Inappropriate shock delivery by an implantable cardioverter-defibrillator due to electrical interference with a refrigerator in a 4-year-old child
Published online by Cambridge University Press: 23 September 2021
Abstract
We report a patient with long QT syndrome who received an inappropriate implantable cardioverter-defibrillator shock due to electrical interference from a refrigerator. This electrical interference was mistakenly detected as an episode of ventricular fibrillation and ended with an inappropriate delivery of shock without any warning symptoms before.