Hostname: page-component-586b7cd67f-2plfb Total loading time: 0 Render date: 2024-11-28T15:08:22.589Z Has data issue: false hasContentIssue false

Comparison of virus aerosol concentrations across a face shield worn on a healthcare personnel during a simulated patient cough

Published online by Cambridge University Press:  23 August 2023

Alessandra A. Pratt*
Affiliation:
University of Iowa, Department of Occupational and Environmental Health, Iowa City, Iowa Iowa City Veterans Affairs Health Care System, Iowa City, Iowa
Grant D. Brown
Affiliation:
Department of Biostatistics, University of Iowa, Iowa City, Iowa
Eli N. Perencevich
Affiliation:
Iowa City Veterans Affairs Health Care System, Iowa City, Iowa Department of Internal Medicine, Carver College of Medicine, University of Iowa, Iowa City, Iowa
Daniel J. Diekema
Affiliation:
Department of Internal Medicine, Carver College of Medicine, University of Iowa, Iowa City, Iowa Department of Medicine, Maine Medical Center, Portland Maine
Matthew W. Nonnenmann
Affiliation:
Department of Environmental, Agricultural and Occupational Health, University of Nebraska Medical Center, Omaha, Nebraska
*
Corresponding author: Alessandra Pratt; Email: [email protected]
Rights & Permissions [Opens in a new window]

Abstract

Background:

Patients diagnosed with coronavirus disease 2019 (COVID-19) aerosolize severe acute respiratory coronavirus virus 2 (SARS-CoV-2) via respiratory efforts, expose, and possibly infect healthcare personnel (HCP). To prevent transmission of SARS-CoV-2 HCP have been required to wear personal protective equipment (PPE) during patient care. Early in the COVID-19 pandemic, face shields were used as an approach to control HCP exposure to SARS-CoV-2, including eye protection.

Methods:

An MS2 bacteriophage was used as a surrogate for SARS-CoV-2 and was aerosolized using a coughing machine. A simulated HCP wearing a disposable plastic face shield was placed 0.41 m (16 inches) away from the coughing machine. The aerosolized virus was sampled using SKC biosamplers on the inside (near the mouth of the simulated HCP) and the outside of the face shield. The aerosolized virus collected by the SKC Biosampler was analyzed using a viability assay. Optical particle counters (OPCs) were placed next to the biosamplers to measure the particle concentration.

Results:

There was a statistically significant reduction (P < .0006) in viable virus concentration on the inside of the face shield compared to the outside of the face shield. The particle concentration was significantly lower on the inside of the face shield compared to the outside of the face shield for 12 of the 16 particle sizes measured (P < .05).

Conclusions:

Reductions in virus and particle concentrations were observed on the inside of the face shield; however, viable virus was measured on the inside of the face shield, in the breathing zone of the HCP. Therefore, other exposure control methods need to be used to prevent transmission from virus aerosol.

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America

Severe acute respiratory coronavirus virus 2 (SARS-CoV-2) is a positive sense, single-stranded, RNA virus in the coronavirus family. The virus was originally detected in humans in 2019 and has since caused a global pandemic with >533 million cases and 6.3 million deaths as of June 2022. SARS-CoV-2 causes coronavirus disease 2019 (COVID-19). The disease has symptoms including fever, cough, shortness of breath, fatigue, sore throat, nausea, and diarrhea. In the United States, ∼80 million patients have been hospitalized with COVID-19.

There may be a dose response associated with COVID-19 severity, Reference Van Damme, Dahake, van de Pas, Vanham and Assefa1 meaning that the dose of virus in the initial inoculum could affect the severity of disease. Reference Van Damme, Dahake, van de Pas, Vanham and Assefa1 Patients can aerosolize virus, which exposes the HCP. Therefore, reducing SARS-CoV-2 aerosol concentration and worker exposure could reduce the disease burden among HCPs. Virus containing aerosols can be reduced by following the industrial hygiene hierarchy of controls, which provides guidance on reducing exposure to hazards through elimination, substitution, engineering controls, administrative controls, or personal protective equipment (PPE). PPE is often used in healthcare to reduce exposure to viruses and can be worn by the HCP. Reference Setti, Passarini and de Gennaro2 PPE can include gowns, gloves, face shields, goggles, masks, and respirators. Face shields have been used in the past and are currently being used as PPE to reduce exposure to bioaerosols and as a barrier device for eye protection. Reference Setti, Passarini and de Gennaro2 During the COVID-19 pandemic, 70% of US hospitals reported requiring eye protection. Reference Rhee, Baker and Klompas3 Thus, it is crucial that we understand the effectiveness of a disposable face shield at reducing virus containing aerosol concentrations in the breathing zone of an HCP when they are exposed to coughing patients as part of effectively protecting workers in a high-risk environment.

