Autoclaves and their dangers and safety in laboratories

Published online by Cambridge University Press:  15 May 2009

E. H. Gillespie
Affiliation:
Public Health Laboratory, Sheffield, and Trent Regional Health Authority
S. A. Gibbons
Affiliation:
Public Health Laboratory, Sheffield, and Trent Regional Health Authority
Rights & Permissions [Opens in a new window]

Summary

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the ‘Save PDF’ action button.

Using a laboratory downward displacement vertical autoclave with the help of thermocouples recorded on a 12 point multichannel strip recorder, the risk of failing to sterilize laboratory discard buckets has been demonstrated. The use of proper temperature and time controls can prevent this risk.

A load in a bucket with perforated sides is more easily sterilized than in a solid bucket. Wire baskets, where appropriate, facilitate the sterilizing practice. The addition of water to a bucket does not reduce the time of heating up.

It is desirable that sealed bottles of media should not be sterilized in simple downward displacement autoclaves, but if used, strict monitoring of temperatures and times is essential both in the heating up stage and especially in the cooling stage. The temperatures in bottles are slow to rise and very slow to fall. Bottles at high temperature 80–405° C. or over have a high internal pressure which can allow the bottles to explode when subjected to thermal shock if removed too early.

It is suggested that all laboratory autoclaves should have a load temperature simulator or similar device to control the temperature of the load during the cycle automatically. For the sterilization of fluid media, it is suggested that, in addition to a simulator there should be accelerated cooling to reduce damage to the media and, what is more important, to rapidly bring down the temperature and thus the internal pressure in the bottles to a safe level. The opening of the sterilizer door or lid should be automatically controlled by the load temperature simulator.

Type
Research Article
Copyright
Copyright © Cambridge University Press 1975

References

REFERENCES

Cruickshank, R., Duguid, J. P., Marmion, B. P. & Swain, R. H. A. (1973). Medical Microbiology, 12 ed., Edinburgh and London: Churchill Livingstone.Google Scholar
Perkins, J. F. (1969). Principles and Methods of Sterilisation. Illinois: Charles Thomas.Google Scholar
Rubbo, S. D. & Gardner, J. F. (1965). Review of Sterilisation and Disinfection. London: Lloyd-Luke Ltd.Google Scholar