Hostname: page-component-f554764f5-wjqwx Total loading time: 0 Render date: 2025-04-14T19:16:22.965Z Has data issue: false hasContentIssue false

Real-time biofluorescent particle counting compared to conventional air sampling for monitoring airborne contamination in orthopedic implant surgery

Published online by Cambridge University Press:  07 April 2025

Frans Stålfelt*
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
Josefin Seth Caous
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden Division Materials and Production, Methodology, Textiles, and Medical Technology, RISE Research Institutes of Sweden, Gothenburg, Sweden
Karin Svensson Malchau
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
Camilla Björn
Affiliation:
Division Materials and Production, Methodology, Textiles, and Medical Technology, RISE Research Institutes of Sweden, Gothenburg, Sweden
Maziar Mohaddes
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden Department of Clinical Sciences Lund – Orthopedics, Lund University, Lund, Sweden Department Orthopedics, Hässleholms Hospital, Hässleholm, Sweden
*
Corresponding author: Frans Stålfelt; Email: [email protected]

Abstract

Background:

Surgical site infection (SSI) following orthopedic surgery is a complication associated with morbidity and economic burden. Transmission of airborne bacteria that settle into surgical wounds constitutes a risk factor for SSIs. However, monitoring microbial contamination inside operating rooms with conventional methods is resource and time-consuming.

Aim:

This study aimed to assess correlation between a biofluorescent particle counter (BFPC) and conventional air sampling, to enable real-time monitoring of airborne contamination. Additionally, the study aimed to analyze correlation between particles near the surgical site and particles 1 meter away, to evaluate the feasibility of distance-based measurements.

Methods:

Correlation analysis was conducted to compare colony-forming units (CFU) collected using a Sartorius MD8 air sampler with biofluorescent viable particles detected by BioTrak 9510-BD, both positioned near the surgical site. Additionally, correlation between particle counts measured by AeroTrak 6510, positioned 1 meter away, and total particle counts measured by the BioTrak near the surgical site was evaluated. Sampling took place in two operating rooms: one with turbulent mixed airflow (TMA) and one with unidirectional airflow (UDAF).

Results:

Negligible to low correlation between biofluorescent particles and CFU was observed, both in UDAF (n = 100) and TMA (n = 22). However, strong correlation was found between BFPC and particle counter measurements of total numbers of particles (Rp = 0.634–0.769, P < .001).

Conclusion:

While BFPCs offer real-time monitoring of airborne contamination, their predictive ability for CFU levels remains uncertain. Yet, the strong correlation between particles in the surgical site and particles measured 1 meter away suggests feasibility to conduct future studies with larger cohorts.

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America

Introduction

Surgical site infection (SSI) after orthopedic implant surgery constitutes a serious complication, Reference Berrios-Torres, Umscheid and Bratzler1 where the incidence rates range from 0.76% to 1.24% for total hip arthroplasty (THA) and 0.88% to 1.28% for total knee arthroplasty (TKA). Reference Springer, Cahue, Etkin, Lewallen and McGrory2 Additionally, SSI leads to prolonged patient suffering and substantial economic ramification, imposing significant societal costs and necessitating additional healthcare resources. Reference Graf, Ott and Vonberg3 While the precise infectious pathways remain largely elusive, one possible way for bacteria-carrying particles to contaminate the surgical wound is via airborne transmission, Reference Lidwell, Lowbury, Whyte, Blowers, Stanley and Lowe4Reference Darouiche, Green and Harrington7 either directly or indirectly via surgical instruments or implants. Reference Seth Caous, Svensson Malchau and Petzold8,Reference Alfonso-Sanchez, Martinez, Martin-Moreno, Gonzalez and Botia9 Bacteria are typically sized between 1 and 2 µm per cell however often clustering together to form larger aggregates, which together with the carrying fragment of eg skin define the total size of the particle. Reference Noble, Lidwell and Kingston6 Identified risk factors for particle emission include the shedding of skin fragments from staff, door openings, and the type of clothing worn in operating rooms (ORs). Reference Noble10,Reference Andersson, Bergh, Karlsson, Eriksson and Nilsson11 This measure, complemented by other preventive actions, such as antibiotic prophylaxis, proper hygienic protocols, and skin preparation, are important to minimize the risk for SSI. Reference Calderwood, Anderson and Bratzler12

