Hostname: page-component-586b7cd67f-r5fsc Total loading time: 0 Render date: 2024-11-22T22:14:09.482Z Has data issue: false hasContentIssue false

Advanced Chemical Analysis Using an Annular Four-Channel Silicon Drift Detector

Published online by Cambridge University Press:  06 March 2017

Ralf Terborg*
Affiliation:
Bruker Nano GmbH, Am Studio 2D, 10317 Berlin, Germany
Andi Kaeppel
Affiliation:
Bruker Nano GmbH, Am Studio 2D, 10317 Berlin, Germany
Baojun Yu
Affiliation:
Bruker Nano Analytics Division, 19F, Xinyuan Technology Building, No.418 Guiping Rd, Caohejing Development Zone, Shanghai 200233, China
Max Patzschke
Affiliation:
Bruker Nano GmbH, Am Studio 2D, 10317 Berlin, Germany
Tobias Salge
Affiliation:
Natural History Museum, Cromwell Road, London SW7 5BD, UK
Meiken Falke
Affiliation:
Bruker Nano GmbH, Am Studio 2D, 10317 Berlin, Germany

Abstract

Type
X-ray Microanalysis
Copyright
Copyright © Microscopy Society of America 2017 

Introduction

Peltier-cooled silicon drift detectors (SDDs) have become standard instrumentation for energy-dispersive X-ray spectrometry (EDS) in electron microscopes [Reference Strìder1, Reference Goldstein2]. State-of-the-art SDDs reach an energy resolution of 121 eV at Mn-Kα, show good low-energy performance below 100 eV, and produce Gaussian peak shapes undistorted by detector artifacts such as shelf and tail effects. This is beneficial for peak separation over the entire available energy range. But peak separation is particularly important for overlaps of low-energy lines, such as the K-lines of light elements (Be - F) and the L-lines (K - Ni), M-lines (Mo – La), and N-lines (Hf - U) of heavier elements.

The use of multiple detectors allows a substantially higher acquisition speed and higher throughput because of greater geometrical collection efficiency from a larger solid angle for X-ray collection. Such multiple detector arrangements need multiple accessible EDS ports, which may not be possible because of geometric limitations inside and outside the microscope. Thus, further increasing the solid angle and throughput by using more than two individual detectors becomes impractical. Fortunately, thanks to advances in SDD technology, other alternatives are available.

Multi-channel SDDs with four separate detector segments integrated into one single SDD chip have been developed to deliver improved EDS performance. The combined parallel readout of the detector segments provides the increased count rate that would be expected in the case of several individual detectors. The detection geometry can be optimized using an annular arrangement of the SDD segments, as implemented in the annular Bruker XFlash® FlatQUAD system.

The present article shows this detector optimized for different tasks. We show results of this dealing with the challenges of meteorite exploration, polymer research, and low-yield life science samples.

Materials and Methods

The annular detector and its setup

The four integrated SDD segments of the XFlash® FlatQUAD detector are arranged in radial symmetry around a hole through which the electron beam passes. The detector is inserted horizontally and placed like a backscattered electron (BSE) detector between sample and pole piece. Thus the annular active SDD area resides directly above and very close to the sample, achieving both a high take-off angle between 40° and 70°, which minimizes shadowing effects, and a relatively large solid angle of collection. Each of the four SDD segments has an active area of 15 mm², resulting in a total of 60 mm². The respective annular geometry (Figure 1) delivers a solid angle of more than 1 steradian (sr) [Reference Terborg and Rohde3Reference Demers6]. To avoid alteration of X-ray spectroscopy results by electrons coming from the sample and to maintain low-energy X-ray sensitivity, three polymer windows of different thickness can be placed semiautomatically in front of the chip depending on the acceleration voltage used. Because this annular SDD array is closer to the specimen, the solid angle value is 100 times higher than that typically available in a 10 mm² detector with a 35° takeoff angle and over 10 times higher than that available from two 30 mm2 detectors or a single 100 mm2 detector in conventional setups. High count rates can be obtained even with low probe currents. This is achieved by processing the counts from the annular detector through four separate electronic channels in parallel, leading to a maximum output count rate of up to 2,400 kilo-counts per second (kcps) [Reference Terborg and Rohde3, Reference Soltau7].

