Hostname: page-component-78c5997874-8bhkd Total loading time: 0 Render date: 2024-11-19T04:32:35.963Z Has data issue: false hasContentIssue false

Secondary Electron Imaging in an Aberration-Corrected STEM

Published online by Cambridge University Press:  09 September 2016

D. R. G. Mitchell*
Affiliation:
Electron Microscopy Centre, Australian Institute for Innovative Materials, University of Wollongong, NSW 2500, Australia
G. Casillas
Affiliation:
Electron Microscopy Centre, Australian Institute for Innovative Materials, University of Wollongong, NSW 2500, Australia

Abstract

Type
Secondary Electron Microscopy
Copyright
Copyright © Microscopy Society of America 2016 

Introduction

Secondary electrons (SEs) are produced when an electron beam interacts with matter. This signal is widely used in the scanning electron microscope (SEM). The peak intensity of the SE energy distribution occurs at about 2 eV, and such electrons escape from within a few nanometers of the surface. Thus, the SE signal is highly surface-specific, and the image resolution is a function of the imaging probe [Reference Goldstein1]. Typical field emission gun (FEG) SEMs have imaging resolutions of just a few nm in SE mode. The great strength of SE imaging is in providing high-resolution surface (topographic) information.

Scanning transmission electron microscopy (STEM) uses a scanning electron probe, much like in SEM. However, the STEM signals originate from electrons transmitted through a thin specimen. Conventional STEM imaging of crystalline specimens is based on brightfield (BF) (Bragg diffraction contrast) and high-angle annular dark-field (HAADF) signals. These provide crystallographic and mass-thickness information, respectively. The resulting projection images impart little direct information about the surface morphology. In the case of regularly shaped particles (for example, spheres), the morphology can be inferred from the image contrast. However, for irregularly shaped particles of varying composition, this becomes difficult.

Despite its utility, SE imaging is not available on most conventional STEM instruments. SE detectors were an option on some JEOL STEMs of the 1980s/90s, such as the JEM-2000FX. At that time, this type of STEM-based SE imaging surpassed the capabilities of some of the best (LaB6) SEMs of the day [Reference Mitchell2].

SE detectors have been installed on a small number of dedicated STEMs and have achieved sub-nm resolution [Reference Allen3Reference Liu and Cowley5]. When combined with aberration correction, atomic-resolution SE imaging has been demonstrated [Reference Inada6, Reference Zhu7]. However, for atomic-scale imaging, HAADF produces higher contrast and intensity and has a much lower sensitivity to surface preparation than SE mode. The major advantage of STEM SE imaging is as a complementary technique to HAADF, typically at magnifications well below those necessary for atomic resolution.

The JEOL company has recently released a novel multi-channel plate analyzer (BEI) detector option for their aberration-corrected JEM-ARM200F platform, which enables efficient imaging with secondary electrons (SEs) and backscattered electrons (BSEs). STEM BF and HAADF signals may be acquired concurrently with SE or BSE signals, enabling conventional STEM images to be correlated with SEM-like information. This article describes some practical applications of STEM-based SE and BSE imaging for a range of materials.

Materials and Methods

Various specimens (on carbon films) were imaged: sputtered gold to demonstrate atomic resolution, amorphous boron nitride spheres to show depth effects, mineralized biochar and porous titanium dioxide to reveal surface details, sulfur nanoparticles on graphite to highlight contrast and resolution advantages, and tantalum oxide to show control over depth of field. Particulate specimens were prepared by either light grinding or ultrasonic agitation of the powder in ethanol, then placing a drop of the suspension onto a carbon support film. No subsequent coating of these specimens was performed.

Specimens were examined in a JEOL JEM-ARM200F operating at 200 kV. This was fitted with a JEOL BEI detector operated in SE and BSE topographic modes. This imaging mode is set in software that controls detector biasing. The only user adjustment is brightness and contrast.

Analytical conditions were used for imaging: 40 µm condenser aperture and 0.1 nm probe size (155 pA). Images were acquired using Gatan’s DigiScan hardware and DigitalMicrograph software. No subsequent image filtering or processing was carried out.

