A method for measuring the cleaning effect of flushing disinfectors

Published online by Cambridge University Press:  15 May 2009

Åke Cederberg
Affiliation:
Department of Clinical Bacteriology, Malmö General Hospital, S-214 01 Malmö, Sweden
Kai Österberg
Affiliation:
Department of Clinical Bacteriology, Malmö General Hospital, S-214 01 Malmö, Sweden
Rights & Permissions [Opens in a new window]

Summary

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

A method is presented with which the mechanical cleaning effect of flushing disinfectors can be estimated independently of the thermal disinfecting effect of the hot flushing water. This makes it possible to specify the demands to be placed on the disinfecting effect of flushing with water of 85°C or more.

Bacillus stearothermophilus spores suspended in faeces were used as indicators because of their non-sensitivity to the hot-water temperature. Their elimination by flushing could thus be attributed to the mechanical effect of the water and not to the disinfecting effect of the temperature. A simple bacteriological technique was used, and the elimination factor (EF) was calculated as the ratio of the number of micro-organisms in the contamination before and after flushing. By using flushing water below 50°C for 130 s the EF on a bedpan was about 104–106. The effect of flushing with water of the same temperature for only half that time was somewhat weaker and when the temperature was raised to 85 °C after half the flushing time the effect was somewhat stronger.

It can be presumed that the conventional disinfecting phase with hot (85 °C) water for about 45 s in the commonly used flushing units could be substantially shortened and the costs of their use thereby reduced.

Type
Research Article
Copyright
Copyright © Cambridge University Press 1980

References

REFERENCES

Kallings, L. O. (1970). In Report of the KOD-Committee of the County Council of Stockholm, no. 2/1970 (Investigation on Decontamination by Hot Water).Google Scholar
Koller, W. (1978). Untersuchungen Zur Reinigungs und Desinfektions-leistung dreier handelsüblicher Leibschüssel-Reinigungsgeräte. Hygiene + Medizin 3, 192–7.Google Scholar
Niléhn, B. (1972). A method for the quantitative microbiological check of heat decontaminators. Scandinavian Journal of Infectious Disease 4, 245–53.CrossRefGoogle ScholarPubMed
Spri (Swedish Institute for Planning and Rationalization of Health and Social Welfare Services) (1977). Specification 23501 (Washer and Flushing Disinfectors, General Requirements. Technical Description). Specification 23503 (Flushing Disinfectors). Specification 23504 (Washer and Flushing Disinfectors: Testing of Cleaning Efficiency) (1977). (To be published.) Specification 23505 (Washer and Flushing Disinfectors, Testing of Disinfecting Efficiency).Google Scholar
Williams, R. E. O., Blowers, R., Garrod, L. P., & shooter, R. A. (1966). Hospital Infection, Causes and Prevention. London: Lloyd-Luke.Google Scholar