Figure 1 Left: Solid angle Ω and X-ray output count rate (OCR) versus detector-to-sample distance. Note that the optimum detector-sample distance d of 2.8 mm delivers a solid angle of >1.1 sr. Right: Detector (blue) arrangement between SEM pole piece and sample.

These advantages invite challenging analysis approaches such as the use of low electron accelerating voltages, ultra-low probe currents, and the acquisition of high-resolution element mappings of large sample areas or of single nanoparticles with sizes down to a few nm. Non-destructive sample analysis with high spatial resolution can be very difficult to accomplish using conventional EDS detectors because sample surface charging, electron beam damage, and shadowing effects are common problems in scanning electron microscopy (SEM) that complicate X-ray acquisition.

Low accelerating voltages and low probe currents can be helpful for the analysis of electron-beam-sensitive materials and of nonconductive samples that charge during examination. Compared to low-vacuum analysis, often used to avoid sample charging, the use of low kV and low probe current in high vacuum circumvents the beam skirting effect. Beam skirting means that the spatial resolution of analysis is degraded by beam broadening because of collisions of probe electrons with gas molecules, atoms, and ions. Therefore low probe current under high vacuum conditions is desired. Thin TEM samples or life science samples producing a low X-ray yield are additional situations well suited for analysis with this efficient annular detector. Its geometry was optimized in a way that allows the combination of thin-sample EDS with diffraction experiments in transmission, for example transmission Kikuchi diffraction (TKD) in the SEM [Reference Keller and Geiss8]. Small particles and structures in the nanometer range also produce low X-ray yield, but they can be investigated at low accelerating voltages in SEM because of the small interaction volume attained. High-spatial-resolution EDS mapping on bulk specimens is best accomplished at low kV with the annular detector. Furthermore, fast mapping of sensitive samples and of large sample areas, including stitching maps together for complete analysis and statistics, are applications that benefit greatly from efficient X-ray collection. Last but not least, element mapping of porous or otherwise topographically structured samples present another application for the annular Bruker XFlash® FlatQUAD.

Results

High-vacuum EDS at low beam current of the historic Mocs Meteorite

This large meteorite fell as a bright ball of fire in February 1882 near the village of Mocs, Romania. A sample of this stony meteorite, officially named NHMW-H9898 (Figure 2a), entered the Natural History Museum Vienna (NHMV) collection in 1908 as a donation. A large surface of 60 mm × 47 mm revealing a network of shock veins up to 10 mm, cut and polished but uncoated, was investigated. For the current SEM-EDS investigation, a region of interest was identified after acquiring an 800 pixel by 600 pixel overview map at a 2 µm by 2 µm pixel size over 47 minutes (Figure 2b). The Pb M-line and S K-line were deconvoluted during measurement using a physical background subtraction and a least square fit to library spectra (Figure 2c). To avoid charging of the uncoated nonconductive meteorite surface, a 6 kV beam with a probe current of less than 10 pA was employed under high-vacuum conditions to produce a second map with an input count rate of 3000 cps (Figure 2d). The 6 kV accelerating voltage kept the interaction volumes small for both electrons and X-rays. A smaller region of interest was chosen for an overnight map with a 130 nm pixel size, but otherwise had the same experimental conditions (Figure 2e).

Figure 2 Historic Mocs meteorite. (a) Photograph of the sample (65 mm × 50 mm). (b) Overview X-ray map at 6 kV, 10 pA, input count rate 3 kcps, and 800 × 600 pixels was acquired for 2800 s. X-ray map was overlaid with the secondary electron (SE) image revealing lead contamination from the historical polishing process. (c) Pb M-line (Mα 2.343 keV) and S K-line (Kα 2.307 keV) were deconvoluted during measurement using a physical background subtraction and a least square fit with stored line profiles. (d) A higher magnification composite map showing only the elements C, O, Pb, and S of the rectangular region of interest in (b) overlaid with the SE micrograph. This shows that lead and soot were deposited on the sample surface that may consist of sulfides (orange areas = red + yellow) and oxides (blue shows the O content). (e) Image detail of the outlined area in (d), without S, documenting that carbon features < 300 nm in size (green filaments) can be made visible under high-vacuum conditions using low kV and ultra-low probe current. Sample courtesy of Natural History Museum, Vienna, Dept. of Mineralogy and Petrography. Sample NHMW-H9898 was a donation from industrial magnate Sigmund Sachsel. Special thanks to Ludovic Ferriere for taking the sample to the Bruker Nano laboratories for analysis and the resulting discussion.