Coughing machines have been previously developed to study bioaerosols and particles produced when a human coughs. Reference Lindsley, King and Thewlis4,Reference Wang, Fu and Chao5 We developed a coughing machine to study the effectiveness of a disposable plastic shield at reducing HCP exposure to SARS-CoV-2 via aerosol. MS2 bacteriophage has been used as a safe surrogate for the aerosol transmission of virus. Reference Bae and Schwab6Reference Sala-Comorera9 MS2 is a single-stranded RNA bacteriophage that is nonenveloped. Reference Setti, Passarini and de Gennaro2

Prior studies with coughing machines have not analyzed the effectiveness of PPE with viable virus. Therefore, in the experiment performed in our study, we compared SARS-CoV-2 aerosol particle concentrations, through the surrogate MS2, across a face shield during a simulated cough. We analyzed virus concentrations and viability across a disposable plastic face shield placed on a simulated HCP during a simulated cough from a simulated patient and performed statistical analyses to compare the virus concentrations found across the face shield. We used these data to analyze the effectiveness of the face shield at reducing virus exposure.

Materials and methods

Power analysis

A power analysis was conducted a priori, with an effect size of 1, to determine the sample size needed to result in 80% power when using a paired t test with an α set at the 0.05 using a normal distribution. We performed the calculations using R Core Team 2020 (R: A language and environment for statistical computing, R Foundation for Statistical Computing, Vienna, Austria). The determined sample size was 16.7 for each group; therefore, at mimunm, 17 trials were conducted.

Coughing machine virus aerosolization

A coughing machine was designed to simulate a patient coughing viral aerosolized particles and exposing HCPs (article currently under review). The coughing machine operates by pushing 4.1 L of air, which is mixed with aerosolized virus through a mannequin’s mouth, producing a coughing plume. The resulting plume is like a human cough. A simulated HCP was placed 0.41 m in front of the cough to simulate exposure in a hospital or other care facility. A face shield was placed on the simulated HCP to mimic part of the PPE worn. MS2, the bacteriophage used as a surrogate for SARS-CoV-2 was placed in the liquid reservoir of the coughing machine and was aerosolized during the cough. Three consecutive simulated coughs were performed in each trial to mimic a “coughing fit.”

Virus and particle sampling

Virus and particle concentrations were sampled across the disposable plastic face shield (placed on a simulated HCP) while the coughing machine was operated. Two SKC biosamplers (no. 225-9595, SKC Inc, Eighty-Four, PA) were used to sample the MS2 (virus concentration) and 2 optical particle counters (OPCs) (GRIMM 11-D Aerosol Counters, nos. 619012 and 619025, Durag Group, Germany) were used to sample the particle concentration. The Biosamplers and OPCs were placed on the inside and the outside of the simulated HCP face shield (Fig. 1). The liquid media used in the biosampler was 20 mL phosphate-buffered saline (PBS, Gibco, 1x, pH 7.4 -/-, TX). The biosamplers were each connected to a pump that sampled at 12.5 liters per minute (LPM). The OPCs used a 1-second interval and sampled number concentration information across 16 channel (bin) sizes (range, 0.253–3.515 µm). The biosamplers and OPCs were started 1 minute prior to the first cough and operated nonstop throughout the 3 coughs (∼28 seconds).

Figure 1. Sampling setup. The disposable plastic face shield is placed on the simulated healthcare personnel with the coughing mannequin representing a patient with COVID-19. Biosamplers and optical particle counters (OPCs) are placed on the inside and the outside of the face shield to evaluate the healthcare personnel’s exposure to bioaerosols produced by coughing.

The sampling was conducted in an undisturbed chamber (8.5 m × 3.7 m × 2.4 m). Relative humidity and temperature were also monitored to ensure consistency across trials using a direct reading instrument (Extech Instruments Hygro-Thermometer Model SDL500, serial no. Z335535). After sampling, the chamber-like room and the coughing machine were cleaned with sodium hypochlorite wipes and 70% ethanol.