The conventional method for microbial air assessments in ORs involves active air sampling, where airborne bacteria are captured on agar plates, followed by incubation and quantification of colony-forming units (CFU). 13,14 While effective for detecting bacterial contamination near the surgical site, active air sampling is challenging to perform due to its intrusive equipment, which can distract surgeons. Reference Dharan and Pittet15 Additionally, obtaining results is both time- and resource-intensive, and requires specialized microbiological skills, often necessitating additional staff in the OR. In contrast, industries employing cleanroom technologies have long relied on automated real-time particle monitoring systems to maintain minimal levels of particulate contamination. Reference Scott, Vanbroekhoven and Joossen16,17 Despite similarly rigorous policies regarding airborne contamination in ORs, the widespread implementation of such systems in healthcare remains limited. This hesitation is primarily attributed to the lack of robust evidence linking particulate levels to CFUs. 14 Establishing evidence could enable real-time surveillance in ORs, facilitating timely interventions to reduce microbial contamination. Reference Lidwell, Lowbury, Whyte, Blowers, Stanley and Lowe4,Reference Noble, Lidwell and Kingston6,Reference Whyte, Lidwell, Lowbury and Blowers18,Reference Friberg, Friberg and Burman19

In a previously published systematic review examining the relationship between particulate levels and CFUs, the yielded results were mixed, making it difficult to establish a definitive correlation. Reference Stalfelt, Svensson Malchau, Bjorn, Mohaddes and Erichsen Andersson20 While most prior research in ORs has utilized optical particle counters (PCs) to explore the association between total particulate levels and CFUs, modern biofluorescent particle counters (BFPCs) offer a promising alternative to detect viable particles. Although BFPCs have significant potential for microbial surveillance, Reference Jeys, Herzog, Hybl, Czerwinski and Sanchez21Reference Sandle24 correlation between biofluorescent particles and CFU have not been extensively studied in clinical settings. A study by Dai et. al explored this relationship and found a significant correlation (Pearson’s correlation coefficient = 0.76). However, the limited sample size from this study emphasizes the need for further research to validate these findings during ongoing surgeries. Reference Dai, Zhang and Ma25

The aim of this study was to assess the correlation between biofluorescent particles (BFP) and CFU during orthopedic implant surgeries, to facilitating real-time monitoring of air bacterial contamination during surgical procedures. A secondary aim was to analyze the correlation between total particle counts measured by BFPCs near the surgical wound and total particle counts obtained by a conventional optical PC positioned 1 meter away, to assess the feasibility of accurate measurements at greater distances for future studies.

Method

Study information

This observational investigation took place during the spring of 2022 at Sahlgrenska University Hospital, Mölndal, involving two ORs for orthopedic implant surgery; one equipped with vertical unidirectional airflow (UDAF) ventilation and the other with turbulent mixed airflow (TMA) ventilation. Eligible surgical procedures performed on weekdays were included in the study after approval by the lead scrub nurse. The study included THAs and TKAs performed in the UDAF-ventilated OR, as well as hip hemiarthroplasties (HHAs) conducted in the TMA-ventilated OR. Air sampling procedures adhered to the Swedish Institute of Standards (SIS) technical specification (TS) SIS-TS 39:2015. 14 The OR team consisted of a lead surgeon, an assistant surgeon, a scrub nurse, an assisting nurse, and an anesthetic nurse. All staff wore surgical attire made of a 50/50 cotton/polyester blend, along with tucked-in surgical hoods. Additionally, the surgeons and the scrub nurse wore disposable non-woven surgical gowns (Mölnlycke, Barrier Surgical Gown), double sterile gloves, and face masks. Ethical approval for the study was granted by the Swedish Ethical Review Authority (2021–01689), and the study is registered on ClinicalTrials.gov (NCT05816135).

Materials and instrumental setup

Bacterial samples were acquired using the active air sampler Sartorius MD8 (Sartorius Lab Instruments GmbH & Co. KG, Göttingen, Germany) operating at a flow rate of 100 L/min for a duration of 10 minutes, equivalent to 1 m3/10 min. A sterilized tube (PVC, L = 1.5 m, ⌀ = 10 mm) was affixed to the Sartorius MD8, positioned approximately 0.3 m from the surgical wound (Figure 1A). A sterilized mouthpiece was fixated to the tube’s end, where the filtered air samples impinged upon gelatin filters (Sartorius Lab Instruments GmbH & Co. KG) featuring pore sizes of 3 µm. Following the collection of 1 m3 of filtered air, the gelatin filter was aseptically detached by the scrub nurse and handed to a trained professional proficient in environmental sampling techniques, who then transferred the filter to nonselective horse blood agar plates (Substrate Department, Sahlgrenska University Hospital) as recommended by SIS-TS 39:2015. 14,Reference Lans, Mathijssen, Goswami, van den Dobbelsteen, Luscuere and van der Elst26 Post-surgery, the samples were taken to the biosafety level 2 research laboratory at the Department of Orthopedics, Sahlgrenska University Hospital and incubated at 35 ± 2 °C and inspected after 2 and 5 days for CFU count. 14,Reference Lans, Mathijssen, Goswami, van den Dobbelsteen, Luscuere and van der Elst26 Depending on the duration of the procedure (typically 45–70 minutes), 4–6 air samples were collected from incision to wound closure. The results for each air sample were expressed as colony-forming units per cubic meter (CFU/m³). Identification of bacterial types was not performed.