The overview map of the Mocs meteorite (Figure 2b) shows lead enrichment in cracks that were most likely caused by an old polishing method employing a lead lapping plate that was used in the past. Sufficient data quality (large number of counts) allows deconvolution of the Pb M and S K lines (Figure 2c), providing a separation of lead contamination from sulfide minerals that are part of the meteorite material beneath the contamination. The EDS analysis at higher spatial resolution (Figures 2d and 2e) shows the deposition of lead on oxides and sulfides. Spherical carbon particles and filaments with sizes <300 nm were observed, a sign of surface contamination with soot, very likely related to furnaces used at the NHMV during the last century or from the time before it was added to the NHMV collection.

Polymer with embedded organo-clay particles

Soft and porous materials are challenging not only in terms of beam sensitivity and charging but also in regard to judging the distribution of pores or property-enhancing particles. The following experiment describes the EDS analysis of a polymer compound containing Si-rich organo-clay particles, derived from organically modified montmorillonite. This new type of material was studied for its potential of improved mechanical, thermal, and flame-resistance properties [Reference Esfandiari9, Reference da Silva Dalto10]. For comparison of detector performance, element maps were acquired using the XFlash® FlatQUAD detector and a standard 30 mm² XFlash® SDD under the same acquisition conditions at 3 kV and 220 pA beam current (Figure 3). The annular detector setup was optimized here not for highest solid angle but for highest possible take-off angle. So the working distance was increased above the optimum distance for large solid angle in order to investigate deeper pores. This still led to more than 12.5 times higher input count rate: 10 kcps compared to the 0.8 kcps for the conventional 30 mm2 SDD. Note that the result from the conventional SDD suffers from extreme shadowing effects because of the rough sample surface. Shadowing is nearly eliminated in the map acquired with the annular XFlash® FlatQUAD, allowing the microscopist to properly evaluate the distribution and embedding of the silicon-containing nano-clay particles in the polymer matrix.

Figure 3 Combined C (blue) and Si (red) element maps overlaid with the corresponding SE images of a low-density polyethylene composite with silicon-containing organo-clay particles. The clay particles have formed large agglomerates. (a) Map acquired with a conventional 30 mm² XFlash SDD setup, at 3 kV, 220 pA, input count rate 0.8 kcps, 1024 × 768 pixels, acquired over 320 s. Note the shadowing due to surface roughness. (b) Map acquired with the XFlash® FlatQUAD detector using the same acquisition conditions as in (a) but resulting in a 12.5 times higher count rate and no shadowing. (c) Magnified map area (green square) of map (b) in annular detection geometry. (d) Magnified map area (green square) of image (c) showing no shadowing effects. Sample courtesy of D. P. da Silva Dalto Center for Mineral Technology (CETEM), Brazil, M. J. O. C. Guimarães and M. E. F. Garcia, Federal University of Rio de Janeiro, Brazil [Reference da Silva Dalto10].

