Published online by Cambridge University Press: 15 May 2009
This work is concerned with the cleaning and disinfection by heat of stainless-steel and polypropylene bedpans, which had been soiled with either a biological contaminant, human serum albumin (HSA) labelled with technetium-99m 99m(Tc), or a bacteriological contaminant, streptococcus faecalis mixed with Tc-labelled HSA. Results of cleaning and disinfection achieved with a Test Machine and those achieved by procedures adopted in eight different wards of a general hospital are reported. Bedpan washers installed in wards were found to be less efficient than the Test Machine, at least partly because of inadequate maintenance. Stainless-steel and polypropylene bedpans gave essentially the same results.