We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
This is a phantom study to evaluate the dosimetry effects of using virtual bolus (VB) in TomoTherapy Treatment Planning System (TPS) optimisation for superficial planning target volume (PTV) that extends to the body surface. Without VB, the inverse-planning TPS will continuously boost the photon fluence at the surface of the superficial PTV due to lack of build-up region. VB is used during TPS optimisation only and will not be present in actual treatment delivery.
Materials and methods:
In this study, a dummy planning target was contoured on a cylindrical phantom which extends to the phantom surface, and VB of various combinations of thickness and density was used in treatment planning optimisation with TomoTherapy TPS. The plans were then delivered with the treatment modality TomoTherapy. Radiochromic films (Gafchromic EBT3) were calibrated and used for dose profiles measurements. TomoTherapy Planned-Adaptive software was used to analyse the delivered Dose-Volume Histograms (DVHs).
Results:
The use of 2 mm VB was not providing adequate build-up area and was unable to reduce the hot spots during treatment planning and actual delivery. The use of 4 mm VB was able to negate the photon fluence boosting effect by the TPS, and the actual delivery showed relatively small deviations from the treatment plan. The use of 6 mm VB caused significant dose overestimation by the TPS in the superficial regions resulting in insufficient dose coverage delivered.
Findings:
VB with the combination of 4 mm thickness and 1·0 g/cc density provides the most robust solution for the TomoTherapy TPS optimisation of superficial PTV.