The Recessed Source Geometry for Source Excited X-Ray Fluorescence Analysis
Published online by Cambridge University Press: 06 March 2019
Abstract
Different geometries are considered for source excited energy-dispersive X-ray fluorescence (EDXRF) analysis Systems, including the recessed source geometry introduced in the present work. The calculated physical excitation-detection efficiencies, for the side (or annular), central, receded and recessed source geometries are presented as a function of the target to source distance, for Ca, K, S and Si targets excited with a Fe-55 XBF-3 X-ray source and xenon filled gas proportional scintillation counters. The last two geometries present in gênerai the highest efficiencies. The recessed source geometry présent the best performance with peak efficiencies a factor of 3.3 better than those for the standard side or annular source geometries.