Virus sample analysis

Viability of the biosampler liquid media was assessed using a plaque assay. The sample was concentrated from ∼20 mL to <1,000 µL using an Amicon Ultra 15 centrifugal filter (Cork, Ireland), which was spun in the centrifuge at 4,000 rotations per minute (RPM) in 1-minute intervals. The concentrated sample volume was measured and recorded. A serial dilution was then conducted using 100 µL of the concentrated sample and 900 µL of MS2 broth (Appendix A). Petri dishes (Fisher Scientific, 100 mm × 15 mm, Polystyrene, MA) for the plaque assay were prepared using 10 mL bottom agar (Appendix A), which was left to rest for 30 minutes before any overlay was added. Top agar was prepared and placed in aliquots in 3-mL tubes (FalconBrand, 14-mL polystyrene round bottom tubes, Mexico), which were then incubated for 20 minutes at 45°C. After the incubation period, 10 µL liquid Escherichia coli (107 PFU/mL, ATCC15597) and 100 µL serial dilution sample were added to the 3 mL of top agar, which was mixed and poured over the solidified-bottom agar plates. Each serial dilution was plated in triplicate. The plates were left to solidify for 20 minutes and then sealed with a thermoplastic film (Parafilm wrapping film, Bemis, RPI, WI). The sealed plates were placed upside down in a 37°C incubator for 14–24 hours. After the incubation period, the MS2 plaques formed. The serial dilution plate that produced between 20 and 250 plaques were then counted. The arithmetic mean was calculated using the triplicate sample.

Viruses were quantified by counting plaque-forming units (PFU) per unit volume (eg, PFU/mL). Reference Langlet, Gaboriaud and Gantzer10,Reference Baer and Kehn-Hall11 Viable virus was the primary concern because it is needed to infect other individuals. Reference Letizia, Smith and Ge12 In the case of nosocomial infection, a patient must shed viable virus for the HCP to be infected. Reference Letizia, Smith and Ge12 The viable virus concentration was calculated from the PFU. The plaque count and dilution factor were used to calculate the plaque concentration (Eq. 1). The plaque concentration, concentrated volume, and air volume sampled were then used to determine the airborne concentration of MS2 (Eq. 2).

(1) $$\eqalign{{\rm{Plaque\;}}\,{\rm{Concentration\;}}\left( {{\rm{PFU}}} \right) =\ ({{{\rm{Average\;}}\,{\rm{plaque}}\,{\rm{forming\;}}\,{\rm{units}}}})/ \cr ({{{\rm{Total\;}}\,{\rm{volume}}\,{\rm{plated}}}})}$$
(2) $$\eqalign{{\rm{Sample\;Concentration}}\,{\rm{\;}}({\rm{PFU/}}{{\rm{m}}^3}) =\ {{{\rm{Plaque\;concentration}}\,{\rm{\;}}({\rm{PFU}})}}/\cr {{{\rm{Volume\;of\;air\;sampled\;}}\,{\rm{(}}{{\rm{m}}^3}{\rm{)}}}} }$$

After all 18 trials were completed, the normality of the samples was tested using the Shapiro-Wilks test. A box plot showing the mean, median, outliers, minimum, and maximum was computed across the 18 trials for the inside and outside of the face shield. A paired t test was performed to compare the 2 groups, with a critical P set to 0.05.

Particle sample analysis

The OPCs were connected to 2 computers operating the OPC Software (Spectrometer, GRIMM Control, 1179, V1-0-6, 25-10-2018) where particle data were recorded, stored, and exported to Microsoft Excel (Microsoft Office Professional Plus 2016, version 16.0.5278.1000, Redmond, WA) for analysis. The inside and outside particle concentration data for all the trials were compiled for each channel size. The Shapiro-Wilk test was used to assess for normality, after the data were log transformed. Any censored data (data that were recorded as zero) were assigned a value below the limit of detection (LOD, 5 particles/m3). 13 All OPC zero values were replaced with log (2.5) or 0.39794 (count/m3). Descriptive statistics were calculated using box plots to identify the median of the first quartile, the median of third quartile, standard deviations, and outliers for each experimental condition. For each variable, outliers were identified as values greater than 1.5 interquartile range (IQR) of the third quartile or less than 1.5 IQR of the first quartile. Paired t tests were also performed to examine the differences in number particle concentration between the inside and outside of the face shield. The paired t test was conducted using an unadjusted α of .05 as well as an α that was adjusted for false discovery rate using the Benjamini Hochberg procedure. Reference Benjamini and Hochberg14

Results

Virus concentration

The concentration of the MS2 aerosolized into the coughing machine was 1×109 PFU/m3. This concentration was verified after the aerosolization coughing event as well to ensure that it was the same. Air samples were “blank corrected” throughout the experiment. The calculated airborne concentration on the inside of the face shield, near the breathing zone of the HCP, had an arithmetic mean over the 18 trials of 3.54E+07 PFU/m3 (SD, 4.20E+07) (Table 1). The mean viable virus airborne concentration on the outside of the face shield was 1.15E+08 PFU/m3 (SD, 9.76E+07). There was a statistically significant reduction (P < .0006) in viable virus concentration on the inside of the face shield compared to the outside of the face shield, with a relative risk reduction of 69%.

