Article contents
Flat Crystal X-Ray Optics
Published online by Cambridge University Press: 06 March 2019
Abstract
An investigation was undertaken by the Bureau of Mines at College Park, Md., to determine the effect of various combinations of collimators, analyzing crystals and detectors on line intensities, line-to-backgrouhd ratios, and spectral resolution. The research showed that line broadening due to mosaic crystal surfaces was greatly reduced and line splitting from faults was eliminatedby the use of two fine collimators (0.005 inch, spacing, 4 inch length). Line intensities were reduced, but lineto- background ratios arid line profiles were substantially improved with double collimators. Pulse height discrimination resulted in marked improvement in the line-to-background ratio in the long-wave length region, 2 to 10 A, but was much less effective for shorter wave lengths.
- Type
- Research Article
- Information
- Copyright
- Copyright © International Centre for Diffraction Data 1957
References