Potential impact of contaminated bronchoscopes on novel coronavirus disease (COVID-19) patients

Published online by Cambridge University Press:  02 April 2020

Cori L. Ofstead*
Affiliation:
Ofstead and Associates, St Paul, Minnesota
Krystina M. Hopkins
Affiliation:
Ofstead and Associates, St Paul, Minnesota
Matthew J. Binnicker
Affiliation:
Mayo Clinic Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota
Gregory A. Poland
Affiliation:
Mayo Clinic Vaccine Research Group, Mayo Clinic, Rochester, Minnesota
*
Author for correspondence: Cori L. Ofstead, E-mail: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Type
Letter to the Editor
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited.
Copyright
© 2020 by The Society for Healthcare Epidemiology of America. All rights reserved.

To the Editor—During the novel coronavirus disease (COVID-19) pandemic, critically ill patients may require therapeutic bronchoscopy or sample collection via bronchoalveolar lavage (BAL), which involves using a bronchoscope to flush lungs with saline solution. Results of BAL assays are used to make clinical decisions that may impact outcomes. Clinicians have reported that COVID-19 patients had bacterial and fungal pulmonary coinfectionsReference Chen, Zhou and Dong1 with potential pathogens including Escherichia, Salmonella, Pseudomonas, and Stenotrophomonas.Reference Ren, Wang and Wu2 Recent research suggests that COVID-19 coinfections are associated with significantly higher mortality rates.Reference Ruan, Yang, Wang, Jiang and Song3

Numerous nosocomial outbreaks and pseudo-outbreaks have been linked to inadequately reprocessed bronchoscopes.Reference Galdys, Marsh and Delgado4,Reference Kovaleva, Peters, van der Mei and Degener5 In 2018, researchers in Wuhan City, China, identified Stenotrophomonas maltophilia in 55.55% of BAL samples.Reference Liu and Tong6 The source was the channel of an improperly reprocessed bronchoscope, and the pseudo-outbreak involved 25 asymptomatic patients undergoing treatment for tuberculosis and other infections. Reprocessing and hand-hygiene deficiencies were identified.

Ofstead et alReference Ofstead, Quick and Wetzler7-Reference Ofstead, Doyle and Eiland9 have conducted prospective studies that evaluated effectiveness of bronchoscope reprocessing in 5 hospitals in the United States. Microbial growth was detected on 23 of 35 bronchoscopes (65.7%), and 10 bronchoscopes (28.6%) harbored high-concern organisms or actionable levels of microbial growth (>100 CFU) (Table 1).Reference Ofstead, Quick and Wetzler7-Reference Ofstead, Doyle and Eiland9 Mold and gram-negative bacteria were detected, including S. maltophilia, Sphingomonas phyllosphaerae, and Escherichia coli/Shigella. At one hospital, high protein levels were detected in 7 of 8 bronchoscopes, indicating that manual cleaning failed to remove soil.Reference Ofstead, Quick and Wetzler7 Visual inspections using magnification and borescopes identified residue or defects in 100% of bronchoscopes.Reference Ofstead, Quick and Wetzler7,Reference Ofstead, Heymann, Quick, Eiland and Wetzler8 Audits evaluating personal protective equipment use and reprocessing guideline adherence (eg, point-of-care precleaning; leak testing; manual cleaning; visual inspection; cleaning verification; high-level disinfection; rinsing; drying; storage; transport and handling) identified breaches in all 5 hospitals.Reference Ofstead, Quick and Wetzler7-Reference Ofstead, Doyle and Eiland9 Technicians in 2 hospitals (sites 1 and 5) performed most reprocessing steps correctly, but bronchoscopes at both sites harbored S. maltophilia due to contaminated rinse water.Reference Ofstead, Quick and Wetzler7,Reference Ofstead, Doyle and Eiland9 In 3 hospitals (sites 2–4), nearly all steps were performed incorrectly or were skipped entirely.Reference Ofstead, Quick and Wetzler7,Reference Ofstead, Heymann, Quick, Eiland and Wetzler8 In light of these breaches and observations that most bronchoscopes were damaged and contaminated, a recommendation was made that procedures in 2 hospitals be halted until strict protocols could be implemented and personnel retrained. In addition, it was recommended that badly damaged bronchoscopes be removed from service and replaced with single-use, sterile bronchoscopes or new reusable bronchoscopes constructed with sterilizable materials.