Table 1. Arithmetic Mean (SD) Concentrations of Viable MS2 Virus on the Inside and the Outside of the Face Shield of the Healthcare Provider During a Simulated Coughing Event

Particle concentration

The number particle concentrations on the inside and outside of the face shield worn by the HCP across all sizes are shown in Figure 2. The particle concentration was significantly lower on the inside of the face shield compared to the outside of the face shield for 12 of the 16 particle sizes measured, using false discovery rate (FDR)–adjusted P values (P < .05) (Table 2). No significant difference was observed across distances for particle sizes of 0.800–0.943 µm, 0.943–1.112 µm, 1.310–1.545 µm, or 2.982–3.515 µm. Significant differences were observed across distances for particle sizes of 0.253–0.298 µm, 0.298–0.352 µm, 0.352–0.414 µm, 0.414–0.488 µm, 0.488–0.576 µm, 0.576–0.679 µm, 0.679–0.800 µm, 1.112–1.310 µm, 1.545–1.821 µm, 1.821–2.146 µm, 2.146–2.530 µm, and 2.530–2.982 µm.

Figure 2. Box plot of average particle concentration comparing the inside and the outside of the face shield placed on HCP exposed to patient cough. The inside of the face shield is solid filled and the outside of the shield is pattern filled. The data are organized by increasing channel size. Statistically significant pairs are marked with stars.

Table 2. Paired t Test Results for Comparing the Inside and Outside of the Face Shield Across Channel Size When the Face Shield is Placed on the Simulated HCP

Note. HCP, healthcare personnel; FDR, false discovery rate; df, degrees of freedom.

* Statistically significant (<.05).

Discussion

HCP can be exposed to aerosolized SARS-CoV-2 when treating infectious COVID-19 patients; therefore, we need to understand the effectiveness of face shields at protecting HCPs. We analyzed viable virus and particle concentrations across a disposable plastic face shield placed on an HCP during a simulated cough from a patient. We performed a statistical analysis to compare the virus concentrations found across the face shield to analyze the effectiveness of the face shield at reducing virus exposure. These results can be used to inform control measures for protecting HCPs. By performing the sampling experiment in a chamber like room, we were able to limit outside variables, such as drafts, from ventilation and foot traffic to ensure that our results pertained only to the aerosolization event itself.

We observed a statistically significant difference in airborne virus concentration between the inside and outside of the face shield (P = .0006). Having a 7-fold reduction in viable virus concentration may not eliminate the risk of infection; however, it may reduce the severity of infection if there is a dose–response relationship; it may reduce the risk of infection by reducing a viral inoculum to below an infectious threshold; or it may increase the effectiveness of medical masks via stepwise reduction in viable aerosolized virus. We observed a statistically significant difference (P < .05) in particle number concentrations between the inside and outside of the face shield for most channel sizes. The observed outliers may be due to the variability in particle concentration when operating the coughing machine across experimental trials.

These results align closely with other studies reporting that face shields may not be the most effective method of reducing virus containing aerosols. Reference Lindsley15,Reference Roberge16 However, these studies did not analyze viable airborne virus concentrations, which has implications for disease transmission and effectiveness. Our study demonstrated that virus aerosol-maintained infectivity inside the face shield, representing a risk of disease transmission. Therefore, HCPs may need to wear respiratory protection when treating patients with an active respiratory virus infection presenting with cough.