Figure 1. A) Instrumental setup showing the biofluorescent particle counter probe alongside the tube and mouthpiece of the Sartorius MD8 air sampler, equipped with a gelatin membrane filter. B) Schematic of the instrument placement inside the operation room (OR) with vertical unidirectional airflow (UDAF) ventilation. The circled area indicates the aerial safe zone generated by the vertical UDAF. Similar instrumental placement was used in the turbulent mixed airflow ventilated OR, with the exception that no additional particle counter was used.

Viable particle count was acquired utilizing the BFPC BioTrak 9510-BD (TSI, Minnesota, USA). The BFPC was connected to a sterilized polyvinyl chloride tube (L = 1.5 m, ⌀ = 8 mm) and positioned in close proximity to the headpiece of the Sartorius MD8 (Figure 1A), in order to ensure comparable samples. The configuration follows the manufacturer’s recommendations and complies with international standard guidelines. 14,17,27 Additionally, the setup was verified in consultation with a cleanroom measurement expert to ensure accuracy and adherence to best practices. Air sampling was conducted continuously throughout the surgical procedure, from the initial incision to the closure of the surgical wound, to evaluate the clinical applicability of the sampling method. The BFPC collected samples at a rate of 0.0283 m3/min, with data logged at one-minute intervals. For the analysis, the collected data used was the mean value to match the 10-minute interval of the Sartorius MD8. Results were automatically converted into the unit of number of particles per cubic meter (m3) for comparability with the results from Sartorius MD8. The BFPC detected particle sizes of 0.5, 0.7, 1, 3, 5, and 10 µm, both viable and total count. However, particle sizes of 0.5 and 0.7 µm were excluded from the analysis due to smaller particles being less likely to carry bacteria. Reference Noble, Lidwell and Kingston6 The BioTrak aggregates particle counts for all particles equal to or greater than a defined size threshold, with the reported count reflecting the total number of particles within the specified size range and above.

A conventional PC, AeroTrak 6510 (TSI, Minnesota, USA), was positioned 1 meter away from the surgical table, within the periphery of the surgical zone (Figure 1B). The AeroTrak operated at a flow rate of 28.3 L/min and employed photodetection for particle count analysis. In contrast to the BioTrak 9510-BD, this particle counter only detected particles of sizes 0.5, 0.7, 1, and 5 µm. Particle sizes were compared consistently with the respective particle channel for the PC and the BFPC (1 and 5 µm). Continuous data sampling was systematically executed from the incision phase to the surgical wound closure, the same as the protocol employed for the BioTrak. Notably, data of conventional particle counting was only accessible for UDAF ventilation, due to the AeroTrak being already installed in this OR.

Statistical methods

The statistical analysis was conducted utilizing IBM SPSS Statistics version 29.0.0.0. The particle measurements were divided into time intervals corresponding to the CFU measurements, with mean values used for both the BFPC and the conventional PC for the correlation analysis. To facilitate the interpretation of the descriptive analysis, the particulate data were transferred into a logarithmic scale. Pearson’s correlation coefficient (Rp) was applied for parametric data. Assumptions were tested using the Shapiro-Wilk test and via visualization through distribution- and QQ-plots. Correlations were ranked accordingly: Negligible correlation (R < 0.1), weak correlation (0.1 ≤ R ≤ 0.39), moderate correlation (0.4 ≤ R ≤ 0.7), and strong correlation (R > 0.7). Reference Schober, Boer and Schwarte28

Results

Sample collection and characteristics

In total, the study observed 22 surgical procedures, including THA (n = 11), TKA (n = 7) and HHA (n = 4). THA and TKA procedures were conducted in an OR equipped with UDAF ventilation, while HHA was performed in an OR utilizing TMA ventilation. A total of 122 paired samples of CFU and BFP were collected from the two ORs, with a distribution of 100 samples in UDAF and 22 in TMA. Across all surgical procedures and ORs, the concentration of smaller airborne particles consistently exceeded that of larger particles. Notably, the mean CFU/m3 was significantly higher in samples collected from TMA-ventilated ORs compared to UDAF-ventilated ORs, as shown in Table 1.

