated patients. Anesthesiology 42(5): 637640.CrossRefGoogle Scholar
Tracheostomy tubes Part 2. Basic requirements for tubes for adults. British Standards Institution (1993).Google Scholar
Treffers, R., De Lange, J. J. (1989) An unusual case of cuff herniation. Acta Anaesthesiologica Belgica 40(1): 8790.Google ScholarPubMed
Ward, C. F., Gamel, D. M., Benumof, J. L. (1978) Endotracheal tube cuff herniation: a cause of delayed airway obstruction. Anesthesia and Analgesia 57: 114116.CrossRefGoogle ScholarPubMed