Hostname: page-component-78c5997874-ndw9j Total loading time: 0 Render date: 2024-11-05T15:24:24.452Z Has data issue: false hasContentIssue false

Evaluation of a Rapid Readout Biological Indicator for 121°C Gravity and 132°C Vacuum-Assisted Steam Sterilization Cycles

Published online by Cambridge University Press:  02 January 2015

Donald Vesley*
Affiliation:
School of Public Health, University of Minnesota, Minneapolis, Minnesota
Melissa A. Nellis
Affiliation:
School of Public Health, University of Minnesota, Minneapolis, Minnesota
Paul B. Allwood
Affiliation:
School of Public Health, University of Minnesota, Minneapolis, Minnesota
*
Environmental and Occupational Health, School of Public Health, University of Minnesota, Box 807 Mayo, 420 Delaware St. SE, Minneapolis, MN 55455

Abstract

Objective:

A biological indicator (BI), based on fluorescent detection of a spore-associated enzyme, has been developed. We evaluated the new BI against five conventional self-contained BIs in a 121°C gravity-displacement sterilizer and, packaged in towel packs and disposable steam test packs, in a 132°C vacuum-assisted sterilizer.

Design:

In the 121°C gravity-displacement sterilizer, rapid readout BIs and conventional BIs were tested together in an otherwise empty chamber. Exposure times were 5, 10, and 15 minutes. In the 132°C vacuum-assisted steam sterilizer, BIs were contained in towel packs or disposable steam test packs. Surgical packs were added to represent a full load. Exposure times were 0, 1, and 2 minutes.

Result:

In gravity-displacement 5-minute cycles, the rapid readout BIs yielded 100% positive results within 30 minutes. The conventional BIs ranged from 72% to 100% positive in 48 hours. At 10 minutes, the rapid readout BIs were 83% positive after 3 hours, whereas the conventional BIs ranged from 0% to 37% positive after 48 hours. All indicators were negative after 15-minute exposures. In vacuum-assisted cycles, all indicators were negative after 3 minutes. After 1-minute exposures, the rapid readout BIs were 40% to 60% positive, whereas the conventional BIs were 15% to 20% positive.

Conclusion:

The rapid readout BIs are a more sensitive indicator of marginal steam sterilization cycles than conventional selfcontained BIs. They yield results within 3 hours, as compared to 24 to 48 hours for the conventional BIs.

Type
Original Articles
Copyright
Copyright © The Society for Healthcare Epidemiology of America 1995

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Association of Operating Room Nurses. Proposed recommended practices for sterilization in the practice setting. AORN J 1994;52:109120.Google Scholar
2. Joint Commission on Accreditation of Healthcare Organizations. Accreditation Manual for Hospitals. Oakbrook Terrace, IL: Joint Commission on Accreditation of Healthcare Organizations; 1994.Google Scholar
3. American Society for Healthcare Central Service Personnel of the American Hospital Association. l&commended Practice for Central Service, Sterilization. Chicago, IL American Hospital Association; 1989.Google Scholar
4. Association for the Advancement of Medical Instrumentation. Good Hospital Practice: Steam Sterilization and Sterility Assurance. Arlington, VA: AAMI;1993.Google Scholar
5. Gamer, JS, Favero, MS. Guidelines for Handwashing and Environmental Control. CDC Publication No. 85923404. Atlanta, GA: Centers for Disease Control; 1985.Google Scholar
6. Vesley, D, Langholz, AC, Rohlfing, SR, Foltz, WE. Fluorimetric detection of a Bacillus stearothermophilus spore-bound enzyme, α-D-glucosidase, for rapid identification of flash sterilization failure. Appl Environ Microbiol 1992;58:717719.CrossRefGoogle Scholar
7. Rutala, WA, Gergen, MF: Weber, DJ. Evaluation of a rapid readout biological indicator for flash sterilization with three biological indicators and three chemical indicators. Infect Control Hosp Epidemiol 1993;14:390394.CrossRefGoogle ScholarPubMed
8. Association for the Advancement of Medical Instrumentation (AAMI). American National Standard for Biological Indicators for Saturated Steam Sterilization Processes in Health Care facilities, ST19. Arlington, VA: AAMI; 1986.Google Scholar