Face shields protect the HCP from larger aerosolized particles by preventing exposure to the mucous membranes on the face (eg, eyes). However, this study was not designed to assess this exposure. Face shields worn as a barrier protection device will prevent HCP direct contact with splashes and sprays of potentially infectious materials. Reference Simmons17 A face shield also prevents the HCP from touching their eyes, nose, and mouth, which can also lead to indirect transmission and infection. 18 A face shield could still be protective if worn by HCPs when treating potentially infectious patients. A limitation of our study design was that we used a bluff body to represent the HCP instead of a breathing mannequin due to the unavailability of a simulated breathing machine. However, we chose a sampler that represented a breathing rate similar to humans to minimize this limitation. We expect that if a breathing machine had been implemented it would have driven the results toward the null because more mixing would have occurred. Future experiments could be done with a breathing machine as the HCP to improve upon these findings. In addition, the coughing machine, representing a simulated patient, produced an exhaled cough and did not inhale air. Future studies should generate a coughing machine that can both inhale and exhale.

Another limitation was using a plaque assay to determine the viable aerosolized virus concentration to estimate infection risk. Due to the mechanical nature of the coughing machine, some of the viable virus could be inactivated during aerosolization, simulated coughing and sampling. This error did not likely affect our experimental observations because we used the same methods to assess both the inside and the outside of the face shield. We also reduced the impact of experimental variability in the plaque assays by running each serial dilution in triplicate and averaging the concentration across trials.

MS2 may not be representative of SARS-CoV-2 virus. MS2 has been used successfully as a surrogate for SARS-CoV-2 as well as norovirus, calicivirus, and poliovirus. Reference Bae and Schwab6,Reference Dawson8,Reference Sala-Comorera9 In addition, MS2 is a positive-sense, single-stranded, RNA phage like SARS-CoV-2. Reference Valegård19 MS2 is smaller than SARS-CoV-2 (23–28 nm vs 100 nm), but the particles that transmit through coughing are likely much larger because they occur in conglomerations of liquid particles. Reference Valegård19 MS2 is also nonenveloped whereas SARS-CoV-2 is enveloped, meaning that it may have different biophysical properties such as susceptibility to dehydration. Future experiments should confirm the results of this study using an enveloped surrogate like Phi 6 or with SARS-CoV-2.

Also, we also did not correct for background aerosol concentration in the chamber, so some of the particles measured might have been from room background aerosol. This limitation did not affect the results of our experiment because we measured live virus concentrations across the face shield rather than solely relying on airborne particle concentrations.

In conclusion, a face shield worn on an HCP was effective at reducing exposure to virus containing aerosol from simulated coughing. Future studies should be conducted to evaluate other scenarios (eg, sneezing, tidal breathing etc) and PPE ensembles (eg, N95 and face shield combination) using simulated breathing machine. In addition, future studies quantifying the benefit that face shields provide for eye protection are warranted.

Acknowledgments

Financial support

This research was supported by the Centers for Disease Control and Prevention and the National Center for Emerging and Zoonotic Infectious Diseases (CDC/NCEZID)–University of Chicago (COVID supplement grant no. CK000557/CK18-001 to M.W. Nonnenmann) and CDC/NCEZID–University of Maryland (COVID supplement grant no. CK000557/CK18-001to M.W. Nonnenmann). These sponsors had no role in the study design, methods, analyses, and interpretation, or in preparation of the manuscript and the decision to submit it for publication.

Competing interests

All authors report no Competing interests relevant to this article.

Appendix A

MS2 Broth Solution Recipe

Makes 500mL. All solutions were autoclaved without added supplement at 121°C for 20 minutes then placed in a water bath (55°C) to cool. The supplement was added and mixed once the solution cooled.

Solution Recipe of Supplement

Makes 500 mL. Filter solution with 0.22-µm pore filter prior to use.

