Introduction
Atomic force microscopy (AFM) has been recognized since the mid-eighties [Reference Binnig, Quate and Gerber1] as an excellent technique to image a wide range of samples in their near-natural environment. Although the primary function of AFM is to generate three-dimensional (3D) profiles of the scanned surface, much more information can be delivered via this technique. In 1993, TappingMode was developed [Reference Zhong, Innis, Kjoller and Elings2], which prevents tip and sample damage due to friction and shear forces and allows qualitative mechanical property mapping through phase imaging [Reference Chernoff and Bailey3, Reference Tamayo and García4]. About the same time, force spectroscopy [Reference Lee, Kidwell and Colton5] and force volume (FV) [Reference Radmacher, Cleveland, Fritz, Hansma and Hansma6] were developed to study tip-sample forces at a point or over an area, respectively. To date, force spectroscopy and FV are the most commonly used AFM modes for measuring nanometer-scale mechanical forces in a quantitative manner. Unfortunately, force spectroscopy and FV suffer from slow acquisition speed and a lack of automated tools; these operating characteristics limit their use because of the hundreds or thousands of curves that are required for good statistics.
In 2010, the PeakForce QNM mode was developed [Reference Pittenger, Erina and Su7], based on Bruker's PeakForce Tapping technology. In this technique, the probe is oscillated around 1 kHz and the peak force (maximum nominal force applied to the sample) is used for feedback control. Each time the tip interacts with the sample, a force curve is collected and analyzed by PeakForce QNM. The resulting signals are extracted and quantitatively displayed as a false-colored, real-time image. Currently available data types include the peak force, the adhesion, the Young's modulus, the deformation, and the dissipation. This mode has been successfully tested on a wide range of samples [Reference Berquand8], including bio-polymers [Reference Adamcik, Berquand and Mezzenga9, Reference Young, Monclus, Burnett, Broughton, Ogin and Smith10] and live eukaryotic cells [Reference Pletikapic, Berquand, Misic and Svetlicic11], as well as human models [Reference Heu, Berquand, Caille-Elie and Nicold12]. This article describes the application of this method to measurements on live cells.
Comparing FV and PeakForce QNM
Because measuring and mapping mechanical properties of live samples are of high importance in biology, the capabilities of force spectroscopy, FV, and PeakForce QNM AFM are all useful for studying cell mechanics. Table 1 summarizes the characteristics of each technique and compares their principal benefits and drawbacks. PeakForce QNM is best for high-resolution imaging or relatively high-speed imaging where material properties are of interest. FV is useful as a comparison to PeakForce QNM and for cases where the loading rate dependence of the measurement is important, such as studies of viscoelastic behavior and kinetic binding or unfolding measurements. Single-point force spectra are best for situations where mapping is not as important, where there are only a few points of interest in a sample, or where force measurements need to be separated by some distance to avoid the interaction of one measurement with the next. The key differences between PeakForce QNM and the more traditionally used FV (and force spectroscopy) are:
1. PeakForce QNM uses a sinusoidal modulation of the base of the cantilever relative to the sample surface, whereas FV uses linear ramping. This allows PeakForce QNM to acquire thousands of ramps per second, whereas FV ramping is usually limited to a maximum of around ten ramps per second. The higher ramp rate of PeakForce QNM permits the acquisition of more detailed material property maps in much less time than FV.
2. PeakForce QNM controls the normal force of the tip-sample interaction by detecting the peak force of each tap and feeding the information into a feedback loop that is running continuously. This force control benefits from the results of previous taps and from the fact that the sinusoidal waveform causes the tip velocity to approach zero as the tip approaches the peak, allowing for an ultra-low interaction force (the peak force can be as low as ten pN). This is in contrast to FV where each tap is treated separately and the tip approaches the surface at full speed. Once the trigger force is detected, the system attempts to retract instantaneously, leading to overshoot and larger forces, especially at higher ramp rates.
3. PeakForce QNM is usually more stable than FV because there is less time for the system to drift within an image, and because the superior force control inhibits tip wear and contamination.
These properties make PeakForce QNM significantly better for material property mapping than FV in most cases. On the other hand, FV and force spectroscopy allow more control of the details of the ramp, such as independent control of approach and retract tip velocities and surface hold. Also, FV and force spectroscopy taps can be separated by some distance because the triggering is treated independently, whereas PeakForce QNM taps must be close together in order to achieve the best feedback performance.