Table 1. Descriptive statistics of colony-forming unit (CFU) and logarithmic particle levels of all measured particle sizes

IQR, interquartile range.

* Not applicable.

Correlation between BFP, PC and CFU in UDAF and TMA

Correlation analysis between the measured concentrations of CFU and BFP of sizes 1 µm, 3 µm, 5 µm, and 10 µm, in UDAF-ventilated OR showed an overall negligible to weak correlation, with no statistical significance. Similarly, a negligible correlation was observed between particles of sizes 1 µm and 5 µm measured by the PC, and CFU. In TMA-ventilated operating rooms the correlation analysis of BFP/m3 and CFU/m3 showed week correlation for all particle sizes, as seen in Table 2.

Table 2. Correlation coefficients in unidirectional airflow (UDAF)-ventilated operation room (OR) (n = 100) and turbulent mixed airflow (TMA)-ventilated OR (n = 22) between colony-forming unit (CFU) and BFP and particle counter (PC), and between total particle count measured by the biofluorescent particle counter (BFPC) and the PC

Results are displayed as Pearsons’s Correlation Coefficient (95% Confidence Interval).

Correlation significant at the 0.05 level (two-tailed).

a Total particle count.

Correlation between BFPC and PC in the UDAF-ventilated OR

The correlation analysis between total particle counts using BFPC and PC, positioned close to the surgical wound and further away from the operating table, respectively, displayed a strong significant correlation for 1 µm, Rp = 0.769, P < .001 and a moderate significant correlation for 5 µm, Rp = 0.634, P < .001, as outlined in Table 2 and visualized in Figure 2.

Figure 2. Scatter plot of total count of particles (1 µm) measured with a biofluorescent particle counter positioned close to the surgical site and with a conventional particle counter, positioned approximately 1m from the surgical site in a unidirectional airflow-ventilated OR.

Discussion

SSI following orthopedic implant surgery poses significant risks to patient outcomes and healthcare resources. This study evaluated the accuracy of using a BFP to measure airborne bacteria levels during surgery and examined the correlation between CFU and biofluorescent particles BFP under different ventilation conditions. The results indicated a weak correlation between BFP and CFU in both UDAF and TMA-ventilated ORs for all particle sizes. These findings align with the results from a systematic review of the correlation between particles and CFU during ongoing surgery, revealing no discernible evidence for correlation between particles and CFU. Reference Stalfelt, Svensson Malchau, Bjorn, Mohaddes and Erichsen Andersson20 The small sample size in the TMA-ventilated OR likely contributed to the weak and unsignificant correlation between CFU and BFP. Also, the low CFU levels observed in the UDAF-ventilated OR suggests that more samples are needed to facilitate a more robust correlation analysis.

However, mean CFU levels were higher in TMA-ventilated ORs, reinforcing the importance of airflow control, with UDAF systems providing better air quality. This also aligns with previous research indicating that UDAF systems create a more controlled environment with lower bacterial counts due to the establishment of an aerial safe zone over the operating table. Furthermore, the moderate and strong correlation between BFPC and conventional PC suggests that PCs can reliably measure total particle counts even when placed away from the surgical site. This could streamline monitoring without distracting surgeons or requiring additional staff involvement in future research.

The limitations of this study should be considered when interpreting the results. Firstly, the sample size, particularly in TMA-ventilated ORs (n = 22), was small, potentially limiting statistical power and generalizability. The results from the TMA-ventilated OR should therefore be used with great caution. Furthermore, the COVID-19 pandemic led to a reduction in the overall number of surgeries and measurements initially planned for inclusion, with 25 procedures per ventilation setting similar to other comparable studies. Reference Stalfelt, Svensson Malchau, Bjorn, Mohaddes and Erichsen Andersson20 This limitation may have introduced bias by restricting the capture of the full range of operating room conditions. The lack of randomization, where elective surgeries were performed in UDAF-ventilated ORs and trauma surgeries in TMA-ventilated ORs, introduces potential confounders such as variations in door openings, staff presence, and surgery duration, making direct comparison between the two ventilation systems difficult. Additionally, the AeroTrak PC was only used in UDAF-ventilated ORs, restricting comparisons in particle counts between ventilation types. Overall, larger-scale, randomized studies are needed to confirm these findings and provide more robust evidence regarding the impact of ventilation systems on air cleanliness and SSI.