Table 1. Microbial Culture Results From Fully Reprocessed Bronchoscopes in 5 Hospitals

Note. CFU, colony-forming units; NA, not applicable; TNTC, too numerous to count; GPC, gram-positive cocci.

a Results from a swab of the ultrasound component of an EBUS bronchoscope appear in parentheses.

There is currently an urgent need to reduce the number of patients requiring hospitalization or intensive care, in part because of shortages of ventilators and personal protective equipment. Given the high bronchoscope contamination rates found during routine use in previous studies, we must now consider the possibility of bronchoscopy-associated transmission of COVID-19 or other pathogens that could cause secondary infections. Theoretically, high-level disinfection should eliminate these risks when bronchoscopes are well-maintained and reprocessed according to manufacturer instructions and professional guidelines. However, even during normal patient loads, practices are frequently substandard, and pathogens are commonly present on patient-ready endoscopes. The presence of gastrointestinal pathogens found in bronchoscopes and BAL samples suggests the possibility of cross-contamination caused by intermingling bronchoscopes and gastrointestinal endoscopes during reprocessing. This hypothesis is supported by findings at one hospital where protein and bioburden levels on brand-new bronchoscopes increased significantly following manual cleaning prior to any clinical use.Reference Ofstead, Quick and Wetzler7

Researchers recently reported COVID-19 patients presenting with diarrhea and abdominal pain, with fecal carriage of SARS-CoV-2 among severely ill and asymptomatic patients. Thus, extreme care must be taken to minimize cross-contamination during all endoscope reprocessing.

Reprocessing effectiveness has not been evaluated in epidemic settings, and research is needed to confirm that COVID-19, influenza viruses, and other pathogens are eliminated in these settings. The use of sterile, disposable bronchoscopes would substantially reduce the risks for patients and reprocessing personnel, and this approach has been recommended by the American Association for Bronchology and Interventional Pulmonology.Reference Wahidi, Lamb and Murgu10 However, single-use bronchoscopes are not universally available and may not be sufficient for advanced bronchoscopy. When reusable bronchoscopes must be used, they should be segregated from gastrointestinal endoscopes and sterilized rather than relying on high-level disinfection.

We urgently recommend further research assessing potential contamination of reusable bronchoscopes with viral, bacterial, and fungal pathogens. Laboratory methods should include bacterial/fungal cultures and molecular assays (eg, real-time PCR) for respiratory viruses, including COVID-19. To optimize the accuracy of results, samples should be taken from multiple components using a friction-based technique (eg, flush-brush-flush for sampling ports and channels). Laboratories should utilize methods that foster growth of microbes that are viable but not easily culturable (eg, using neutralizers to counteract residual reprocessing chemicals that could suppress growth, concentrating samples, and/or incubating for at least 5–7 days or 6–8 weeks when culturing for Mycobacteria). Due to the relative insensitivity of viral culture and potential safety concerns related to cultivating COVID-19, molecular testing (ie, targeted real-time PCR and multiplex respiratory panels) could be considered to assess contamination with viral pathogens.

No patient should suffer from preventable nosocomial infections due to bronchoscopy. Using bronchoscopes that have physical defects and harbor viruses, bacteria, or fungi puts vulnerable patients at risk and could have adverse effects on public health. Institutions are obligated to protect both patients and reprocessing personnel and ensure bronchoscope reprocessing practices adhere to guidelines and manufacturer instructions. The urgency of the current COVID-19 situation underscores the need for robust quality management practices, including audits or virtual audits by qualified experts, visual inspection, and biochemical tests to verify reprocessing effectiveness. These measures are essential for protecting healthcare workers and preventing erroneous BAL test results and bronchoscopy-associated pathogen transmission due to the use of contaminated bronchoscopes.

