We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
The purpose of this study was to determine if Clostridioides difficile (C. diff) was present on the electrocardiogram (ECG) right arm leads, blood pressure cuffs, and fingertip pulse oximetry sensors of monitor/defibrillators used in the prehospital setting.
Methods:
On March 22, 2019, a total of 20 prehospital monitor/defibrillators located at an Emergency Medical Service (EMS) station in Alabama (USA) were assessed for C. diff. The inside area of the fingertip pulse oximetry sensor, patient contact side of the blood pressure cuff, and right arm ECG lead of monitor/defibrillators (n = 60) were swabbed using a sterile cotton-tipped applicator saturated in a 0.85% Sodium Chloride solution. These cotton-tipped applicators were then inserted, scored, and released into Banana Broth vials. The vials were then sealed tightly and immediately transported to the laboratory, where they were incubated at 36°C for 72 hours. Colorimetric change from red to yellow was considered a positive indication for the presence of C. diff.
Results:
Of 20 blood pressure cuffs, 15 had C. diff contamination (75%); C. diff was also present on 19 of 20 fingertip pulse oximeter sensors (95%) and 20 of 20 ECG right arm monitor leads (100%).
Conclusion:
Prehospital monitor/defibrillators may represent a significant reservoir of C. diff and other pathogenic bacteria. Improved disinfection protocols for reusable monitoring equipment and transition to disposable monitoring equipment used in the prehospital setting may reduce the risk of patient and EMS provider infection.