Correction for Non-Linearity of Proportional Counter Systems in Electron Probe X-Ray Microanalysis

Published online by Cambridge University Press:  06 March 2019

Kurt F. J. Heinrich
Affiliation:
National Sureau of Standards, Washington, D.C.
Donald Vieth
Affiliation:
National Sureau of Standards, Washington, D.C.
Harvey Yakowitz
Affiliation:
National Sureau of Standards, Washington, D.C.
Get access

Abstract

While the theoretical basis for the correction of non-linearity of detector systems is well known, methods for the determination of dead-time effects must be adapted to electron probe microanalyzer systems. Two such methods, one employing both X-ray and current measurements and the other employing simultaneous X-ray measurements on two spectrometers, are described. The effect of pulse-height shrinkage at high counting rates on the linearity of the detector system is discussed. When the proposed corrections for the dead-time of X-ray detector systems employing proportional counters are applied to the X-ray intensity measurements obtained with the electron probe microanalyzer, count rates as high as 50,000 counts/sec can be used.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1965

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Friedman, H., Electronics 18: 132, 1945.Google Scholar
2. Lonsdale, K., Acta Cryst. 1: 12, 1948.Google Scholar
3. Schrff, L. J., Phys. Rev. 50: 88, 1936.Google Scholar
4. Ruark, A. and Brammer, F. E., Phys. Rev., 52: 322, 1937.Google Scholar
5. Hole, N., Arkiv. Mat., Astro. Pys. 33A (H), 1947.Google Scholar
6. Jost, R., Helv. Phys. Acta 20: 173, 1947.Google Scholar
7. Beckman, M. and Michiels, J. L., Proc. Phyt. Soc. (London) 60: 594, 1948.Google Scholar
8. Evans, R. D., The Atomic Nucleus, MacGraw-Hill Book Company, Inc., New York, 1955.Google Scholar
9. Wittry, D. B., in: X-Ray and Electron Probe Analysis, ASTM Special Technical Publication No. 349, ASTM, 1963.Google Scholar
10. Short, M. A., Rev. Sci. Inst. 31: 618, 1960.Google Scholar
11. Bender, S. L. and Rapperport, E. J., “The Electron Microprobe,” in: Proceedings of the Symposium on Electron Probe Microanalysis, Electrochemical Society, John Wiley & Sorts, Inc., New York, 1966.Google Scholar
12. Heinrich, K. F. J., in: W. M. Mueller, G. R. Mallett, and M. J. Fay (eds.), Advances in X-Ray Analysis, Vol 7, Plenum Press, New York, 1963, p. 325.Google Scholar
13. Petzold, W., Z. Angeio. Phys. 15: 158, 1963.Google Scholar
14. Heinrich, K. F. J., to be published.Google Scholar