Results

Atomic resolution SE imaging

Figure 1 shows an image sequence obtained from a gold particle on a carbon film. The HAADF image (Figure 1a) shows the multiply twinned gold lattice. A fast Fourier transform (FFT) analysis (not shown) indicated an image resolution of 0.102 nm. The SE image taken concurrently with the HAADF image (Figure 1b) also shows lattice resolution (0.123 nm). The HAADF and SE images of this region were acquired continuously using 20 s acquisition times. After 80 s of continuous scanning, the SE lattice contrast had faded considerably (Figure 1d); whereas, the HAADF image acquired after 120 s of scanning still showed strong lattice contrast (Figure 1c). Minor shape changes and tilting of the crystal occurred due to irradiation effects, and a minor loss of HAADF contrast was observed. The significant degradation of the SE signal is not due to tilting, since the HAADF signal is more sensitive to orientation than the SE signal [Reference Inada6]. A build-up of a thin layer of carbon contamination during the continuous bombardment by the electron beam degraded the SE signal from the gold. The low energy SEs (about 2 eV) are readily absorbed by carbon layers of just a few nm in thickness [Reference Inada6]. In contrast, the high-energy (200 keV) HAADF signal is much less affected by such layers.

Figure 1 Image sequence of a twinned gold particle on carbon. Under continuous beam scanning, atomic-resolution HAADF and SE images were acquired simultaneously: (a) HAADF at 0 s, (b) SE at 0 s, (c) HAADF at 120 s, and (d) SE at 80 s electron beam exposure.

The atomic-resolution SE imaging shown is an order of magnitude better than that which can be routinely achieved with typical FEG SEM instruments. However, the HAADF image is generally much superior for atomic-resolution imaging. Thus, the primary benefit of SE imaging is realized at lower magnifications where morphology, rather than atomic structure, is of interest.

Depth information and BSE imaging

Figure 2 shows HAADF, SE, and BSE images of amorphous boron nitride spheres. The HAADF image (Figure 2a) is a projection image and thus contains little depth information. In contrast, both the SE and BSE images show which particle clusters lie above and which lie below the carbon film. For X-ray microanalysis, particles projecting over the holes in the film are preferred. However, usually nanoparticulates reside on the support film leaving the holes unpopulated. It is preferable to analyze particles that sit on top of the support film to avoid absorption effects, especially for light element analysis.

Figure 2 Amorphous boron nitirde spheres on lacey carbon. (a) HAADF image showing no depth information. (b) SE image and (c) BSE image. Both of the latter images highlight which particle clusters lie above, and which lie below, the carbon support film.

The BSE image (Figure 2c) has a much lower signal-to-noise ratio than that of the corresponding SE image (Figure 2b). The BSE signal yield increases with atomic number (Z). The very low Z elements of boron, carbon, and nitrogen in this specimen therefore represent a worst case for imaging with this mode.

Thickness and phase disposition information

Figure 3 shows a biochar particle (pyrolysed organic material) following soil exposure. Soil chemistry and microbial activity cause a range of mineral phases to precipitate on the biochar. Understanding the chemistry and mineralogy of this process helps in understanding how biochars improve soil fertility and water retention [Reference Lehmann8].

Figure 3 Mineralized biochar fragment. (a) HAADF image and (b) SE image of the same particle. Fe-rich mineral particles decorate the surface of the slab-shaped biochar particle.

The HAADF contrast of the biochar (Figure 3a) is complex because of the irregular shape and compositionally diverse components; the contrast being a function of both Z and thickness. This makes interpretation of the projection image difficult. Here, the SE image (Figure 3b) has a transformational effect for understanding the material. The particle shape and orientation clarify the contrast variations related to thickness. The small bright regions in the HAADF image are associated with iron-rich particulates.

Internal versus external structure

Highly porous oxides are ideal supports for catalysts. The interior of the TiO2 crystal in Figure 4a is extremely porous. However, this BF image provides little topographic information. The SE image (Figure 4b) shows that only a limited number of pores extend to the surface, and this may act as a bottleneck to gas exchange. The mottling on the crystal surface is due to small steps or terraces. Such features appear bright since SE emission can occur from the upper surface and exposed side simultaneously [Reference Goldstein1].

