Article contents
FIB Preparation of a NiO Wedge-Lamella and STEM X-Ray Microanalysis for the Determination of the Experimental k(O-Ni) Cliff-Lorimer Coefficient
Published online by Cambridge University Press: 03 January 2013
Abstract
A method for the fabrication of a wedge-shaped thin NiO lamella by focused ion beam is reported. The starting sample is an oxidized bulk single crystalline, ⟨100⟩ oriented, Ni commercial standard. The lamella is employed for the determination, by analytical electron microscopy at 200 kV of the experimental k(O-Ni) Cliff-Lorimer (G. Cliff & G.W. Lorimer, J Microsc103, 203–207, 1975) coefficient, according to the extrapolation method by Van Cappellen (E. Van Cappellen, Microsc Microstruct Microanal1, 1–22, 1990). The result thus obtained is compared to the theoretical k(O-Ni) values either implemented into the commercial software for X-ray microanalysis quantification of the scanning transmission electron microscopy/energy dispersive spectrometry equipment or calculated by the Monte Carlo method. Significant differences among the three values are found. This confirms that for a reliable quantification of binary alloys containing light elements, the choice of the Cliff-Lorimer coefficients is crucial and experimental values are recommended.
Keywords
- Type
- Software, Techniques and Equipment Development
- Information
- Copyright
- Copyright © Microscopy Society of America 2013
References
- 4
- Cited by