Hostname: page-component-586b7cd67f-rcrh6 Total loading time: 0 Render date: 2024-11-25T08:08:55.607Z Has data issue: false hasContentIssue false

Malposition of left atrial disc and feasibility of recapturing Figulla-II Occlutech septal occluder

Published online by Cambridge University Press:  18 February 2020

Masood Sadiq*
Affiliation:
Department of Paediatric Cardiology, The Children’s Hospital and Institute of Child Health, Lahore, Pakistan
Ahmad U. Qureshi
Affiliation:
Department of Paediatric Cardiology, The Children’s Hospital and Institute of Child Health, Lahore, Pakistan
Zahid Amin
Affiliation:
Division of Pediatric Cardiology, Congenital and Structural Heart Disease, Children’s Hospital of Georgia, Augusta University, Augusta, GA, USA
*
Author for Correspondence: Prof Masood Sadiq, FRCP, FRCPCH, Dean Office, Admin Block, The Children’s Hospital and Institute of Child Health, Ferozepur Road, Lahore, Pakistan. Tel: +92-42-99230614; Fax: +92-42-99230358. E-mail: [email protected]

Abstract

We report a case of Figulla-II Occlutech septal occluder malposition with residual shunt at posteriosuperior margin of an atrial septal defect. Improvising its bioptome type delivery cable, same system was used to recapture the device and redeploy it successfully. This report highlights a potential malfunction of Figulla-II Occlutech disc and the advantage of its delivery system for retrieval of the device.

Type
Brief Report
Copyright
© The Author(s), 2020. Published by Cambridge University Press

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Abaci, A, Unlu, S, Alsancak, Y, Kaya, U, Sezenoz, B. Short and long term complications of device closure of atrial septal defect and patent foramen ovale: meta-analysis of 28,142 patients from 203 studies. Catheter Cardiovasc Interv 2013; 82: 11231138.CrossRefGoogle Scholar
Moore, J, Hegde, S, El-Said, H, et al.Transcatheter device closure of atrial septal defects: a safety review. JACC Cardiovasc Interv 2013; 6: 433523.CrossRefGoogle ScholarPubMed
Haas, NA, Soetemann, DB, Ates, I, et al.Closure of secundum atrial septal defects by using the Occlutech occluder devices in more than 1300 patients: the IRFACODE Project: a retrospective case series. Catheter Cardiovasc Interv 2016; 88: 571581.CrossRefGoogle ScholarPubMed
Kilic, T, Ural, E, Sahin, T. Cobrahead deformity in the right atrial disc of a new-generation Occlutech Figulla Flex II atrial septal defect occluder device. Tex Heart Inst J 2015; 42: 454457.CrossRefGoogle ScholarPubMed
Sadiq, M, Rehman, AU, Mehmood, A. Successful retrieval of a figulla occlutech septal occluder-embolized device stability and potential solutions. Ann Pediatr Card 2016; 9: 161163.CrossRefGoogle ScholarPubMed
Jang, S, Kim, S. The retrieval of the embolized Occlutech Figulla Flex II in the left ventricle of child. J Am Coll Cardiol 2015; 65: S307S309.CrossRefGoogle Scholar
Jahrome, AK, Stella, PR, Leijdekkers, VJAbdominal aortic embolization of a Figulla Atrial Septum Occluder Device at the level of the celiac axis after an atrial septal defect closure: hybrid attempt. Vascular 2010; 18: 5961.CrossRefGoogle ScholarPubMed
Early and mid-term follow up outcomes of percutaneous closure of atrial septal defects using recent generation devices - A single center experience. Korean Circ J 2019; 49: 326335.CrossRefGoogle Scholar
Shebani, SO, Rehman, R, Taliotis, D, Magee, A. Techniques for transcatheter retrieval of the Occlutech ASD Device; United Kingdom–European multicenter report. Catheter Cardiovasc Interv 2017; 89: 690698.CrossRefGoogle ScholarPubMed
Goel, PK, Kapoor, A, Batra, A, Khanna, R. Transcatheter retrieval of embolized Amplatzer septal occluder. Tex Heart Inst J 2012; 39: 653656.Google ScholarPubMed