Results
Preliminary testing on bacteria.Figure 1a shows a pair of E. coli bacteria rendered in 3D, based on the topography of the sample, with the brightness of the image based on the value of Young's modulus. This image was collected using PeakForce QNM in about 10 minutes. This is in contrast to the FV image in Figure 1b, which took about 35 minutes to collect. The improvement in speed is convenient and allows more and higher-resolution images (in this case, 256 × 256 versus 64 × 64) to be collected during the life of a cell. The bacteria are still alive during imaging (in liquid at room temperature), and the one on the right appears to be in the process of dividing. Note that the dividing cell on the right is significantly softer (darker) than the cell on the left (~2 MPa versus ~15 MPa). Note the presence of some softer components on the substrate, including the bacterial flagella in the lower-right corner. For these images, the modulus was calculated using the Sneddon (conical) model [Reference Chernoff and Bailey3], assuming a tip with an 18-degree half-angle (typical for the DNP-A type probe used).
It is useful to examine individual force curves from interesting parts of the image. PeakForce Capture allows the simultaneous capture of the force curves that were used to create the property maps in PeakForce QNM. As before, QNM images are still acquired normally, but a force curve for each pixel in the property map is saved separately in a Quantitative Force-Volume file format. These files can be opened in NanoScope analysis for additional exploration with the full Quantitative Force-Volume analysis tools. Additionally, calibration parameters can be adjusted and different property models can be compared (for example, Sneddon [Reference Sneddon13] versus the Derjaguin-Muller-Toporov (DMT) [Reference Derjaguin, Muller and Toporov14]). Because the same tools are used to analyze both FV and PeakForce Capture files, it is simple to compare results collected at ramp rates from 0.1–10 Hz (FV) to those at ramp rates from 250 Hz and higher (PeakForce Capture). This allows investigation of the time dependence of the tip-sample interaction.
Figure 1c shows a pair of force curves (approach only) from PeakForce Capture. One was collected on top of the cell, whereas the other was collected on the sample substrate as indicated in Figure 1a. The slope of the curve from the substrate (red) is initially smaller than that of the bacteria, but it increases rapidly, becoming nearly vertical at the most extended point. This occurs because there is a soft, thin gelatin coating on the substrate that is intended to aid immobilization [Reference Doktycz15] of the bacteria cells. When the tip penetrates deeper into the gelatin, it begins to feel the underlying glass substrate, which is much stiffer than the cells. The curve collected on the cell (green) is more uniform in slope, indicating that the substrate is not influencing the measurement significantly.
Probing agarose gels at various ramp rates. Agarose is a polysaccharide derived from agar. It is most widely used as a medium for gel electrophoresis measurements, but it is appearing in recent applications as a tissue mimicking material (for example, tissue phantoms) [Reference Luo, Yang, Ying, Awad, Choti and Taylor16, Reference Lobsien, Dreyer, Stroh, Boltze and Hoffmann17, Reference Deepthi, Bhargavi, Jagadeesh and Vijaya18, Reference Pichardo, Kivinen, Melodelima and Curiel19]. The mechanical properties of agarose gels are concentration-dependent, with higher agarose concentrations resulting in “stiffer” gels. This allows preparation of agarose gels having different elastic modulus values in the biologically relevant range of tens to thousands of kilopascal (kPa). To demonstrate the ability of FV and PeakForce QNM to measure these very soft samples, gels containing 1%–5% by wt. of agarose were prepared and imaged in phosphate buffered saline (PBS) buffer. Figure 2a shows typical approach curves collected on the 3% agarose sample with ramp rates from 1 Hz (FV) to 250 Hz (PeakForce QNM) are overlaid on the same plot to demonstrate the similarity of results from the two techniques even over several orders of magnitude of ramp rate. If Sneddon modulus is calculated for each of these curves, the results were 353 kPa at 1 Hz, 340 kPa at 5 Hz, 357 kPa at 10 Hz, and 351 kPa at 250 Hz. These results are essentially identical within the uncertainty of the measurement.
Figures 2b, 2c, and 2d show the same type of experiment in a more statistically relevant way. The FV results (1 Hz) are shown in green with the PeakForce QNM results at 250 Hz in red. The PeakForce QNM histogram provides much better statistics because it is based on 16,384 (128 × 128) measurements compared to 256 (16 × 16) measurements for FV. Despite the large difference in the number of samples for each measurement, the acquisition times are similar: ~11 minutes for PeakForce QNM versus 4 minutes for FV. Thus, PeakForce QNM provides significantly better statistics than FV at minimal cost. The Young's modulus can be calculated by directly analyzing the color contrast of the captured images (bearing analysis) or by exporting and processing the selected force curves. When thousands of curves are involved, automated analysis becomes critical.