Conclusion

This study underscores the complexity of airborne bacterial transmission during orthopedic surgeries, emphasizing the influence of ventilation, airflow dynamics, and environmental factors. Controlled environments, such as UDAF-ventilated operating rooms, demonstrated lower bacterial contamination compared to TMA-ventilated ORs. While BFPCs offer real-time monitoring, their correlation with CFU is limited and should be used with caution. However, the strong correlation between total particle counts measured near and further away from the surgical site suggests the potential for less intrusive monitoring. Future improvements in BFPC technology, along with identifying surgical stages that generate high particle counts, could enhance infection control strategies.

Supplementary material

The supplementary material for this article can be found at https://doi.org/10.1017/ash.2025.61

Financial support

The funding of FS and the particle counters utilized in this study (the BioTrak 9510-BD and the AeroTrak 6510) were provided by Getinge Group (Getinge, Sweden). Getinge Group were not involved in the planning, methodology, data retrieval, or interpretation of results and conclusions generated from this study.

Competing interests

Neither Getinge Group nor the authors have any affiliations with the manufacturer (TSI, Minnesota, USA) of these particle counters. The authors declare no conflicts of interest.

References

Berrios-Torres, SI, Umscheid, CA, Bratzler, DW, et al. Centers for disease control and prevention guideline for the prevention of surgical site infection, 2017. JAMA Surg 2017;152:784791.CrossRefGoogle Scholar
Springer, BD, Cahue, S, Etkin, CD, Lewallen, DG, McGrory, BJ. Infection burden in total hip and knee arthroplasties: an international registry-based perspective. Arthroplast Today 2017;3:137140.Google ScholarPubMed
Graf, K, Ott, E, Vonberg, RP, et al. Surgical site infections–economic consequences for the health care system. Langenbecks Arch Surg 2011;396:453459.CrossRefGoogle ScholarPubMed
Lidwell, O, Lowbury, E, Whyte, W, Blowers, R, Stanley, S, Lowe, D. Airborne contamination of wounds in joint replacement operations: the relationship to sepsis rates. J Hosp Infect 1983;4:111131.CrossRefGoogle ScholarPubMed
Persat, A, Nadell, CD, Kim, MK, et al. The mechanical world of bacteria. Cell 2015;161:988997.CrossRefGoogle ScholarPubMed
Noble, WC, Lidwell, OM, Kingston, D. The size distribution of airborne particles carrying micro-organisms. J Hyg (Lond) 1963;61:385391.Google ScholarPubMed
Darouiche, R, Green, D, Harrington, M, et al. Association of airborne microorganisms in the operating room with implant infections: a randomized controlled trial. Infect Control Hosp Epidemiol 2017;38:310.Google ScholarPubMed
Seth Caous, J, Svensson Malchau, K, Petzold, M, et al. Instrument tables equipped with local unidirectional airflow units reduce bacterial contamination during orthopedic implant surgery in an operating room with a displacement ventilation system. Infect Prev Pract 2022;4:100222.Google Scholar
Alfonso-Sanchez, JL, Martinez, IM, Martin-Moreno, JM, Gonzalez, RS, Botia, F. Analyzing the risk factors influencing surgical site infections: the site of environmental factors. Can J Surg 2017;60:155161.CrossRefGoogle ScholarPubMed
Noble, WC. Dispersal of skin microorganisms*. Br J Dermatol 1975;93:477485.CrossRefGoogle ScholarPubMed
Andersson, AE, Bergh, I, Karlsson, J, Eriksson, BI, Nilsson, K. Traffic flow in the operating room: an explorative and descriptive study on air quality during orthopedic trauma implant surgery. Am J Infect Control 2012;40:750755.Google Scholar
Calderwood, MS, Anderson, DJ, Bratzler, DW, et al. Strategies to prevent surgical site infections in acute-care hospitals: 2022 update. Infect Control Hosp Epidemiol 2023;44:695720.Google ScholarPubMed
Standardization GIf. DIN 1946-4. In: Ventilation in Buildings and Rooms of Health Care. Plzen: European Standard; 2018.Google Scholar
Institute SS. SIS-TS 39:2015. In: Microbiological Cleanliness in the Operating Room – Preventing Airborne Contamination – Guidance and Fundamental Requirements. Stockholm: Swedish Institute for Standards; 2015.Google Scholar