Acknowledgments

None.

Financial support

Our work on this manuscript was completed without outside funding, and no research sponsors or clients were aware of or had any role in developing or reviewing this manuscript.

Conflicts of interest

C.L.O. and K.M.H. have received research grants, study materials, educational materials, or consulting contracts from 3M Company, Ambu/Invendo, Auris Health, Advanced Sterilization Products, Boston Scientific, Laborie/Cogentix, Convergascent, Endoscopy Repair Specialists, Fortive, Healthmark, Cantel/Medivators, Mobile Instrument, Nanosonics, and Steris. MJB is an advisory board member for DiaSorin Molecular. GAP has provided consultation for Merck Research Laboratories, Avianax, Valneva, Medicago, Sanofi Pasteur, GlaxoSmithKline, Emergent Biosolutions, and Dynavax.

References

Chen, N, Zhou, M, Dong, X, et al. Epidemiological and clinical characteristics of 99 cases of 2019 novel coronavirus pneumonia in Wuhan, China: a descriptive study. Lancet 2020;395:507513.CrossRefGoogle ScholarPubMed
Ren, LL, Wang, YM, Wu, ZQ, et al. Identification of a novel coronavirus causing severe pneumonia in human: a descriptive study. Chin Med J (Engl) 2020 [Epub ahead of print]. doi: 10.1097/CM9.0000000000000722.CrossRefGoogle ScholarPubMed
Ruan, Q, Yang, K, Wang, W, Jiang, L, Song, J. Clinical predictors of mortality due to COVID-19 based on an analysis of data of 150 patients from Wuhan, China. Intensive Care Med 2020 [Epub ahead of print]. doi: 10.1007/s00134-020-05991-x.Google Scholar
Galdys, AL, Marsh, JW, Delgado, E, et al. Bronchoscope-associated clusters of multidrug-resistant Pseudomonas aeruginosa and carbapenem-resistant Klebsiella pneumoniae . Infect Control Hosp Epidemiol 2019;40:4046.CrossRefGoogle ScholarPubMed
Kovaleva, J, Peters, FT, van der Mei, HC, Degener, JE. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Rev 2013;26:231254.CrossRefGoogle ScholarPubMed
Liu, B, Tong, S. An investigation of Stenotrophomonas maltophilia-positive culture caused by fiberoptic bronchoscope contamination. BMC Infect Dis 2019;19:1072.CrossRefGoogle ScholarPubMed
Ofstead, CL, Quick, MR, Wetzler, HP, et al. Effectiveness of reprocessing for flexible bronchoscopes and endobronchial ultrasound bronchoscopes. Chest 2018;154:10241034.CrossRefGoogle ScholarPubMed
Ofstead, CL, Heymann, OL, Quick, MR, Eiland, JE, Wetzler, HP. Residual moisture and waterborne pathogens inside flexible endoscopes: evidence from a multisite study of endoscope drying effectiveness. Am J Infect Control 2018;46:689696.CrossRefGoogle ScholarPubMed
Ofstead, CL, Doyle, EM, Eiland, JE, et al. Practical toolkit for monitoring endoscope reprocessing effectiveness: identification of viable bacteria on gastroscopes, colonoscopes, and bronchoscopes. Am J Infect Control 2016;44:815819.CrossRefGoogle ScholarPubMed
Wahidi, M, Lamb, C, Murgu, S, et al. American Association for Bronchology and Interventional Pulmonology (AABIP) statement on the use of bronchoscopy and respiratory specimen collection in patients with suspected or confirmed COVID-19 infection. AABIP website. https://aabronchology.org/wp-content/uploads/2020/03/2020-AABIP-Statement-on-Bronchoscopy-COVID.GAE-updated-Version.pdf. Published March 2020. Accessed April 24, 2020.Google Scholar
Figure 0

Table 1. Microbial Culture Results From Fully Reprocessed Bronchoscopes in 5 Hospitals