Investigations into falsely elevated somatic cell counts of bulked herd milk
Published online by Cambridge University Press: 01 June 2009
Summary
Bulk milk cell counts obtained by a Coulter Milk Cell Counter were shown to be falsely elevated because of the formation of non-cellular particles (NCP). The NCP, which were detected by the use of a particle size analyser, were removed if the milk samples were heated to 55 °C for 15 min before the addition of formalin. A survey of bulk milk supplies in New South Wales showed that NCP occurred in milk from 2 % of farms, the counts being falsely elevated by about 200000 cells/ml. Surveys of milking equipment of farms where the bulk milk had shown this phenomenon demonstrated a relationship between the use of centrifugal milk pumps and the presence of NCP with a much lower prevalence in herds where spit chamber releasers were used. These particles were not air bubbles since air was removed from the milk by heating to 60 °C for 5 min, vacuum at 65 mmHg for 15 minor 0·5% v/voctan-2-ol; only the heat treatment removed the NCP.