Numerical Calculation of Axial Divergence Profiles
Published online by Cambridge University Press: 06 March 2019
Abstract
The diffractometer instrumental profiles convoluted by Alexander from aberration profiles calculated using the methods developed Easterbrook, Pike and Wilson include two profiles which are not amenable to exact analytical solution. These two profiles are generated by the axial divergence and flat specimen aberrations. Using ray tracing techniques the exact profiles of these two aberrations have been computed for the Philips PW1050 and Scintag DA5 diffractometers. sample profiles at 2θ = 28°, 90° and 152° have been convoluted with the third asymmetric profile, that due to specimen transparency, and plots of the resultant profiles are shown. Shifts of the centroids and the peaks of the profiles from the true Bragg angle have also been computed and plotted over the range 2θ = 0°-180°.