Figure 4 Highly porous TiO2 catalyst support imaged in (a) STEM brightfield and (b) SE image mode. Despite the high degree of internal porosity, very few pores penetrate the surface.

Contrast-limited resolution

Figure 5 shows a dispersion of crystalline sulfur on graphitic carbon. The HAADF image (Figure 5a) shows good contrast between the larger sulfur particles and the thinner regions of graphitic carbon. The smallest sulfur particle that can be resolved is around 6 nm. This limit of particle detection is controlled not by the probe diameter (0.1 nm), but by contrast limitations. Sub-6 nm sulfur particles located on thick regions of carbon produced insufficient contrast to be visible. Superposition of particles on the upper and lower surface of the carbon also degrade the contrast. The SE image (Figure 5b) reveals only particles on the upper surface, detecting crystals as small as 3 nm. The SE contrast between the particles and carbon is largely a function of topography, with the carbon support thickness having little effect on the resolution and contrast [Reference Inada6]. For particle size determination, the SE image is superior to the HAADF.

Figure 5 Sulphur nanoparticles dispersed on graphite: (a) HAADF image and (b) SE image. Inset (top left) shows details of the same region in both images. Contrast and resolution are superior in the SE image. Width of inset box = 120 nm.

Depth of field

Aberration-corrected STEM probes typically have a high convergence semi-angle. High image resolution is obtained at the expense of depth of field. The depth of field (Δ eff ) can be estimated using Equation 1 [Reference Erni9], where λ is the electron wavelength (2.51 × 10-12 m), α is the beam convergence semi-angle, Cc is the chromatic aberration coefficient (1.4 × 10-3 m), ΔE is the energy spread of the electron source (about 0.3 eV for a cold FEG), and Eo is the beam energy (200 keV).

(1) $$\Delta eff\,\approx\,\sqrt {\left( {{\lambda \over {\alpha ^{2} }}} \right){\plus}\left( {{{{\rm Cc}\Delta {\rm E}} \over {{\rm E}_{0} }}} \right)^{2} } $$

Improved depth of field can be achieved by using a smaller condenser aperture. For example, an aperture diameter of 40 µm makes α = 24.9 mrad and yields Δ eff = 6.1 nm; whereas, an aperture of 10 µm makes α = 6.2 mrad yielding Δ eff = 67 nm.

SE images of tantalum oxide obtained with condenser apertures of 40 µm and 10 µm are shown in Figure 6a and 6b, respectively. A 10 µm condenser aperture improves the depth of field by a factor of 10 compared with the 40 µm aperture. However, the major downside is that it produces a 16-fold reduction in the probe current. To partially compensate for this, a 3-fold increase in acquisition time was used. Nevertheless, although not easily visible in Figure 6, the smaller aperture will cause a lower signal-to-noise and a slightly poorer image resolution.

Figure 6 SE images of tantalum oxide powder showing the effect of condenser aperture size on depth of field and signal-to-noise ratio: (a) 40 µm and (b) 10 µm.

Discussion

The introduction of a BEI detector to the JEOL ARM200F platform greatly extends its imaging capabilities. The detector is an in-lens type, located above the upper objective lens polepiece. This geometry ensures efficient signal collection [Reference Griffin10]. In SEM instruments, on-axis, in-lens detectors have been shown to produce improved contrast and topographic information compared with conventional off-axis Everhart-Thornley detectors [Reference Kuo11].

In BSE (topographic) mode, the detector provides morphological information for BN spheres on a carbon film. However, when compared with the HAADF signal, the contrast and signal-to-noise ratio are lower. Because of the high energy (200 keV) of the primary electron beam, contrast is not entirely topographic in either BSE or SE mode. Overlapping particles and those particles located on the underside of the support film show weak contrast. This is to be expected for the high-energy BSE signal. However, the BSE signal can also generate SE signals as it exits the support film’s upper surface. In the case of SE imaging on non-continuous support films, low-energy SEs from particles on the underside of the film can also pass upward through holes. Such SE signals are brightest where particle edges are close to a hole in the support (for example, lower left in Figure 2b).

