Application of Tapered Monocapillary in a Laboratory Mxrf Set-Up
Published online by Cambridge University Press: 06 March 2019
Abstract
Simulation and experimental work that compare the performance of straight and tapered monocapillaries when used with laboratory x-ray sources are reported. Detailed simulations for various taper profiles give several important conclusions for optimizing the design of a tapered monocapillary. Several tapered monocapillaries were prepared. With a 16W x-ray source, beam intensities of 4×105 photon/sec/μm2 and 3×105photon/sec/μm2 of Cu Kα x rays were obtained from the tapered monocapillaries for output diameters of 8μm and 3.5μm, respectively. These intensities are 1.4 and 1.5 times that obtained from straight capillaries with the same output beam sizes at the experimental set-up optimized for a straight capillary. In addition to the gain in x-ray flux, the tapered monocapillaries produce output beams with significantly reduced high energy bremsstrahlung radiation and increased flux stability with respect to shifts of the x-ray source spot.