Impact micro-crater in aluminum foil

Another fascinating analysis problem is the examination of artificial impact craters in aluminum foils produced in the laboratory under the collection conditions for NASA’s Stardust Mission. The primary task of this mission was to collect dust samples of a comet and return them to Earth for analysis. In order to find suitable instrumentation to analyze the precious samples from space, an analogous test experiment was created on earth [Reference Kearsley11]. Polymineralic aggregate projectiles were produced by aerosol impregnation of a dried slurry of olivine, diopside, and pyrrhotite powders. These projectiles were fired at 6.05 km s-1 from a light gas gun onto aluminum foils, forming impact micro-craters. The generated micro-craters were analyzed using a conventional SDD at 20 kV and an annular XFlash® FlatQUAD at 6 kV (Figure 4). Compositional analysis of the deeper part of the crater was not accessible using the conventional detector setup because of shadowing effects. In comparison, the analysis using the annular detector shows residues of the projectile that can be clearly distinguished from the aluminum target foil across the whole field of view without shadowing effects. Residues from the three mineral components were detected mostly in the crater interior. Silicate and sulfide melt with flow texture originating from olivine (Mg-green), pyrrhotite (S-blue), and diopside (Ca-red) were found. Mg-rich fragments, probably from olivine, occur as well as a few particles rich in Ca or S.

Figure 4 Impact experiment producing a crater in aluminum. (a) Element map of the polymineralic projectile aggregate showing Mg (green), S (blue), and Ca (red) from olivine, diopside, and pyrrhotite, respectively. (b, c) Height map and profile of the micro-crater in aluminum foil, generated during the test experiment. (d) Crater mapped with a conventional SDD setup at 20 kV, 3nA, 1024 × 768 pixels acquired over 15.9 hours. The deeper part of the crater is not accessible. (e) Crater map acquired using the annular XFlash® FlatQUAD detector at 6 kV, input count rate 175 kcps, 800 × 600 pixels collected in 3.9 hours. Residues from the three minerals olivine (Mg in green), diopside (Ca in red), and pyrrotite (S in blue) appear as the three primary colors, with mixtures of diopside with pyrrhotite appearing purple and of mixtures of olivine with diopside appearing orange to yellowish green. Sample and data courtesy of A. T. Kearsley1, T. Salge2,1, P. J. Wozniakiewicz1,3, M. C. Price3, R. Terborg2, M. J. Burchell3, and M. J. Cole1: 1 Core Research Laboratories, Natural History Museum, London, SW7 5BD, UK; 2 Bruker Nano GmbH, Berlin, Germany, 3 School of Physical Science, University of Kent, Canterbury, CT2 7NH, UK [Reference Kearsley11].

Titanium dioxide nanotubes decorated with gold nanorods

Titanium dioxide nanotubes attract attention in biomedicine because of their biocompatibility and bioactivity. It has been shown that they can promote the adhesion and proliferation of rat-bone-marrow-derived mesenchymal stem cells and hydroxyapatite nucleation and growth. Composite materials consisting of TiO2 nanotubes decorated with metal nanoparticles, for example, Au nanoparticles or Au nanorods, have an antibacterial effect that can be sustained in darkness. This effect is based upon the formation of a Schottky barrier between TiO2 and the conductive Au, which can set up a lethal reaction at the cell walls of the bacteria. Thus a porous TiO2 structure decorated with Au nanoparticles or nanorods is a promising composite material for orthopedic and dental implants [Reference Yang12]. Figure 5 shows the Au map of a TiO2 nanotube material containing Au nanorods of 6 nm to 12 nm in diameter. This Au X-ray map was acquired in 4 minutes at 5 kV accelerating voltage and superimposed on the SEM image. The XFlash® FlatQUAD with its high take-off angle and high X-ray collection angle allows measurement of the distribution, attachment, and size of the Au nanorods on the highly topographic TiO2 nanotube surface in one experiment. Furthermore, the Au nanorod distribution from the EDS element map could be correlated with data from fluorescence microscopy. This correlation shows which particular sites on this material attract bone mesenchymal stem cells and promote their spreading and which have antibacterial effects. This sample of Au nanorods on a porous TiO2-nanotube structure represents a type of TEM-suitable sample investigated here in the SEM; it shows the promise of highly sensitive EDS systems applied to STEM in SEM, so-called T-SEM EDS [Reference Hodoroaba13].

Figure 5 X-ray element map of the O and Au distribution in a Au-decorated TiO2 nanotube matrix, acquired using the XFlash® FlatQUAD, superimposed on an SEM image: 5 kV, input count rate 38 kcps, 1024 × 768 pixels, acquisition time 240 s. Sample and data courtesy of Guangxi University, which also provided measurement time at the SEM, and special thanks to Prof. Xuanyong Liu and Prof. Yi Zeng from Shanghai Institute of Ceramics, Chinese Academy of Sciences, for generously providing the sample [Reference Yang12].

