We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure [email protected]
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
To deliver radiation doses with higher accuracy, radiation treatment through megavoltage photon beams from linear accelerators, is accepted widely for treating malignancies. Before calibrating the linear accelerators, it is essential to make a complete analysis of all photon beam profile parameters. The main objective of this exploration was to investigate the 6 and 15 MV photon beam profile characteristics to improve the accuracy of radiation treatment plans.
Methods
In this exploration, treatment parameters like depth, field size and beam energy were varied to observe their effect on dosimetric characteristics of beam profiles in a water phantom, generated by linear accelerator Varian Clinac.
Results
The results revealed that Dmax and Dmin decreased with increasing depth but increased with increasing field sizes. Both left and right penumbras increased with increasing depth, field size and energy. Homogeneity increased with field size but decreased with depth. Symmetry had no dependence on depth, energy and field size.
Conclusion
All the characteristics of photon beam dosimetry were analysed and the characteristics like homogeneity and symmetry measured by an ion chamber in a water phantom came within clinically acceptable level of 3 and 103%, respectively, thus fulfilled the requirements of standard linear accelerator specifications. This exploration can be extended to the determination of beam profile characteristics of electron and photon beams of other energies at various depths and field sizes for designing optimum treatment plans.