Hostname: page-component-586b7cd67f-dsjbd Total loading time: 0 Render date: 2024-11-23T12:55:01.350Z Has data issue: false hasContentIssue false

A Polycrystalline Thin Film Diffractometer for Asymmetric Diffraction Using Parallel Beam and High Resolution Parallel Slits

Published online by Cambridge University Press:  06 March 2019

H. Toraya
Affiliation:
Ceramics Research Laboratory, Nagoya Institute of Technology Asahigaoka, Tajimi 507, Japan
J. Yoshino
Affiliation:
Ceramics Research Laboratory, Nagoya Institute of Technology Asahigaoka, Tajimi 507, Japan
Get access

Abstract

A polycrystalline thin film diffractometer for asymmetric diffraction at a fixed incident angle using parallel beam and high-resolution-parallel slits has been developed. Well collimated CuK α1 incident beam was obtained by coupling of a rotating anode generator with a channel-cut crystal monochromator. The diffracted beam was analyzed with parallel slits with 0.17° angular aperture. The observed profiles of CeO2 standard powder sample were analyzed by using an individual profile fitting technique. The minimum full-width at half-maximum (FWHM) was 0.15° in the 20 range of 40° ∼ 70°, and the variation of FWHM with 20 was almost independent of the incident angle. Intensity in asymmetric diffraction was enhanced compared to that in the symmetric 0-20 scan, and its variation was well in accordance with the theory. The use of the parallel slits compromizes the intensity and resolution for polycrystalline thin film diffraction in the present setup, but gives the high quality of diffraction data for structural study.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1995

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Feder, R. & Berry, B. S. (1970). J. Appl. Cryst. 3, 372379.Google Scholar
2. Matyi, R. J. (1986). Adv. X-ray Anal, 29, 353-366.Google Scholar
3. Iyengar, S. S., Santana, M. W., Windischmann, H. & Engler, P. (1987). Adv. X-ray Anal. 30, 457464.Google Scholar
4. Larsen, R. A., McNulty, T. F., Goehner, R. P. & Crystal, K. P. (1988). Adv. X-ray Anal. 32, 311321.Google Scholar
5. Goehner, R. P. & Eatough, M. O. (1992). Powder Diffraction 7, 25.Google Scholar
6. Segmuller, A. (1986). Adv. X-ray Anal. 29, 353366.Google Scholar
7. Huang, T. C. (1990). Adv. X-ray Anal. 33, 91107.Google Scholar
8. Parrish, W. & Hart, M. (1987). Zeit. Krist. 179, 161173.Google Scholar
9. Hart, M. (1991). Mater. Sci. Forum79-82, 447454.Google Scholar
10. Toraya, H., Huang, T. C. & Wu, Y. (1993). Adv. X-ray Anal. 36, 609616.Google Scholar
11. Toraya, H., Huang, T. C. & Wu, Y. (1993), J. Appl. Cryst. 26, 774777.Google Scholar
12. Toraya, H. & Yoshino, J. (1994). J. Appl. Cryst. 27, 961966.Google Scholar
13. Hart, M. (1992). Adv. X-ray Anal. 35, 329-332.Google Scholar
14. Toraya, H. (1986). J. Appl. Cryst. 19, 440447.Google Scholar
15. Hastings, J. B., Thomlinson, W. & Cox, D. E. (1984). J. Appl. Cryst. 17, 8595.Google Scholar
16. Caglioti, G., Paoletti, A. & Ricci, F. P. (1958). Nucl. Instrum. 3, 223228.Google Scholar
17. Toraya, H. & Okuda, T. (1995). J. Phys. Chem. Solids. 56, 13171322.Google Scholar
18. Kogan, V., Bethke, J. & Verman, B. (1995). Abstract in Denver X-ray Conference, 123.Google Scholar
19. Schuster, M. & Goebel, H. (1995). Abstract in Denver X-ray Conference, 145.Google Scholar