We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
The purpose of this study was to estimate technical treatment accuracy in fractionated stereotactic radiosurgery (fSRS) using the Extend™ system (ES) of Gamma Knife (GK).
Methods and materials:
The fSRS with GK relies on a re-locatable ES where the reference treatment position is estimated using repositioning check tool (RCT). A patient surveillance unit (PSU) monitors the head and neck movement of the patient during treatment and imaging. The quality assurance test of RCT was performed to evaluate a standard error (SE) associated with a measurement tool called digital probe. A ‘4-mm collimator shot’ dose plan for a head–neck phantom was investigated using EBT3 films. CT and MR distortion measurement studies were combined to evaluate SEimaging. The combined uncertainty from all measurements was evaluated using statistical methods, and the resultant treatment accuracy was investigated for the ES.
Results:
Four sets of RCT measurements and 20 observations of associated digital probe showed SERCT of ±0·0186 mm and SEdigital probe of ±0·0002 mm. The mean positional shift of 0·2752 mm (σ = 0·0696 mm) was observed for 20 treatment settings of the phantom. The differences between radiological and predefined isocentres were 0·4650 and 0·4270 mm for two independent experiments. SEimaging and SEdiode tool were evaluated as ±0·1055 and ±0·0096 mm, respectively. An expanded uncertainty of ±0·2371 mm (at 95% confidence level) was observed with our system.
Conclusions:
The combined result of the positional shift and expanded uncertainty showed close agreement with film investigations.