Atomic-resolution SE imaging is readily achieved. More importantly, at lower magnifications the topographic information provides insights into specimen location with respect to support films, which is of use in X-ray microanalysis. It also greatly aids interpretation of the complex contrast in HAADF images that can arise in topographically and compositionally irregular materials. For highly porous materials, the correlation of the internal structure (BF imaging) with the surface morphology (SE imaging) provides a more complete understanding of the material. For nanoparticulate materials distributed onto the surface of thick supports, the smallest particles may not be resolved in HAADF images because of insufficient contrast. Because the SE contrast is largely independent of the support thickness, smaller particles may be detected in such regions. Because only the upper surface of the specimen is imaged, feature superposition found in projection images is eliminated. The highly converged probe used in aberration-corrected STEM results in a small depth of field. Use of the smallest condenser aperture helps mitigate this problem, albeit at the expense of signal intensity and resolution.

Conclusions

The addition of a SE/BSE detector to the JEM-ARM200F platform greatly enhances its imaging capabilities. SE images at up to atomic resolution can be readily obtained. However, the real benefit is found at lower magnifications where topographic information greatly aids interpretation of conventional STEM projection images. Correlation of STEM images with SEM data from a separate instrument is no longer necessary, as the microscope described here enables simultaneous acquisition of both.

Acknowledgements

This research used equipment funded by the Australian Research Council (ARC) with a Linkage, Infrastructure, Equipment and Facilities (LIEF) grant (LE120100104) located at the UOW Electron Microscopy Centre. S. Joseph and M. S. Hossain of the University of Wollongong are thanked for providing specimens.

References

[1] Goldstein, JI et al., Scanning Electron Microscopy and X-Ray Microanalysis, 3rd ed., Springer, New York, 2003.Google Scholar
[2] Mitchell, DRG et al., Cement Concrete Res 28 (1998) 15711584.Google Scholar
[3] Allen, RM, Ultramicroscopy 10 (1982) 237245.Google Scholar
[4] Hembree, GG et al., Ultramicroscopy 31 (1989) 111115.Google Scholar
[5] Liu, J and Cowley, JM, Ultramicroscopy 37 (1991) 5071.Google Scholar
[6] Inada, H et al., Ultramicroscopy 111 (2011) 865876.Google Scholar
[7] Zhu, Y et al., Nature Materials 8 (2009) 808812.Google Scholar
[8] Lehmann, J et al., Soil Biol Biochem 43 (2011) 18121836.Google Scholar
[9] Erni, R, Aberration-corrected Imaging in Transmission Electron Microscopy: An Introduction, Imperial College Press/World Scientific Pub. Co., London, 2010, p. 304.Google Scholar
[10] Griffin, BJ, Scanning 33 (2011) 162173.Google Scholar
[11] Kuo, W et al., Platinum Metals Review 58 (2014) 106110.Google Scholar
Figure 0

Figure 1 Image sequence of a twinned gold particle on carbon. Under continuous beam scanning, atomic-resolution HAADF and SE images were acquired simultaneously: (a) HAADF at 0 s, (b) SE at 0 s, (c) HAADF at 120 s, and (d) SE at 80 s electron beam exposure.

Figure 1

Figure 2 Amorphous boron nitirde spheres on lacey carbon. (a) HAADF image showing no depth information. (b) SE image and (c) BSE image. Both of the latter images highlight which particle clusters lie above, and which lie below, the carbon support film.

Figure 2

Figure 3 Mineralized biochar fragment. (a) HAADF image and (b) SE image of the same particle. Fe-rich mineral particles decorate the surface of the slab-shaped biochar particle.

Figure 3

Figure 4 Highly porous TiO2 catalyst support imaged in (a) STEM brightfield and (b) SE image mode. Despite the high degree of internal porosity, very few pores penetrate the surface.

Figure 4

Figure 5 Sulphur nanoparticles dispersed on graphite: (a) HAADF image and (b) SE image. Inset (top left) shows details of the same region in both images. Contrast and resolution are superior in the SE image. Width of inset box = 120 nm.

Figure 5

Figure 6 SE images of tantalum oxide powder showing the effect of condenser aperture size on depth of field and signal-to-noise ratio: (a) 40 µm and (b) 10 µm.