References

Van Damme, W, Dahake, R, van de Pas, R, Vanham, G, Assefa, Y. COVID-19: does the infectious inoculum dose–response relationship contribute to understanding heterogeneity in disease severity and transmission dynamics? Med Hypoth 2021;146:110431.10.1016/j.mehy.2020.110431CrossRefGoogle ScholarPubMed
Setti, L, Passarini, F, de Gennaro, G, et al. Airborne transmission route of COVID-19: why 2 meters/6 feet of inter-personal distance could not be enough. Int J Environ Res Public Health 2020;17:2932.10.3390/ijerph17082932CrossRefGoogle ScholarPubMed
Rhee, C, Baker, MA, Klompas, M. Survey of coronavirus disease 2019 (COVID-19) infection control policies at leading US academic hospitals in the context of the initial pandemic surge of the severe acute respiratory coronavirus virus 2 (SARS-CoV-2) omicron variant. Infect Control Hosp Epidemiol 2023;44:597603.CrossRefGoogle ScholarPubMed
Lindsley, WG, King, WP, Thewlis, RE, et al. Dispersion and exposure to a cough-generated aerosol in a simulated medical examination room. J Occupat Environ Hyg 2012;9:681690.10.1080/15459624.2012.725986CrossRefGoogle Scholar
Wang, C, Fu, S, Chao, CY. Short-range bioaerosol deposition and recovery of viable viruses and bacteria on surfaces from a cough and implications for respiratory disease transmission. Aerosol Sci Technol 2021;55:215230.CrossRefGoogle Scholar
Bae, J, Schwab, KJ. Evaluation of murine norovirus, feline calicivirus, poliovirus, and MS2 as surrogates for human norovirus in a model of viral persistence in surface water and groundwater. Appl Environ Microbiol 2008;74:477484.CrossRefGoogle Scholar
Bastin, G, Galmiche, A, Talfournier, F, et al. Aerobic conditions and endogenous reactive oxygen species reduce the production of infectious MS2 phage by Escherichia coli . Viruses 2021;13:1376.10.3390/v13071376CrossRefGoogle ScholarPubMed
Dawson, D, et al. Survival of viruses on fresh produce, using MS2 as a surrogate for norovirus. J Appl Microbiol 2005;98:203209.CrossRefGoogle ScholarPubMed
Sala-Comorera, L, et al. Decay of infectious SARS-CoV-2 and surrogates in aquatic environments. Water Res 2021:117090.CrossRefGoogle ScholarPubMed
Langlet, J, Gaboriaud, F, Gantzer, C. Effects of pH on plaque-forming unit counts and aggregation of MS2 bacteriophage. J Appl Microbiol 2007;103:16321638.CrossRefGoogle ScholarPubMed
Baer, A, Kehn-Hall, K. Viral concentration determination through plaque assays: using traditional and novel overlay systems. J Visual Exper 2014:e52065.Google ScholarPubMed
Letizia, AG, Smith, DR, Ge, Y, et al. Viable virus shedding during SARS-CoV-2 reinfection. Lancet Respir Med 2021;9:e56e57.10.1016/S2213-2600(21)00219-8CrossRefGoogle ScholarPubMed
Dustdecoder 11-D portable aerosol spectrometer, model no. 11-D Durag Group website. https://www.grimm-aerosol.com/products-en/dust-monitors/the-dust-decoder/11-d/. Accessed June 20, 2023.Google Scholar
Benjamini, Y, Hochberg, Y. Controlling the false discovery rate: a practical and powerful approach to multiple testing. J Royal Stat Soc B (Method) 1995;57:289300.10.1111/j.2517-6161.1995.tb02031.xCrossRefGoogle Scholar
Lindsley, WG, et al. Efficacy of face masks, neck gaiters and face shields for reducing the expulsion of simulated cough-generated aerosols. Aerosol Sci Technol 2021;55:449457.CrossRefGoogle ScholarPubMed
Roberge, RJ. Face shields for infection control: a review. J Occupat Environ Hyg 2016;13:235242.CrossRefGoogle ScholarPubMed
Simmons, M, et al. Bench evaluation: three face-shield CPR barrier devices. Respir Care 1995;40:618623.Google ScholarPubMed
Personal protective equipment (PPE): coaching and training frontline healthcare professionals. Centers for Disease Control and Prevention website. https://www.cdc.gov/infectioncontrol/pdf/strive/PPE103-508.pdf. Accessed June 2023.Google Scholar
Valegård, K, et al. The three-dimensional structure of the bacterial virus MS2. Nature 1990;345:3641.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. Sampling setup. The disposable plastic face shield is placed on the simulated healthcare personnel with the coughing mannequin representing a patient with COVID-19. Biosamplers and optical particle counters (OPCs) are placed on the inside and the outside of the face shield to evaluate the healthcare personnel’s exposure to bioaerosols produced by coughing.

Figure 1

Table 1. Arithmetic Mean (SD) Concentrations of Viable MS2 Virus on the Inside and the Outside of the Face Shield of the Healthcare Provider During a Simulated Coughing Event

Figure 2

Figure 2. Box plot of average particle concentration comparing the inside and the outside of the face shield placed on HCP exposed to patient cough. The inside of the face shield is solid filled and the outside of the shield is pattern filled. The data are organized by increasing channel size. Statistically significant pairs are marked with stars.

Figure 3

Table 2. Paired t Test Results for Comparing the Inside and Outside of the Face Shield Across Channel Size When the Face Shield is Placed on the Simulated HCP