Discussion

The applications presented show that the annular XFlash® FlatQUAD detector can overcome some of the restrictions of conventional detectors. For the uncoated historic Mocs Meterorite, low kV and low beam current were necessary to avoid charging. This analysis is also possible with conventional detectors, but the large solid angle of the annular detector reduces the acquisition time significantly.

The same applies to samples where only small amounts of material are available for analysis, such as small particles or thin layers on bulk and electron transparent samples. Furthermore, light element samples produce low X-ray yields, so the annular design with four detectors is beneficial because it increases X-ray collection efficiency. Also, a low accelerating voltage may be used to produce a small interaction volume for high spatial resolution; again a high collection efficiency is important.

Energy-dispersive X-ray analysis in the cracks of the meteorite, in the porous polymer, and at the bottom of the impact microcrater is only possible because of the high take-off angle and signal acquisition from four different directions. There is a tradeoff between high take-off angle and solid angle. For a higher take-off angle the distance between sample and detector has to be increased. So using a take-off angle of more than 65° reduces the solid angle from 1.1 sr to less than 0.5 sr (Figure 1).

The signal acquisition, amplification, and processing of the four channels is completely separated. This increases the count-rate capability by a factor of four compared to the detector area of one segment. Three-dimensional information can be obtained using the different detector segments separately.

Conclusion

The annular XFlash® FlatQUAD detector is a device that combines a large solid angle (>1.1 sr) with output count rates up to 2.4 million counts per second. These properties can be used for efficient element mapping-applications such as ultra-fast mapping or large-area mapping. The collection efficiency allows elemental analysis of samples at both low accelerating voltages to achieve a small interaction volume and at low beam currents. These operating conditions provide high spatial resolution and high detection sensitivity without the necessity of applying a conductive coating or working at low vacuum. The high take-off angle annular detector arrangement enables element distribution mapping of highly topographic samples. Furthermore, use of this annular detector allows the analysis of small amounts of material. New analytical approaches in various fields become possible, especially in biological science and cultural heritage studies.

References

[1] Strìder, L et al., Microsc Microanal 4 (1998) 622631.CrossRefGoogle Scholar
[2] Goldstein, JI et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, New York, 2003.Google Scholar
[3] Terborg, R and Rohde, M, Microsc Microanal 17 (Suppl.2) (2011) 892893.Google Scholar
[5] Zaluzec, NJ, Microsc Microanal 20 (2014) 13181326.Google Scholar
[6] Demers, H et al., Microsc Microanal 19 (Suppl.2) (2013) 364365.CrossRefGoogle Scholar
[7] Soltau, H et al., Microsc Microanal 15 (Suppl.2) (2009) 204205.CrossRefGoogle Scholar
[8] Keller, RR and Geiss, RH, J Microscopy 245 (2012) 245251.Google Scholar
[9] Esfandiari, A et al., Journal of Applied Sciences 8 (2008) 545561.Google Scholar
[10] da Silva Dalto, DP, Master’s thesis, Escola de Química, UFRJ, Rio de Janeiro, Brazil (2010).Google Scholar
[11] Kearsley, A et al., 44th Lunar and Planetary Science Conference (2013).Google Scholar
[12] Yang, T et al., Colloid Surface B 145 (2016) 597606.Google Scholar
[13] Hodoroaba, D-V et al., EMAS 2015; IOP Conf. Series: Materials Science and Engineering, 109 (2016) 012006–1; DOI: 10.1088/1757-899X/109/1/012006.Google Scholar
Figure 0

Figure 1 Left: Solid angle Ω and X-ray output count rate (OCR) versus detector-to-sample distance. Note that the optimum detector-sample distance d of 2.8 mm delivers a solid angle of >1.1 sr. Right: Detector (blue) arrangement between SEM pole piece and sample.