Dharan, S, Pittet, D. Environmental controls in operating theatres. J Hosp Infect 2002;51:7984.Google ScholarPubMed
Scott, A, Vanbroekhoven, A, Joossen, C, et al. Challenges encountered in the implementation of bio-fluorescent particle counting systems as a routine microbial monitoring tool. PDA J Pharm Sci Technol 2023;77:29.CrossRefGoogle ScholarPubMed
Standardization IOf. ISO 14644 - Cleanrooms and Associated Controlled Environments. Geneva: International Organization for Standardization; 2015.Google Scholar
Whyte, W, Lidwell, O, Lowbury, E, Blowers, R. Suggested bacteriological standards for air in ultraclean operating rooms J Hosp Infect 1983;4:133139.Google ScholarPubMed
Friberg, B, Friberg, S, Burman, LG. Inconsistent correlation between aerobic bacterial surface and air counts in operating rooms with ultra clean laminar air flows: proposal of a new bacteriological standard for surface contaminatio. J Hosp Infect 1999;42:287293.Google Scholar
Stalfelt, F, Svensson Malchau, K, Bjorn, C, Mohaddes, M, Erichsen Andersson, A. Can particle counting replace conventional surveillance for airborne bacterial contamination assessments? A systematic review using narrative synthesis. Am J Infect Control 2023;51:14171424.Google ScholarPubMed
Jeys, TH, Herzog, WD, Hybl, JD, Czerwinski, RN, Sanchez, A. Advanced trigger development. Linc Lab J 2007;17:2962.Google Scholar
Sandle, T, Leavy, C, Jindal, H, Rhodes, R. Application of rapid microbiological methods for the risk assessment of controlled biopharmaceutical environments. J Appl Microbiol 2014;116:14951505.CrossRefGoogle ScholarPubMed
Behrens, D, Schaefer, J, Keck, CM, Runkel, FE. Application of biofluorescent particle counters for real-time bioburden control in aseptic cleanroom manufacturing. Appl Sci 2022;12:8108.CrossRefGoogle Scholar
Sandle, T. Real-time counting of airborne particles and microorganisms: a new technological wave? Clean Air Containment Rev 2012;9:46.Google Scholar
Dai, C, Zhang, Y, Ma, X, et al. Real-time measurements of airborne biologic particles using fluorescent particle counter to evaluate microbial contamination: results of a comparative study in an operating theater. Am J Infect Control 2015;43:7881.Google Scholar
Lans, JLA, Mathijssen, NMC, Goswami, PR, van den Dobbelsteen, JJ, Luscuere, PG, van der Elst, M. Baseline study of ultra-clean air change rate, number, and type of microorganisms and level of particles during trauma surgery. HERD 2024:19375867241302406.Google ScholarPubMed
GmbH SLI. Operating Instructions - Air Sampler MD8 Airscan. Göttingen: Sartorius Stedim Biotech GmbH; 2011.Google Scholar
Schober, P, Boer, C, Schwarte, LA. Correlation coefficients: appropriate use and interpretation. Anesth Analg 2018;126:17631768.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. A) Instrumental setup showing the biofluorescent particle counter probe alongside the tube and mouthpiece of the Sartorius MD8 air sampler, equipped with a gelatin membrane filter. B) Schematic of the instrument placement inside the operation room (OR) with vertical unidirectional airflow (UDAF) ventilation. The circled area indicates the aerial safe zone generated by the vertical UDAF. Similar instrumental placement was used in the turbulent mixed airflow ventilated OR, with the exception that no additional particle counter was used.

Figure 1

Table 1. Descriptive statistics of colony-forming unit (CFU) and logarithmic particle levels of all measured particle sizes

Figure 2

Table 2. Correlation coefficients in unidirectional airflow (UDAF)-ventilated operation room (OR) (n = 100) and turbulent mixed airflow (TMA)-ventilated OR (n = 22) between colony-forming unit (CFU) and BFP and particle counter (PC), and between total particle count measured by the biofluorescent particle counter (BFPC) and the PC

Figure 3

Figure 2. Scatter plot of total count of particles (1 µm) measured with a biofluorescent particle counter positioned close to the surgical site and with a conventional particle counter, positioned approximately 1m from the surgical site in a unidirectional airflow-ventilated OR.

Supplementary material: File

Stålfelt et al. supplementary material

Stålfelt et al. supplementary material
Download Stålfelt et al. supplementary material(File)
File 20 KB