If the sample has a significant time-dependent deformation mechanism such as viscoelasticity, the results of FV and PeakForce QNM may be different, because the time dependence allows different amounts of deformation within the time scale of the ramp. Figures 2b, 2c, and 2d show that there is very little difference between measurements at 250 Hz (PeakForce QNM) and 1 Hz (FV) for the 1% and 3% gels, but the 5% gel has some increase in modulus at higher frequencies. Note that the Sneddon and DMT models only include elastic deformation mechanisms (without time dependence). To quantify this time-dependent behavior, new analysis techniques are needed.
FV versus PeakForce QNM on live cells. Many phenomena relating to the lifecycle and behavior of cells (for example, cancer cells) are related to cell mechanics or the mechanics of the constituent parts of the cell. The faster, higher-resolution mapping made possible with PeakForce QNM allows more detailed mapping of these processes as they occur. The limited duration of these processes and cell lifetimes make speed of acquisition a critical aspect of many experiments. Figure 3 demonstrates the resolution possible on the lamellipodium of a mouse B16 cell imaged in HEPES buffer. This image was collected in 8.5 minutes; a similar FV map collected at a ramp rate of 2 Hz with the same resolution would take about 9 hours.
Figure 3a shows a 3D topographic rendering of the lamellipodium showing some actin fibrils. Figures 3b and 3c compare modulus maps calculated from the DMT (sphere) model using the retract curves and the Sneddon (cone) model using the approach curves. The substrate image (upper right) is saturated to white because the cantilever spring constant for the probe used is too small to provide a quantitative modulus measurement on the much stiffer surface of the glass bottom Petri dish. The DMT modulus map is brighter than in the Sneddon map, which we expect to be more accurate because the deformation depth on the cell was between 100 and 200 nm. Because the tip radius is ~30 nm, the tip shape can be approximated by that of a cone. To demonstrate the difference, a single curve was extracted from the PFC data and analyzed with both the DMT model (b) and the Sneddon model (c). Comparing the green fit curve to that of the data, it is easy to see that the Sneddon model fits the data much better than the DMT model.
During image acquisition, force curves (similar to those in Figures 3b and 3c) were collected and analyzed along with the topography, providing the simultaneous maps of DMT modulus, Sneddon modulus, peak force error, deformation, dissipation, and adhesion. Figure 4 shows six examples of these maps. Details such as the actin fibrils in the cell cytoskeleton are visible in many of the channels, but there are differences. For example, the Peak Force Error channel and Deformation channel show many small fibrils, whereas fewer fibrils are visible in the modulus channels and none are visible in the adhesion map.
Although PeakForce QNM can correct for some background deflection not related to the short-range tip-sample interaction, it is not always possible to correct all of the background in fluid due to variation of apparent liquid viscosity near the sample surface. For highest measurement accuracy, it is best to minimize this viscous background by using the smallest modulation amplitude possible and <1 kHz modulation frequency. These images were collected with modulation amplitude of 200 nm at 250 kHz. The probe was also selected with the viscous background in mind.
FV is well accepted as a tool for studying the mechanical properties of cells at ramp rates of less than 10–20 Hz. Figure 5 compares the Sneddon modulus maps from Figure 3 with those from FV at 1 Hz and 5 Hz and with PeakForce Capture at 250 Hz. All of the maps show that the cell's lamellipodium has about the same modulus (compare the color of the lower-left part of each image in Figures 5a and 5b) independent of ramp rate and technique. Figure 6 shows a histogram of all of the images, showing that all of the peaks are around 20 kPa (0.02 MPa). The lack of variation indicates that there is little viscoelasticity or any other time-dependent deformation mechanism active in the range of frequencies between 1 Hz and 250 Hz for the lamellipodium of this cell. The FV images were collected at low resolution (16 × 16 pixels) to save time, but the resulting modulus maps do not have enough resolution to clearly identify the actin fibrils in the cytoskeleton that are visible in the PFC and PF-QNM images.
Conclusion
The mechanical properties of biological samples often affect their structure and functional activity and are hence important to biologists. FV has been accepted since the mid 1990s and is still a powerful tool for measuring and mapping the mechanical properties of biological samples. FV is optimized for mapping with low ramp rates (~0.5–10 Hz) and relatively low resolution. PeakForce QNM improves upon FV in terms of resolution and speed (with ramp rates ~250 Hz–2 kHz), making it more practical to collect and analyze significantly more data for better detail and statistics. Together, FV and PeakForce QNM provide opportunities for comparisons of material response over about four orders of magnitude of ramp rate in air or in a liquid environment.
Special thanks to Manfred Radmacher (University of Bremen, Germany) for assistance with the mouse cell measurements.