Figure 1

Figure 2 Historic Mocs meteorite. (a) Photograph of the sample (65 mm × 50 mm). (b) Overview X-ray map at 6 kV, 10 pA, input count rate 3 kcps, and 800 × 600 pixels was acquired for 2800 s. X-ray map was overlaid with the secondary electron (SE) image revealing lead contamination from the historical polishing process. (c) Pb M-line (Mα 2.343 keV) and S K-line (Kα 2.307 keV) were deconvoluted during measurement using a physical background subtraction and a least square fit with stored line profiles. (d) A higher magnification composite map showing only the elements C, O, Pb, and S of the rectangular region of interest in (b) overlaid with the SE micrograph. This shows that lead and soot were deposited on the sample surface that may consist of sulfides (orange areas = red + yellow) and oxides (blue shows the O content). (e) Image detail of the outlined area in (d), without S, documenting that carbon features < 300 nm in size (green filaments) can be made visible under high-vacuum conditions using low kV and ultra-low probe current. Sample courtesy of Natural History Museum, Vienna, Dept. of Mineralogy and Petrography. Sample NHMW-H9898 was a donation from industrial magnate Sigmund Sachsel. Special thanks to Ludovic Ferriere for taking the sample to the Bruker Nano laboratories for analysis and the resulting discussion.

Figure 2

Figure 3 Combined C (blue) and Si (red) element maps overlaid with the corresponding SE images of a low-density polyethylene composite with silicon-containing organo-clay particles. The clay particles have formed large agglomerates. (a) Map acquired with a conventional 30 mm² XFlash SDD setup, at 3 kV, 220 pA, input count rate 0.8 kcps, 1024 × 768 pixels, acquired over 320 s. Note the shadowing due to surface roughness. (b) Map acquired with the XFlash® FlatQUAD detector using the same acquisition conditions as in (a) but resulting in a 12.5 times higher count rate and no shadowing. (c) Magnified map area (green square) of map (b) in annular detection geometry. (d) Magnified map area (green square) of image (c) showing no shadowing effects. Sample courtesy of D. P. da Silva Dalto Center for Mineral Technology (CETEM), Brazil, M. J. O. C. Guimarães and M. E. F. Garcia, Federal University of Rio de Janeiro, Brazil [10].

Figure 3

Figure 4 Impact experiment producing a crater in aluminum. (a) Element map of the polymineralic projectile aggregate showing Mg (green), S (blue), and Ca (red) from olivine, diopside, and pyrrhotite, respectively. (b, c) Height map and profile of the micro-crater in aluminum foil, generated during the test experiment. (d) Crater mapped with a conventional SDD setup at 20 kV, 3nA, 1024 × 768 pixels acquired over 15.9 hours. The deeper part of the crater is not accessible. (e) Crater map acquired using the annular XFlash® FlatQUAD detector at 6 kV, input count rate 175 kcps, 800 × 600 pixels collected in 3.9 hours. Residues from the three minerals olivine (Mg in green), diopside (Ca in red), and pyrrotite (S in blue) appear as the three primary colors, with mixtures of diopside with pyrrhotite appearing purple and of mixtures of olivine with diopside appearing orange to yellowish green. Sample and data courtesy of A. T. Kearsley1, T. Salge2,1, P. J. Wozniakiewicz1,3, M. C. Price3, R. Terborg2, M. J. Burchell3, and M. J. Cole1: 1 Core Research Laboratories, Natural History Museum, London, SW7 5BD, UK; 2 Bruker Nano GmbH, Berlin, Germany, 3 School of Physical Science, University of Kent, Canterbury, CT2 7NH, UK [11].

Figure 4

Figure 5 X-ray element map of the O and Au distribution in a Au-decorated TiO2 nanotube matrix, acquired using the XFlash® FlatQUAD, superimposed on an SEM image: 5 kV, input count rate 38 kcps, 1024 × 768 pixels, acquisition time 240 s. Sample and data courtesy of Guangxi University, which also provided measurement time at the SEM, and special thanks to Prof. Xuanyong Liu and Prof. Yi Zeng from Shanghai Institute of Ceramics, Chinese Academy of Sciences, for generously providing the sample [12].